background image

GE H

EALTHCARE

D

IRECTION

 GA091568, R

EVISION

 5

  

 VIVID E9 S

ERVICE

 M

ANUAL

 

8 - 96

Section 8-6 - Top Console Parts Replacement

8-6-6

LCD Mount Lock replacement

8-6-6-1

Preparations

When preparing for the replacement, you must perform the following steps:

1.) Power down the VIVID E9.
2.) Disconnect the Mains Power Cable from the wall outlet.
3.) Disconnect all probes and external I/O cabling.
4.) Remove the Bulkhead Cover.

WARNING

WARNING

ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9.
WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO 
VOLTAGE GREATER THAN 30 VOLTS:

1. FOLLOW LOCK OUT/TAG OUT PROCEDURES.
2. TURN OFF THE BREAKER. 
3. UNPLUG THE VIVID E9.
4. MAINTAIN CONTROL OF THE POWER PLUG.
5. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE, AS THERE ARE NO TEST POINTS TO 

VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON WILL TURN OFF.

BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER 
IS TURNED OFF WHEN THE CORD IS STILL PLUGGED INTO THE AC OUTLET.

Signed

Date

TAG

 

LOCKOUT

Summary of Contents for Vivid E9

Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...

Page 2: ......

Page 3: ...QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES QUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS T CONSULT ET COMPRIS LE NON RESPECT DE CET AVERTISSE...

Page 4: ...TE DEVIDO A CHOQUES EL TRICOS MEC NICOS OU OUTROS ESTE MANUAL DE ASSIST NCIA EST DISPON VEL APENAS EM INGL S SE QUALQUER OUTRO SERVI O DE ASSIST NCIA T CNICA QUE N O A GEHC SOLICITAR ESTES MANUAIS NOU...

Page 5: ...TERIAALIA TARVITTAVAN K NN KSEN HANKKIMINEN ON ASIAKKAAN VASTUULLA L YRIT KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT JA YMM RT NYT T M N HUOLTO OHJEEN MIK LI T T VAROITUSTA EI NOUDATETA SEUR...

Page 6: ...HO PERSON LU NEBO PACIENT VLIVEM ELEKTRICK HOP PROUDU RESPEKTIVE VLIVEM K RIZIKU MECHANICK HO PO KOZEN NEBO JIN MU RIZIKU DENNE SERVICEMANUAL FINDES KUN P ENGELSK HVIS EN KUNDES TEKNIKER HAR BRUG FOR...

Page 7: ...SU EISTAS D L ELEKTROS SM GIO MECHANINI AR KIT PAVOJ DENNE SERVICEH NDBOKEN FINNES BARE P ENGELSK HVIS KUNDENS SERVICELEVERAND R TRENGER ET ANNET SPR K ER DET KUNDENS ANSVAR S RGE FOR OVERSETTELSE IK...

Page 8: ...AR PUTEA DUCE LA R NIREA DEPANATORULUI OPERATORULUI SAU PACIENTULUI N URMA PERICOLELOR DE ELECTROCUTARE MECANICE SAU DE ALT NATUR OVAJ PRIRU NIK ZA SERVISIRANJE DOSTUPAN JE SAMO NA ENGLESKOM JEZIKU A...

Page 9: ...PROUZRO ENE STRUJNIM UDAROM MEHANI KIM I DRUGIM OPASNOSTIMA T TO SERVISN PR RU KA JE K DISPOZ CII LEN V ANGLI TINE AK Z KAZN KOV POSKYTOVATE SLU IEB VY ADUJE IN JAZYK AKO ANGLI TINU POSKYTNUTIE PREKLA...

Page 10: ...N NG L ZCE DI INDAK B R D LDE OLMASINI STERSE KILAVUZU TERC ME ETT RMEK M TER N N SORUMLULU UNDADIR SERV S KILAVUZUNU OKUYUP ANLAMADAN EK PMANLARA M DAHALE ETMEY N Z BU UYARININ G Z ARDI ED LMES ELEKT...

Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...

Page 12: ...All of GE s electrical work on these products will comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qualified personnel i e GE s field...

Page 13: ...any form in whole or in part without prior written permission of GE Healthcare GE Healthcare may revise this publication from time to time without written notice TRADEMARKS All products and their name...

