BIOGAS 5000 gas analyser
OMBIO5KN4.1
Copyright Geotech
Page
39
View data
3)
Toggle through the reading by selecting ‘Key 4 – Scroll left’ and ‘Key 6 – Scroll right’
on the analyser. Select ‘Key 2 – Page up’ and ‘Key 8 – Page down’ to page through
the auxiliary channels listed.
4)
Select the soft-key ‘Filter’ to filter the data by sample point ID, or specify before or
after date. Press the soft-key ‘Exit’ to exit the filter menu and return to the ‘View
Data’ screen.
Filter data
5)
Select the soft-key ‘Delete’ followed by the appropriate soft-key to delete a single
reading or all filtered readings. Press soft-key ‘Cancel’ to cancel the deletion
request.
6)
Select the soft-key ‘Exit’ to exit the view data screen.
7.4.4
Set alarms
This option enables the operator to define the conditions for which an
alarm will be triggered. These conditions apply to the general operation
of the instrument and are not ID specific. A summary of the alarm
settings can be found in ‘Key 3 – Summary’.
1)
Select the ‘Menu’ button on the front of the analyser to display the ‘Device Settings’
menu. Press the soft key to display ‘Operation Settings’.
2)
Select ‘Key 4 – Set Alarms’ and the following menu is displayed: