Section 2-2 - General console requirements
Section 2-2
General console requirements
2-2-1
Console environmental requirements
2-2-1-1
If the unit is very cold or hot
2-2-1-2
Environmental specifications for VIVID E9 scanners
Temperature Limits:
•
Operation:
10 to 35 ºC (50 to 95 ºF)
•
Storage and Transport:
-20 to 60 ºC (-4 to 140 ºF)
Humidity Limits:
•
Operation:
30 - 60% rH non-condensing
•
Storage and Transport:
30 - 95% rH non-condensing
Air Pressure Limits:
•
Operation:
700-1060 hPa
•
Storage and Transport:
700-1060 hPa
Heat Dissipation:
•
3800 BTU/h
CAUTION
IF THE UNIT IS VERY COLD OR HOT, DO NOT TURN ON ITS POWER UNTIL IT HAS HAD A
CHANCE TO ACCLIMATE TO ITS OPERATING ENVIRONMENT.
Table 2-1
VIVID E9 Acclimate Time
o
C
-40
-35
-30
-25
-20
-15
-10
-5
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
o
F
-40
-31
-22
-13
-4
5
14
23
32
41
50
59
68
77
86
95
104 113 122 131 140 149 158
Hrs
20
18
16
14
12
10
8
6
4
2
0
0
0
0
0
0
0
2
4
6
8
10
12
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......