Advanced cardiac measurements and analysis
Vivid S70 / S60
–
User Manual
8-13
BC092760-1EN
01
Figure 8-5. TSI Generic Time to peak measurement screen
Segment Time to peak measurements
1. Acquire TSI loops from all three apical views.
2. Press
Measure
and select the
TSI time
study.
The TSI loop freezes at the TSI end frame.
The first measurement in the study is automatically selected
(see Figure 8-6).
3. Place a point in the middle of the corresponding segment in
the TSI image.
The Time to peak value for the segment is displayed in the
Measurement result
window.
4. Perform a measurement for all basal and mid-level
segments in all three apical views.
In addition to the Time to peak value for each segment, the
following TSI indexes are calculated:
•
Septal lateral delay
: difference in Time to peak velocity
in the basal lateral wall and basal septum.
•
Septal posterior delay
: difference in Time to peak
velocity in the basal posterior wall and the basal
antero-septum.
•
Basal seg. max diff.
: difference between the maximum
and minimum time to peak measurements in the six
basal segments. Requires at least four of the six basal
segment measurements.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......