GEMORO PLATINUM PRO1000V
8
CAUTION: Never attempt to calibrate the PRO1000V
unless you first have the required 1000g calibration
weight (sold separately Item #9748) with the scale.
1. Press the ON/MEM button to turn the PRO1000V
on. Allow the scale to warm up and stabilize for
approximately two minutes before commencing
with the calibration procedure.
2. Remove all items from the weighing platform and
press the tare button to zero the display.
3. After observing a stable 0.0g while in the (g) gram
mode, press and hold the MODE/CAL button until
the calibration weight value of 1000g appears on
the display.
4. Gently place the 1000g calibration weight on the
center of the stainless steel weighing platform.
Wait approximately 4 seconds as the unit performs
internal calibration. IMPORTANT- Do not disturb
the PRO1000V during this time as it may impact
the accuracy of the calibration.
5. The display will then show “F” which indicates that
the calibration procedure has been successful.
If “E” appears on the display instead of “F”,
this indicates that an error has occurred in
the calibration procedure and the calibration
procedure should be repeated.