Using Q Analysis
Vivid S70 / S60
–
User Manual
9-11
BC092760-1EN
01
Manual tracking of the sample area (dynamic anchored sample area)
The sample area can be moved within the loop to ensure that
data in the trace is generated from the same anatomical location
during the cyclic motion of the heart.
1. Freeze the image and set a sample area over a region of
interest.
Note the anatomical location of the sample area.
2. Press
Trackball
until the
Scroll
trackball assignment is
selected. Select a new frame.
3. Press
Trackball
until the
QA
trackball assignment is
selected.
4. Drag the sample area to the corresponding anatomical
location in the new frame.
When the sample area is anchored in more than one frame,
linear interpolation is performed, so that the sample area is
smoothly moved between the anchored positions in the
selected frames when running the cineloop.
NOTE:
In the original frame and this particular frame the sample
area is marked with an anchor.
5. Press
Trackball
until the
Scroll
trackball assignment is
selected.
6. Scroll through the cineloop and control that the sample area
follows the moving anatomical structure.
7. Add anchored sample areas in several frames to obtain a
more accurate displacement of the sample area.
To move a dynamic anchored sample area
1. Freeze the image.
2. Press
Trackball
until the
Scroll
trackball assignment is
selected.
3. Scroll through the cineloop to display one of the frames
where the sample area was anchored.
NOTE:
In these frames the sample area is marked with an anchor.
4. Press
Trackball
until the
QA
trackball assignment is
selected.
5. Place the cursor on the sample area to move, in one of the
Cineloop
windows.
6. Press
Select
.
The sample area is unanchored.
7. Drag the sample area to a new location.
8. Press
Select
to anchor the sample area to the new location.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......