Introduction

How to Operate

COMBO METER

pH-EC-TDS-TEMP

General Description

GI-98129 and GI-98130 are waterproof pH, EC/TDS and

temperature meters.  The housing has been completely sealed

against humidity and designed to float.  All pH and EC/TDS

readings are automatically temperature compensated (ATC),

and temperature values can be displayed in °C or °F units.

For EC/TDS readings, the EC/TDS conversion factor (CONV)

is selectable by the user, as well as the temperature

compensation coefficient ß (BETA).  The meters can be

calibrated at one or two points for pH (with auto buffer

recognition and at one point for EC/TDS). Measurements are

highly accurate with a unique stability indicator right on the LCD.

A low battery symbol warns the user when the batteries are to

be replaced.  In addition, the Battery Error Prevention System

(BEPS) avoids erroneous reading caused by low voltage level

by turning the meter off.  The pH electrode supplied with the

meter is replaceable by the user.  The stainless steel

encapsulated temperature sensor facilitates faster and more

accurate temperature measurement and compensation.

Pocket Pen

Models GI-98129

and GI-98130

Familiarizing yourself with your new meter

A. Turning your meter “on”:

 To turn your meter on, press and

hold the O/MODE button for 2-3 seconds. Notice that every

time you turn the pen on, it will flash all icons used on the LCD

for a few seconds. This is immediately followed by the

percentage of remaining battery life – example: % 95 BATT).

After the battery life flashes the pen will settle into “normal

measurement mode”.  Now would be a good time to get familiar

with the measurement options.  This is located to the right of

the ATC icon and above the “primary display”. Press the Set/

Hold button repeatedly and the measurement options will scroll

between PPM, uS, & pH (or ppt, mS, and pH for the GI-98130).

B. To turn your meter “off”:

 Make sure the pen is normal

measurement mode, then press and hold O/MODE button.  The

LCD will read “OFF”.  Immediately release the O/MODE button,

your meter will now be off.

C.

 

Freezing the display:

 press the set/hold button for 2-3

seconds until the secondary display reads “HOLD”.  Pressing

either button will return you to normal measurement mode.

NOTE

: Before taking a measurement, follow the calibration

instructions to ensure accurate readings.  For accuracy it is

recommended to carry out a two point calibration. For best

results always use Genesis Instrument buffers and calibration

solutions. If measurements are taken in different samples

successively, rinse the electrode thoroughly to eliminate cross-

contamination and after cleaning, rinse the electrode with some

of the sample to be measured. When not in use, rinse the

electrode with water and store in a few drops of storage solution.

Do not use distilled water. If electrode dries out soak in storage

solution for one hour. Clean monthly in cleaning solution.

Electrode Replacement

The pH electrode can be easily removed with the tool provided.

Insert the tool into the electrode cavity as shown below.

Remove the electrode by rotating it counterclockwise and then

pull out. Insert the new electrode by following the above

instruction in reverse.

Specifications:

                               

                    GI 98129 low range

               

GI 98130 high range

Range

                                   

EC

          

                0 to 3999 µS/cm

                       

0.00 to 20.00 mS/cm

                                              

TDS

        

                 0 to 2000 ppm

                   0.00 to 10.00 ppt

                                              

pH

          

                 0.00 to 14.00 pH                   0.00 to 14.00 pH

                                            

Temp

     

                 0 to 60°C or 32 to 140°F          0 to 60°C or 32 to 140°F

Resolution

                           

EC

          

                1 µS/cm

                    0.01 mS/cm

                                               

TDS

        

                1 p

pm

                                         

0.01 ppt

                                               

pH

                            

0.01

                                            

0.01

                                               

Temp

     

                0.1 °C or 0.1 °F

                   0.1 °C or 0.1 °F

Accuracy @ 20°C/68°F

      

EC/TDS

  

               ±2% F.S.

                   ±2% F.S.

                                               

pH

           

               ±0.05

                   ±0.05

                                               

Temp

       

              ±0.5 °C or ±1 °F

                   ±0.5 °C or ±1 °F

Typical EMC Deviation

      

EC/TDS

  

               ±2% F.S.

                   ±2% F.S.

                                               

pH

                            

±0.02

                   ±0.02

                                              

Temp

    

                 ±1 °

C o

r ±

1 °

F                          ±

1 °

C o

r ±

1 °

F

Calibration

                           

EC/TDS

                         

Automatic at 1 point

                                               

pH

         1 or 2 points  with  (pH 4.01, 7.01 & 10.01 )

Temp Compensation

         

EC/TDS/pH

   

               Automatic

EC / TDS Conv Factor

         Adjustable from 0.45 to 1.00

Environment

          

0 to 50°C (32 to 122°F); R.H. 100%

Battery Type / Life

          

4 x 1.5V / 100 hours

Auto Shut-off

          

After 8 minutes

Dimensions

          

163 x 40 x 26 mm (6.4 x 1.6 x 1.0")

Weight

           

85 g (3 oz.)

Specifications

Genesis Instruments is a trademark of Green Air Products Inc.

Genesis Instruments are warranted for one (1) year against

defects in workmanship and materials when used for their

intended purpose and maintained according to instructions. The

electrode is warranted for a period of six months. This warranty

is limited to repair or replacement free of charge. Damages

due to accident, misuse, tampering or lack of prescribed

maintenance are not covered. If service is required, contact

the dealer from whom you purchased the instrument. If under

warranty, report the model number, date of purchase, serial

number and the nature of the failure. If the repair is not covered

by the warranty, you will be notified of the charges incurred.

When shipping any instrument, make sure it is properly

packaged for complete protection.

Warranty