Advanced Programming
7-13
September 2007
GAS VALUE EXP FILTER
Option (cont.)
The
Set Tau(up)
option (cont.):
You may now do one of the following:
•
Press
[N]
as many times as necessary to select another
Gas Value
Exp Filter
menu option. Then, press
[Y]
and proceed to the
appropriate section of this chapter for instructions.
•
Press
[N]
until you reach the
Done?
option of the
Gas Value Exp
Filter
menu, then press
[Y]
to return to the
Calibrate System
menu.
The
Set Tau(down)
option:
The
Tau(down)
parameter is the diffusion constant of the background
gas into oxygen. Its default value is 27.0, but values of 15.0-35.0 are
acceptable (oscillation increases as the value increases). To set this
parameter, proceed as follows:
You may now do one of the following:
•
Press
[N]
as many times as necessary to select another
Gas Value
Exp Filter
menu option. Then, press
[Y]
and proceed to the
appropriate section of this chapter for instructions.
•
Press
[N]
until you reach the
Done?
option of the
Gas Value Exp
Filter
menu, then press
[Y]
to return to the
Calibrate System
menu.
Gas Value Exp Filter:
You have now completed the
programming of this submenu.
Set Tau(up)?
Gas Value Exp Filter:
Press
[N]
until this prompt
appears, then press
[Y]
or
[Enter]
.
Set Tau(down)?
Tau(down) Value:
To accept the current value, press
[Y]
or
[Enter]
. To change the
current value, use the numeric
keys to enter a new value and
press
[Y]
or
[Enter]
twice.
Tau [xx.x]
Gas Value Exp Filter:
You have now completed the
programming of this submenu.
Set Tau(down)?
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...