Cleaning and Disinfection
The devices of the manufacturer were designed in such a way
that minimal effort is required for cleaning and disinfection.
This is why just a few tasks are necessary to keep Ambistik PFT
/ Ambistik CPET functional and clean. The following intervals
apply:
Component
Interval
Method
All touchable parts
Weekly
Wipe disinfection with
a soft cloth and a
weak soap solution.
9.1
Disinfection
When using a disinfectant that has not been tested and
approved by the manufacturer, the following steps must be
observed:
•
Preferably use agents that correspond in composition
to the approved agents. The composition is available
on data sheets, which we can provide on request.
•
Check bactericidal and virucidal effect suitable for the
intended use.
•
Use only disinfectants listed in public databases (e.g.
RKI)
•
Check data sheet for material compatibility with
plastics (especially polyoxymethylene, polystyrene,
acrylonitrile butadiene styrene, Makrolon®) as well as
metal
•
First test the disinfectant for material compatibility in an
inconspicuous place.
•
During the regular visual inspection for damage, also
pay attention to material changes (discoloration,
cracks, embrittlement...)
•
Observe long-term trend of calibrations for changes in
measuring function