Page 14: ...0 Updated per BT10 release 3 2011 MAR 01 Updated per BT 11 M4 release of product 4 2011 SEP 19 Updated per BT 12 release 5 2012 DEC 10 Updated with changes since Rev 4 Pages Revision Pages Revision Pa...

Page 15: ...Standard hazard icons 1 7 Product icons 1 8 Safety considerations 1 11 Introduction 1 11 Human safety 1 11 Mechanical safety 1 13 Electrical safety 1 15 Labels locations 1 16 Label on Front of LCD Mo...

Page 16: ...warnings 1 32 Lockout Tagout LOTO requirements 1 33 Returning Shipping Probes and Repair Parts 1 33 Electromagnetic compatibility EMC 1 34 What is EMC 1 34 Compliance 1 34 Electrostatic discharge ESD...

Page 17: ...2 Console environmental requirements 2 2 Electrical requirements 2 4 EMI limitations 2 7 Probes environmental requirements 2 8 Time and manpower requirements 2 8 Facility needs 2 9 Purchaser responsi...

Page 18: ...rom the Carton Box 3 12 Packing materials for the Wooden Box recycling information 3 16 Packing materials for the Carton Box recycling information 3 17 Preparing for setup 3 19 Verify Customer Order 3...

Page 19: ...41 Introduction 3 41 Compatibility 3 41 Select TCP IP Screen 3 42 Changing the AE Title and or Port Number Port No 3 43 DHCP setup 3 44 Set the Remote Archive s Network Information 3 47 Save the New S...

Page 20: ...to VIVID E9 as ADM 4 15 Moving and Transporting the VIVID E9 4 17 Recording important settings and parameters 4 19 Cleaning the Trackball from the outside OP 5 4 31 Cleaning the Trackball OP 1 to OP 4...

Page 21: ...this section 5 9 Introduction 5 9 VIVID E9 general description 5 10 VIVID E9 block diagram 5 10 Signal flow overview 5 11 System configuration and software 5 12 The electronics 5 12 VIVID E9 intercon...

Page 22: ...tion 5 55 BEP6 Face Top and Rear connections 5 56 BEP5 Face and Top connections 5 58 Input DC Voltages 5 63 Input Signals 5 64 Bi directional signals 5 65 Outputs 5 65 LEDs 5 66 BEP Power Supply 5 69...

Page 23: ...ct manuals 5 99 User documentation 5 99 Service documentation 5 99 Common Service Desktop overview 5 100 Purpose of this section 5 100 Introduction 5 100 iLinq Interactive Platform Features 5 100 Glob...

Page 24: ...5 Cautions and Warnings 6 5 Accessing the Main LCD and Touch Screen Backlight Adjustments 6 5 Backlight and Blue Tint Adjustment 6 6 Adjust LCD Brightness and Contrast 6 6 Test Images 6 7 Touch Scree...

Page 25: ...nt End Boards 7 5 Back End Processor 7 6 Operator Panel 7 7 Probes 7 7 Software 7 7 Network Connectivity Troubleshooting 7 8 First Status 7 8 Cannot connect to anything via the network 7 8 No Verify 7...

Page 26: ...17 XY Brake Motors Troubleshooting 7 17 Z Movement fails 7 17 Difficult to lock and release the alphanumeric keyboard 7 18 USB Footswitch 7 19 USB Device s stopped working 7 19 System Temperature Too...

Page 27: ...IVID E9 at delivery or after an upgrade 8 9 Data Management moving all images 8 10 Backing up the Patient Archive and System Configurations 8 10 Recording important settings and parameters 8 11 When t...

Page 28: ...ement 8 116 Upper Bezel replacement 8 118 Frame UI Upper replacement 8 119 Options Holder Left or Right Support replacement 8 121 Knobs for Encoders and Slidepots replacement 8 122 Replacing the Opera...

Page 29: ...sembly replacement 8 198 Replacing Fan Screen and Fan Screen Frame 8 200 Main Cable replacement 8 202 Subwoofer replacement 8 205 Casters and Brakes replacement 8 208 Rear Casters replacement 8 208 Fr...

Page 30: ...Cable DVR to BEP6 MBD replacement 8 287 DVI Flex Main Yggdrasil replacement 8 288 DVR Audio cable replacement 8 289 BEP6 to GFI and Card Rack Backplane Cable replacement 8 290 Main Power Supply repla...

