September 2007
5-10
Basic Programming
SET LOW INPUT VALUE
Option
The
Set Low Input Value option
lets you specify the %O
2
value in the
gas that corresponds to a 4.00 mA current at the analog output.
Note:
This option appears if
DriftCal
is
ON
or
OFF
and if a
one- or
two-gas calibration
has been selected.
To use the
Set Low Input Value
option, proceed as follows:
1.
Press
[N]
until the following prompt appears:
2.
At the above screen, do one of the following:
•
Press
[Y]
or
[ENTER]
to accept the current low input value (0.00
in the above example). Then, go to Step 4.
•
Use the numeric keys to enter a new low input value (5.00 in
the example that follows). Then, go to Step 3.
3.
If you entered a new low input value:
4.
Complete the procedure as follows:
You may now do one of the following:
•
Press
[N]
as many times as necessary to select another
Basic Menu
option. Then, press
[Y]
and proceed to the appropriate section for
instructions.
•
Press
[N]
repeatedly until the
Resume?
option appears. Then, press
[Y]
to return to
Operate Mode
.
BASIC MENU
Press
[Y]
or
[ENTER]
to set the
low (4 mA) input value.
Set Low Input Value?
mA Output 4 mA Value
Go to Step 2.
%O2 [0.00]:
mA Output 4 mA Value
Press
[Y]
or
[ENTER]
to accept
the new low (4 mA) input value.
%O2 [0.00]:5.00
BASIC MENU
The
Low Input Value
has been
set.
Set Low Input Value?
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...