Measurement Package
Vivid S70 / S60
–
User Manual
12-21
BC092760-1EN
01
Measurement package configuration - example
The following example based on calculation of AV CO (Cardiac
Output by Aortic Flow) describes how to configure the
measurement package so that necessary measurements and
the resulting calculations are displayed on screen.
Calculation of Cardiac Output by Aortic Flow requires the
measurement of:
•
AV diameter located in the folder
Dimension
(2D mode)
•
AV VTI located in the folder
Aortic
(Doppler AV Trace).
•
Heart rate
If a calculated parameter (e.g. AV CO in AV Trace
measurement) requires another parameter to be calculated (e.g.
AV Diam) the user must first measure the required parameter
(e.g. AV Diam) before the dependent parameter (e.g AV CO in
AV Trace) gets calculated.
Configuration of the Measurement menu
If the AV diameter measurement is not present in the folder
Dimension
in the
Measurement
menu, follow the following
procedure:
1. Press
Utility
/
Config
on the Touch panel and select the
category
Measure/Text
.
The
Measurement menu
sheet is displayed (Figure 12-10).
2. AV Diam is a 2D measurement, make sure that
2D
is
checked in the
Measurement
sheet.
3. Select folder
Dimension
in the
Measurement
menu.
A list of all available measurements for the selected folder is
displayed in the
Measurement menu
sheet.
4. Check the box in front of
AV Diam
.
The AV Diam measurement is displayed in the folder
Dimension
in the
Measurement
menu.
5. For the AV VTI measurement, check
Doppler
in the
Measurement menu
sheet and select the folder
Aortic
in
the
Measurement
menu.
6. Check the box in front of
AV Trace
.
The AV Trace measurement is displayed in the folder
Aortic
in the
Measurement
menu.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......