Page 31: ...x 9 13 Covers and Bumpers 9 15 Top Console parts 9 18 LCD Monitor parts 9 18 Operator Panel OP parts 9 19 Top Console PCBs 9 30 Trackball parts 9 33 Operator Panel Cable Kit for Vivid E9 and Logiq E9...

Page 32: ...Top Console Cables 9 74 XYZ Controller cables 9 76 Main Power Supply cables 9 77 Front End Processor FEP cables 9 78 Back End Processor BEP cables 9 79 Peripherals Cables 9 89 ECG cables VIVID E9 9 91...

Page 33: ...reliminary checks 10 6 Functional checks 10 7 Input Power 10 8 Physical inspection 10 9 Cleaning 10 10 Probe maintenance 10 14 Electrical Safety Tests 10 20 Safety Test Overview 10 21 Leakage Current...

Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...

Page 35: ...ust read and understand all the information presented here before installing or servicing a unit 1 1 2 Contents in this chapter 1 1 Overview 1 1 1 2 Service manual overview 1 2 1 3 Important conventio...

Page 36: ...ntents in this service manual CHAPTER NUMBER CHAPTER TITLE DESCRIPTION 1 Introduction Contains a content summary and warnings 2 Site preparations Contains pre installation requirements for the VIVID E...

Page 37: ...v112 1 x v112 1 x GA200890 BEP5 w 4D Nvidia v104 3 x v112 0 x or higher v112 1 x GA000950 Vivid E9 100 230V 4D Expert Option 19 LCD GB200001 BEP6 w 4D v104 3 4 or higher v112 0 7 or higher v112 1 x G...

Page 38: ...00 BEP5 w 4D Nvidia v104 3 2 v110 1 x GA200890 or GA200800 BEP5 w 4D Nvidia v104 2 x v104 1 x v110 0 x GA000830 VIVID E9 100 230V 2D 17 LCD GA200804 VE9 Card Rack Complete w MLA4 GB200002 BEP6 wo 4D v...

Page 39: ...full spectrum of operating frequencies Signal flow from the Probe Connector Panel to the Front End and then over to the Back End Processor and finally to the monitor and peripherals System configurat...

Page 40: ...personal are labeled in one of three ways DANGER WARNING CAUTION When a hazard is present that can cause property damage but has absolutely no personal injury risk a NOTICE is used Example Disk drive...

Page 41: ...make you aware of specific types of hazards that could possibly cause harm Even if a symbol isn t used in this manual it may be included for your reference Some others icons make you aware of specific...

Page 42: ...the box symbol IEC 878 02 05 indicates equipment having a floating applied part having a degree of protection suitable for direct cardiac contact Probe connectors and ECG connector On newer systems a...

Page 43: ...for a Group 1 Class B Medical Device as stated in EN 60601 1 2 IEC 60601 1 2 REAR OF CONSOLE CAUTION The equilateral triangle is usually used in combination with other symbols to advise or warn the us...

Page 44: ...earth grounding terminal USED SEVERAL PLACES INSIDE THE SCANNER Equipotential Indicates the terminal to be used for connecting equipotential conductors when interconnecting grounding with other equip...

Page 45: ...ROUS VOLTAGES CAPABLE OF CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING WARNING WARNING IF THE COVERS ARE REMOVED FROM AN OPERATING VIVID E9 SOME M...

Page 46: ...G DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A DEFINITE SAFETY HAZARD WARNING WARNING DO NOT SUBSTITUTE PARTS OR M...

Page 47: ...G NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR EVEN IF IT LOOKS OK IT MAY BE DAMAGED WARNING WARNING THE SYSTEM SHOULD NOT BE MOVED WITH THE OPERATING PANEL EXTENDED POSITION THE OPERATING PANEL IN...

Page 48: ...UNCONTROLLED MOTION AND COSTLY DAMAGE ALWAYS BE SURE THE PATHWAY IS CLEAR USE SLOW CAREFUL MOTIONS USE TWO PEOPLE WHEN MOVING ON INCLINES OR LIFTING MORE THAN 16 KG 35 LBS CAUTION DO NOT TRANSPORT VIV...

