Section 6-3 - LCD Monitor Adjustment
6-3-2-3
Color Temperature (RGB)
1.) Press the
UTILITIES
key on the control panel.
2.) Touch the MONITOR button to display the “Monitor Menu” on the screen (see:
3.) Touch the OSD CONTROLS button to invoke the “OSD Adjust” menu.
4.) Click on the
RGB
tab.
5.) Adjust Color Temperature - RED.
6.) Adjust Color Temperature - GREEN.
7.) Adjust Color Temperature - BLUE.
8.) Close the “OSD Adjust” menu with [OK].
9.) Touch the EXIT button to store settings and exit the “Monitor Menu”.
Figure 6-4 OSD Adjust - RGB (e.g., MDM100 at BT12)