5
+ VDC
mA Signal Return
Ground/Shielding Wire
Figure 1-1
5. Wire cable to the connector using the following diagram
6. Completely re-assemble the connector.
7. Plug the 4-20mA transducer into the
“4-20mA signal input” on the control panel
The transducer is now be ready to use with the 4-20mA Display.
The 4-20mA display comes pre-set from Geotech to operate with a 0-30 PSI transducer
and display PSI. If a different range is used (i.e. 0-50PSI) or you would like the display to
show water level, use the following steps to change the display.
Setting Display Units
1. Switch the ON/OFF switch to the ON position.
Screen should display a 0 value
EX 0.00; 0.01 etc.
2. Press the MENU key to enter Programming mode.
3. Press ACK key to access the Setup menu.
4. Press the ACK key to access the Input menu