GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
8-48
Section 8-3 - Internal Component Replacement Procedures
8-3-5
DC-to-DC Power Supply Replacement Procedure
8-3-5-1
Tools
Use the appropriate flat and Phillips-type screw drivers, as indicated in the HVPS replacement
procedure.
8-3-5-2
Time Required
15 minutes
8-3-5-3
Preparation
Shut down the Vivid q N ultrasound unit, as described in
8-3-5-4
DC-to-DC Power Supply Removal Procedure
1) Make sure the Vivid q N LCD Display cover is securely closed and place the scanner face down.
Remove the battery as described in the
"Battery Removal Procedure" on page 8-4
.
2.) Remove the bearing handle, as described in the
"Bearing Handle Removal Procedure" on page 8-2
.
3.) Disconnect each of the two Power Supply cables from the DC Power Supply (HVPS-to-BEP and
HVPS-to-RFI), as shown below.
4.) Loosen the five screws that secure the DC Power Supply to the Bottom Assembly, as shown in
Figure 8-56 Disconnecting the DC Power Supply Cables