Instructions for Use
Spirostik Blue
Serial no.:
2111xxxxx
Page 32
Version: 7 | Release date: 02. August 2021
3.3
Operation / Servicing and Maintenance
Danger to life.
Reason: Electric shock. Explosion. Device damage.
Unpredictable movement of metal parts. For this:
Comply with the required installation and operating
conditions! Therefore:
•
In principle, observe chap. 12
“Technical Specifications”!
•
Do not operate Spirostik Blue if there are
flammable or explosive gases in the room!
•
Do not operate Spirostik Blue near the magnetic
field of an MRT system!
Possible danger to life.
Reason: Electric shock. Cross-contamination.
Misdiagnosis caused by measurement error. Device
damage. Therefore:
•
Ensure that the used components are
undamaged and that you are working in a careful
way!
•
Do not use consumable items with a limited life
span after their use-by date has expired!
•
Prior to each use, visually inspect the complete
system (housing, connectors, etc.) for any
damages!
•
If there are any damages, do not operate the
system. The damaged parts must be replaced or
repaired properly!
•
Calibrate Spirostik Blue at the intervals stated!