Chapter 10 - Care & maintenance
10 - 7
10-5-2
Functional checks
NOTE:
See also Chapter 4.
The functional checks take about 60 minutes to perform. Refer to the VIVID E9 user documentation
whenever necessary.
10-5-2-1
System checks
Table 10-3 System Functional Checks
Step
Item
Description
1.
B-Mode
Verify basic B-Mode (2D) operation. Check the basic controls that affect this mode of operation.
2.
CF-Mode
Verify basic CF-Mode (Color Flow Mode) operation. Check the basic controls that affect this mode
of operation.
3.
Doppler Modes
Verify basic Doppler operation (PW and CW if available). Check the basic system controls that
affect this mode of operation.
4.
M-Mode
Verify basic M-Mode operation. Check the basic controls that affect this mode of operation.
5.
4D Mode
Where applicable, verify basic 4D Mode operation.
Check the basic controls that affect this mode of operation.
6.
*Applicable
Software Options
Verify the basic operation of all optional modes such as Multi-Image, 4D, Contrast, Harmonics,
Cine, Stress Echo,
«
etc.
Check the basic system controls that affect each options operation.
7.
System Diagnostic Perform the Automatic Tests.
8.
Operator Panel
test
Perform the Operator Panel Test Procedure.
9.
Touch Panel
Verify basic Touch Panel display functions.
10.
Monitor
Verify basic Monitor display functions.
11.
Peripherals
See:
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......