8 - 168
Section 8-6 - Top Console Parts Replacement
8-6-34
Replacing the Probe Holder Inserts
8-6-34-1
Introduction
The Probe Holder Inserts are soft rubber inserts, used to protect the probes from scratches, when stored
on the VIVID E9. You can place the Probe Holder Inserts in any of the desired places on the edge of
the Operator Panel. The following types of inserts are available:
•
Probe Holder Insert STD
•
Probe Holder Insert 3D
•
Probe Holder Softinsert Doppler
8-6-34-2
Manpower
One person, 1 minute.
8-6-34-3
Tools
No tools needed.
8-6-34-4
Preparations
1.) Disconnect and remove all the probes.
2.) Store the probes in a safe place.
8-6-34-5
Removal
To remove a Probe Holder Insert, grab it and pull or push it upwards.
8-6-34-6
Installation
To install a Probe Holder Insert, place it so it fits one of the probe positions on the side of the Operator
Panel.
Figure 8-175 Probe holders (used on both sides)
PROBEHOLDER, DOPPLER
PROBEHOLDER, 3D/4D
PROBEHOLDER, STANDARD