General Programming
6-37
September 2007
TEST INPUTS Menu
The
Test Inputs?
menu is used to monitor a variety of XMO2 input
signals. It includes the following nine options:
•
O2 Input (raw oxygen sensor mV signal)
•
Comp Input (background gas compensation mV signal)
•
Pressure Input (pressure sensor mV signal)
•
Raw Bridge Input (mV reading across the Wheatstone bridge)
•
Offset Input
•
Temperature Input (raw temperature sensor mV signal)
•
Bridge Power Input
•
Analog Ground Input
•
Done (exit the menu)
To view the desired test input(s), refer to the menu map in Figure C-6
on page C-6, and proceed to the appropriate section.
O2 INPUT Option
The
O2 Input
option lets you monitor the raw oxygen sensor input
signal. The acceptable range for this signal is -512 mV to +512 mV.
Note:
A reading below -512 mV indicates that the unit has lost
temperature control. To correct the problem, decrease the gas
flow rate to 1 SCFH. If the problem persists, autozero the
Wheatstone bridge (see Chapter 7,
Advanced Programming
).
The current
O2 Input
signal is displayed (see your Calibration Sheet
for the expected O2 mV readings at each calibration point).
This completes the prompts for the
O2 Input
option, and the XMO2
returns to the
Test Inputs?
menu.
GENERAL MENU
Press
[Y]
or
[Enter]
to enter the
menu, or press
[N]
to move to the
next
General Menu
option.
Test Inputs?
TEST INPUTS
Press
[Y]
or
[Enter]
to view the
O2 input, or press
[N]
to move to
the next
Test Inputs?
option.
O2 Input?
O2 Input
When ready, press any key to
return to the
Test Inputs?
menu:
-xx.x O2 mV
Summary of Contents for XMO2
Page 8: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 41: ...Chapter 4 ...
Page 52: ...Chapter 5 ...
Page 66: ...Chapter 6 ...
Page 114: ...Chapter 7 ...
Page 146: ...Chapter 8 ...
Page 159: ...Appendix A ...
Page 165: ...Appendix B ...
Page 177: ...Appendix C ...