Section 7-2 - Service Safety Considerations
P R E L I M I N A R Y
7-5-4-6
Hardware Problems
A hardware issue will typically be an error/malfunction on a card.
7-5-4-7
Software Problems
Check if a newer software version is available. A software update may include noise fixes. If needed,
update the software.
7-5-5
Different Power Outlet
Connect the unit to another power outlet and verify if the noise changes or disappear.
NOTE:
GE requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This
dedicated power shall originate at the last distribution panel before the Ultrasound system.
The Vivid™ S60/Vivid™ S70 will function on voltages from 100-240 Volts and 50 or 60 Hz
.
However, if
using 220 volt power in North America, then a center tapped power source is required
.
Sites with a mains power system with defined Neutral and Live
:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet
.
Sites with a mains power system without a defined Neutral
:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet
.
7-5-6
Different System
Try another Vivid™ S60/Vivid™ S70 scanner at the same location and look for the same noise. If the
noise is present on the new system too, the noise is most likely from an external source/equipment.
7-5-7
Different Location
Move the scanner to another location and verify if the noise changes or disappear. This may help you
to locate an external noise source.
Try to move the scanner to:
•
another location inside the room
•
another room
•
another floor
7-5-8
Disconnect External Cables
Disconnect all external cables (network, all unused probes, ECG leads and verify if the noise
disappears.
This page was intentionally left blank.
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......