GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 10 - Care & maintenance
10 - 3
Section 10-3
Maintenance task schedule
10-3-1
How often should maintenance tasks be performed?
The Care and Maintenance task schedule (provided on page
) specifies how often your VIVID E9
should be serviced and outlines items requiring special attention.
NOTE:
It is the customer’s responsibility to ensure the VIVID E9 care & maintenance is performed as
scheduled in order to retain its high level of safety, dependability and performance.
Your GE Service Representative has an in-depth knowledge of your VIVID E9 ultrasound scanning
system and can best provide competent, efficient service. Contact GE for coverage information and/or
price for service.
The service procedures and recommended intervals shown in the Care and Maintenance Task
Schedule assumes that you use your VIVID E9 for an average patient load (10-12 per day) and not use
it as a primary mobile unit which is transported between diagnostic facilities.
NOTE:
If conditions exist which exceed typical usage and patient load, then it is recommended to
increase the periodic maintenance frequencies.
.
Table 10-1 Customer Care Schedule sheet 1 of 2
Service at Indicated Time
Daily
Weekly
Monthly
Per
Facilities
QA
Program
Notes
Clean Probes
•*
* or before each use
Clean Probe Holders
•
Clean Air Filters
•
more frequently depending on
your environment
Clean Console
•
Inspect AC Mains Cable
•
Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Monitor and Touch Panel
•
Inspect Wheels, Casters, brakes and
Swivel Locks
•
Mobile Unit Check Daily
Check Operator Panel Movement
•
Mobile Unit Check Daily
Console Leakage Current Checks
•
also after corrective
maintenance
Peripheral Leakage Current Checks
•
also after corrective
maintenance
Surface Probe Leakage Current Checks
•
also after corrective
maintenance
Endocavity Probe Leakage Current
Checks
•
also after corrective
maintenance
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......