Isolation
Input signal to chassis 1000 V RMS, 600 V CAT II (REINFORCED)
Channel to chassis 1000 V RMS, 600 V CAT II (REINFORCED)
Channel to channel 2000 V RMS, (BASIC)
±
10
0
0
V RMS
60
0
V CAT
II
Rein
forced
-
+
ADC
Isolated channel
±
1000
V RMS
600
V CAT II
Reinforced
±
2
00
0
V
RMS
±
2
0
00
V RMS
-
+
ADC
Isolated channel
Chassis
±
1
00
0
V
RMS
6
00
V
CA
T
I
I
±1000 V RMS
600 V CAT II Reinforced
±1000 V RMS
600 V CAT II Reinforced
Figure 12.13:
Isolation 1kV card overview
l
The isolation between the channel and chassis is classified as
REINFORCED
. This can be seen as double isolation, which is necessary
because the chassis might be accessible (conductive parts can be touched)
to users (personal safety).
l
Isolation between channels is
BASIC
, since a channel is not accessible.
Therefore, there is no direct risk to users (product safety).
l
REINFORCED
or
DOUBLE
isolation has higher test values than
BASIC
isolation.
GEN3t
266
I3995-3.1 en HBM: public