GM7702 User Manual
- 36 -
Exception code response
Code
Definition
Description
02
illegal data address
It means the data address received is illegal.
03
illegal data
The data wrote in is illegal.
04
indicator fault
Indicator fault when operation.
07
invalid
Order received is invalid.
8.2.1 Modbus Address
PLC addr.
Display
addr.
Description
Read only register (function code: 0x03)
40001
0000
present weight value (4 bytes with symbol)
40002
0001
40003
0002
D0
: stable
D1
: zero point
D2
: symbol
D3
: overflow
D4
: weight overlimit
D5
: weight underlimit
D6
: voltage overlimit
D7
: voltage underlimit