K Legacy Information - Input Cards
K.1
Basic amplifier non
-
isolated input cards
K.1.1
GN810 Basic 200kS/s input card
l
8 analog channels
l
Single
-
ended inputs
l
±
1 V to
±
50 V input range
l
User selectable digital Bessel and FIR filters
l
200 kS/s sample rate
l
16 bit resolution
l
128 MB memory
l
Single metal BNC for each channel
The GEN DAQ Basic 200 kS/s input card is a general purpose
signal conditioner for use with voltage inputs, externally
conditioned signals or probes and current clamps.
The basic signal conditioner provides eight channels of single
-
ended voltage inputs from
±
1 V to
±
50 V Full Scale with full offset
and auto
-
zero capability. Every channel is equipped with an
independent full range input amplifier, a seven
-
pole Bessel and
Butterworth anti
-
alias filter, a 16 bit Analog
-
to
-
Digital converter
and several selections of digital filtering.
The on
-
board transient memory size is 64 Mega
-
Samples (128
Mega
-
Bytes). The memory is shared among enabled channels.
Each channel also features two set
-
points for trigger or alarm
purposes. Extensive acquisition and trigger modes allow for many
different ways to capture valuable data, even at the highest sample
rates. All channels are synchronously sampled at full speed without multiplexing
and almost immeasurable crosstalk. The model uses standard metal BNC
connectors, whose shells are connected to ground. The inputs have an
impedance of 1 MΩ and are compatible with probes and current clamps.
For more information on the GN810 Basic 200K input card, please refer to
"B2632-3.1 en (GEN series GN810)" on page 984.
1
2
3
4
5
6
7
8
1 MW 250Vpk
!
1
2
3
!
4
5
6
7
8
GEN3t
950
I3995-3.1 en HBM: public