27
Datalogger#1,2009,94,0939,30,3.22,23.44,9985.479,25.0,3
Datalogger#1,2009,94,0939,35,3.22,23.53,9985.686,25.0,4
To clear the ID enter a <SPACE> character as the ID. When the ID is cleared the arrays from the
logger will display beginning with the year. To display the current ID enter
ID
<ENTER>.
4.10 “LC” - Linear Conversion
Selects the linear conversion method for the instrument reading. See Section 4.8, “Gauge
Settings” for more information.
*LC
Linear conversion selected.
*
4.11 “L” - Display Log Intervals
Displays all six log intervals.
*L
Log Intervals List
-------------------------------------------------
Interval #1 Length: 10 Iterations: 100
Interval #2 Length: 20 Iterations: 90
Interval #3 Length: 30 Iterations: 80
Interval #4 Length: 40 Iterations: 70
Interval #5 Length: 50 Iterations: 60
Interval #6 Length: 60 Iterations: 0
This command has no affect on the current interval (scan or log). If logging is started and log
intervals are enabled the iterations value will be followed by the number of readings left at that
interval. For example:
*L
Log Intervals List
------------------------------------------------
Interval #1 Length: 10 Iterations: 100/58
Interval #2 Length: 20 Iterations: 90/90
Interval #3 Length: 30 Iterations: 80/80