7.3
Inserting the Flow Sensor
1.
Insert Spiraflow [2] into the handle
[1].
For this:
Align the Spiraflow so that the
interrupted guide rib [2a] and the nose
on the handle are on the same side
and
the pressure tapping holes [2b] show
in the direction of the handle.
2.
Snap the Spiraflow [2] into the
handle [1]!
For this:
Apply light pressure.
3.
Spiraflow not only must be fixed, but
also locked in place!
For this:
Check the correct position in the
handle on the basis of the Interrupted
guide rib and the nose and correct it if
necessary.