4.4.2 Remote operation and commands
General Photonics provides a test program for remote control of the POS-002. See
section 4.4.4 for installation and operation instructions for the program.
The following steps and commands are recommended for remote operation of the POS-
002 using the RS-232 communication port.
1.
Connect POS-002 to PC using provided cables.
2.
Make sure that the RS-232 port settings on the control computer are as follows:
a.
Baud rate: 9600 bps
b.
Framing: asynchronous
c.
Data bits:8
d.
Parity bit: None
e.
Stop bit: 1
f.
Flow control: None
3.
Power on the POS-002 by supplying DC ± 12V.
4.
Make sure that the COM port number setting for the communications program
used to send the commands matches the COM port connected to the POS-002.
5.
Referring to Table 2 below, send a command string (e.g. “*IDN?”) to the POS-
002 through the RS-232 port. There are many programming languages that
support serial communications, including Visual Basic, LabView and C.
Table 2 RS-232 Command List:
Query Commands
Command Description
Response
*IDN?
Query the Product Number.
*IDN GP-POS-002V2.0-E#
*VER?
Query the Firmware Version.
*VER POS002-V21-
020513A#
(version dependent)
*MOD?
Query operation status.
*MOD ENA# POS enabled
*MOD DIS# POS disabled
*SER?
Query serial number
*1047000004xx#
Control Commands
*ENA#
Enable polarization tracker.
*E00#
*DIS#
Disable polarization tracker.
*E00#
Setup Commands
*FUN?
Query current tuning mode.
“Variable” or “Fixed”
*VAR#
Set tuning mode to “variable step size”.
*E00#
*FIX#
Set tuning mode to “fixed step size”.
*E00#
Document #: GP-UM-POS-002-21
Page 14 of 30
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
苏州波弗光电科技有限公司
网址:www.bonphot.com
电话:0512-62828421