9.2.1
Flow Unit
1.
Remove the Ergoflow
For this:
Proceed in reverse order as described
in chap. 6.2.3 "Assembly Flow
Measuring System".
2.
Perform a wipe disinfection
For this:
Wipe all accessible surfaces with a wet
disinfecting cloth.
9.2.2
Ergoflow / Flow Sensor
Attention
Bodystik could be damaged.
Reason: destruction of the orifice. Therefore:
•
Do not clean the interior of the sensor, mechanical or
with hard water jet!
•
Do not use any disinfectants which contain high
chloride concentrations!
1.
Remove the flow sensor
(in reverse order as in chap. 6.2.3.2.
"Inserting the Flow Sensor" described)
2.
Visible contamination
wipe off with a soft cloth
3.
Put the sensor in disinfectant fluid
4.
Wash the sensor in distilled water
thoroughly
so that no residual dirt and
disinfectants remain
5.
Air-dry sensor until no moisture is
visible