GM7702 User Manual
- 24 -
4.2.2 Weight Calibration
In order to meet the demands of multiple weight standards, user could set parameters of
weight point 1, weight point 2, weight point 3, weight point 4, which are helpful to calibrate
gain weight.
Weight:
0.000kg
Voltage:
0.000mV
Adding weight
Weight:
0.302kg
Voltage:
1.000mV
Pressing ENTER
In status of empty scale, pressing confirm key to
enter weight value 1, display weight and voltage
will show 0.
Adding weights, also recording both weight
value and voltage