Chapter 10 - Care and Maintenance
10-9
P R E L I M I N A R Y
10-5-3
Physical Inspection
Table 10-8 Physical Checks
Step
Item
Description
1
Labeling
Verify that all Ultrasound system labeling is present and in readable condition.
2
Scratches & Dents
Inspect the exterior for dents, scratches or cracks.
3
Covers
Where applicable, verify all covers are secured in place and are properly aligned with other covers.
Replace any covers that are damaged.
4
Input Power
Refer to:
Mains Cable Inspection on page 10 - 8
5
External I/O
Check all connectors for damage.
6
Wheels and Brakes
• Where applicable, check all wheels and casters for wear and verify operation of foot brake, to stop
the Ultrasound system from moving, and release mechanism.
• Where applicable, check all wheel locks and wheel swiveling for proper operation.
7
Control Panel Movement
• Where applicable, verify ease of Operator Panel (Operator Control Panel) movement in all
acceptable directions.
• Where applicable, ensure that the Control panel latches in position as required.
8
Control Panel Lighting
Check for proper operation of all operator panel and TGC lights.
9
LCD
Inspect the LCD Display for scratches and bad pixels.
Verify proper operation of Contrast and Brightness controls.
Where applicable, confirm that the LCD arm allows:
• Swiveling the screen to the left and to the right
• Folding the screen to the locked position
• Release and adjustment backwards and forwards
• Can be adjusted in the up/down positions.
Note: LCD Arm movement may vary and is not applicable to all Ultrasound systems.
10
Monitor Light
Check for proper operation of any monitor lighting, if available.
11
Cables and Connectors
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
special attention to probe strain or bend reliefs.
12
Shielding and Covers
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems while scanning.
13
Control Panel
Inspect alphanumeric keyboard and Operator Panel. Record any damaged or missing items.
14
Probe Holders
Where applicable, inspect the Probe Holders for cracks or damage.
15
Power and System
Status Indicators
Check for proper operation of all Power and System Status Indicators.
16
Battery
Where applicable, check that the battery is not damaged, does not leak, does not emit an odor, and
is not deformed or discolored. Observe all warnings and cautions for battery handling, recharging,
storing, and/or disposal.
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......