Series 1800 Washers
61301604086
1–11
General Machine Operation
Automatic Seven Phase
Treatment Cycle
The standard treatment cycle consists of a Prewash, Wash 1 and Soak,
Wash 2 and Soak, Rinse 1, Rinse 2, Rinse 3, and an Exhaust. All cycle
phases are adjustable from 0-60 minutes. All wash and rinse treatments are
recirculated under pump pressure. The cycle, once activated, is completely
automatic. Additional cycle treatment phases are available.
Operation
At the start of the processing period, the user places the load to be cleaned
n the chamber, closes the door, selects the desired cycle and presses the
cycle start button. The washer automatically proceeds through the
treatment process and alerts the user when the process is complete. The
user then opens the chamber door and removes the cleaned load.
Treatment Schedule
PREWASH
Hot water from house supply fills the sump or water remaining in the sump
from Rinse 3 of the previous cycle is recirculated through the jet spray
system under pump pressure and pumped to drain upon completion. Phase
time and temperature is user programmable form 0-60 minutes and from
120 - 190°F respectively.
WASH 1
Hot water from the house supply fills the sump and the wash agent is
automatically added. The treatment solution is recirculated through the jet
spray system and is pumped to drain upon completion. The phase time and
temperature are user-programmable from 0 to 60 minutes and from
120 - 190°F respectively.
Soak Option: A subsequent soak period may be programmed from
0-60 minutes, to permit the detergent solution to work on the load
before proceeding to the second wash and/or rinse treatments.
WASH 2
Hot water from the house supply fills the sump and the wash agent is
automatically added. The treatment solution is recirculated through the jet
spray system and is pumped to drain upon completion. The phase time and
temperature are user-programmable from 0 to 60 minutes and from
120 - 190°F respectively.
Soak Option: A subsequent soak period may be programmed from
0-60 minutes, to permit the detergent solution to work on the load
before proceeding to the rinse treatments.
RINSE 1
Hot water from the house supply fills the sump and is recirculated through
the jet spray system under pump pressure and pumped to drain upon
completion. The phase time and temperature are user-programmable from
0 to 60 minutes and from 120 - 190°F respectively.
Summary of Contents for Castle MTP 1800 Series
Page 1: ...SERIES 1800 CAGE AND BOTTLE WASHER OPERATING MANUAL 61301604086 Rev A...
Page 2: ......
Page 68: ...Troubleshooting 5 6...
Page 74: ...Installation Instructions 6 6...
Page 92: ...Optional Equipment 7 18...
Page 97: ......