Instructions for Use
Bodystik
Serial no.: xx|8|301|yyy and 2301xxxxx
Version: 7 | Release date: 02. August 2021
Page 77
7.6
Exchange of Disposable Products / Disinfection
7.6.1
Bacteria- and Virus Filter / Noseclip
Possible danger to life.
Reason: Cross-contamination. For this: Observe the
general medical principles! Therefore:
•
Do not use single use products more than once!
•
Dispose of bacteria / virus filter and noseclip after
each use!
The bacteria and virus filters provided by the manufacturer are
optimally adapted to the measuring systems and, due to the very
high filtration rate, offer maximum patient safety while protecting
against cross-contamination. The filters must also be disposed
after each examination or patient. The same applies to the
noseclip.
7.6.2
Bodystik / Cabin
Parts of the cabin which have come into contact with the patient
during the treatment must be treated with surface disinfection.
For further weekly cleaning and disinfection tasks, please see:
•
Chap. 9 “Cleaning and Disinfection”
Information about disposal of the disposable products can be
found in:
•
Chap. 11.3 "Disposal" / 11.3.3