Probes
13-8
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
Maximum probe temperature
NOTE:
Lens temperature measured under following conditions per
IEC 60601-2-37 Ed2:
1. Thermocouple is placed at the maximum temperature spot
of the lens. Ambient temperature at 23 °C ±3 °C.
Probe
Max Temp (°C)
(Simulated use)
Max Temp (°C)
(Still air)
M5Sc-D (Vivid S70 only)
42.0
42.2
3Sc-RS
41.1
37.1
6S-D
41.3
39.9
12S-D
41.5
33.1
9L-D
41.6
37.6
11L-D
41.8
38.0
L8-18i-D
41.9
36.5
C1-5-D
41.1
40.5
C1-6-D
41.4
39.5
C2-9-D
38.6
37.3
C3-10-D
41.1
36.8
IC5-9-D
41.9
35.4
2D (P2D)
35.7
35.5
6D (P6D)
34.8
28.1
6VT-D
40.8
36.0
6Tc-RS
41.7
37.9
9T-RS
41.6
33.6
10T-D (Vivid S70 only)
41.9
42.5
AcuNav
TM
10F
37.0
34.9
AcuNav
TM
8F
37.8
33.6
SoundStar
TM
3D 10F
37.0
34.9
SoundStar
TM
eco 10F
37.0
34.9
SoundStar
TM
eco 8F
37.8
33.6
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......