7-26
Section 7-4 - Performing Front End (FE) Diagnostics
P R E L I M I N A R Y
Section 7-4
Performing Front End (FE) Diagnostics
NOTE:
When performing Front End Diagnostic tests, it is strongly recommended to start with the
Full System
Test (FE and BEP)
especially if the system shows signs of an intermittent problem. (For details, see
3-1-1 "Full System Test (FE and BEP)" on page 7-22
).
The Front End diagnostic tests can be initiated manually and may be performed in any order. However,
due to the functional dependencies between each of the Front End boards, it is recommended to
perform the tests on the boards in the following order:
1.) RFI board
2.) TR32 boards
3.) PIB board
4.) PSB board
Performing the tests in the proper sequence will help to derive the correct conclusions regarding the
source of the problem.
As described in this section, the testing sequence can be paused and restarted at any stage, providing
a quick and easy method of reviewing any detected faults. For instructions on performing a specific test,
refer to the relevant test in the appropriate sub-sections:
•
Calibration Tests
, as described in the
7-4-2 "Calibration Tests" on page 7-27
.
•
RFI Diagnostic Tests
, as described in the
7-4-3 "Radio Frequency Interface (RFI) Diagnostic
.
•
TR 32 Diagnostic Tests
, as described in the
7-4-4 "TR 32 Diagnostic Tests" on page 7-47
•
Probe Interface Board Diagnostic Tests
, as described in the
Diagnostic Tests" on page 7-62
•
Probe Select Board Diagnostic Tests
Diagnostic Tests" on page 7-63
NOTE:
If the system must be re-booted, do so. Continue testing from the point at which you were required to
reboot. If a failure occurs, re-boot the system and test again to verify the outcome of the test.
Summary of Contents for Vivid S5 N
Page 2: ......
Page 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Page 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Page 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Page 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Page 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Page 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Page 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Page 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Page 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Page 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Page 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Page 833: ......