Page 49: ...e shared with other types of equipment Both the system power cable and the power connector must meet international electrical standards 1 4 4 2 Probes Follow these guidelines before connecting a probe...

Page 50: ...Labels locations Section 1 5 Labels locations 1 5 1 Label on Front of LCD Monitor 1 5 2 Label on Rear of LCD Monitor Table 1 6 Label on Front of LCD Monitor DESCRIPTION ILLUSTRATION LABEL GE LOGO Tabl...

Page 51: ...AMEPLATE GOLD UPPER OP PANEL Used on VIVID E9 with 4D Expert option VIVID E9 NAMEPLATE BRONZE UPPER OP PANEL Used on VIVID E9 with 2D Vivid E9 Nameplate Periwinkle Upper OP Panel Used on VIVID E9 Pro...

Page 52: ...Labels locations 1 5 6 Labels near Connectors on Front Table 1 10 Labels near Connectors on Front DESCRIPTION ILLUSTRATION Label Front Panel Label Patient IO DEFIBRILLATOR PROOF TYPE CF EQUIPMENT Labe...

Page 53: ...on DVD Units 1 5 8 Label on External I O Table 1 11 Labels on DVD units DESCRIPTION ILLUSTRATION Label DVD Label DVR Digital Video Recorder Table 1 12 Labels on External I O DESCRIPTION ILLUSTRATION L...

Page 54: ...ction 1 5 Labels locations 1 5 9 Labels at AC Mains Inlet and Circuit Breaker Table 1 13 Labels on AC Power Supply Main Power Supply DESCRIPTION ILLUSTRATION Label AC Controller Label System Part Numb...

Page 55: ...5 10 1 Label General Info BT 12 Table 1 14 Label General Info BT 12 DESCRIPTION ILLUSTRATION Label General Info BT 12 International Introduced JUN 2012 Label General Info BT 12 International Label Ge...

Page 56: ...ral Info BT 11 BT 10 Table 1 15 Label General Info BT 11 BT 10 sheet 1 of 3 DESCRIPTION ILLUSTRATION Label General Info BT 11 International Introduced SEP 2010 Label General Info BT 11 International I...

Page 57: ...General Info BT 11 and BT 12 Importer GE Medical Systems Egypt Label General Info BT 11 Korean Used after June 1 2012 Label General Info BT 11 Korean Used before June 1 2012 Label General Info BT 11 a...

Page 58: ...RE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 1 24 Section 1 5 Labels locations Label General Info BT 10 Chinese Table 1 15 Label General Info BT 11 BT 10 cont d sheet 3 of 3 DESCRIPTION IL...

Page 59: ...IVID E9 SERVICE MANUAL Chapter 1 Introduction 1 25 1 5 10 3 Label General Info BT 09 Table 1 16 Label on Rear Cover DESCRIPTION ILLUSTRATION Label General Info BT 09 International Label General Info B...

Page 60: ...angerous voltage the lightning flash with arrowhead in equilateral triangle is used to indicate electric shock hazards ATTENTION Consult accompanying documents is intended to alert the user to refer t...

Page 61: ...rted municipal waste and must be collected separately Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment This unit carries...

Page 62: ...ASS I The VIVID E9 ultrasound unit is a Class I device type CF according to Sub clause 14 of IEC60601 1 1988 TYPE CF Equipment Type CF heart in the box symbol IEC 878 02 05 indicates equipment having...

Page 63: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 1 Introduction 1 29 1 5 12 Label on the BEP6 s door Figure 1 2 Label on the BEP6 s door...

Page 64: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 1 30 Section 1 5 Labels locations 1 5 13 Label on the BEP5 s door Figure 1 3 Label on the BEP5 s door...

Page 65: ...ON GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 1 Introduction 1 31 1 5 14 Label Disassembly Nester This label is located on the outside of the Front End Rack s cover inside unit Figure 1 4 Lab...

Page 66: ...HEN HANDLING TESTING AND ADJUSTING WARNING WARNING IF THE COVERS ARE REMOVED FROM AN OPERATING VIVID E9 SOME METAL SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF TOUCHED EVEN WHILE IN...

Page 67: ...ld a part or equipment with visible body fluids be taken or shipped from a clinic or site for example body coils or an ultrasound probe The purpose of the regulation is to protect employees in the tra...

Page 68: ...E9 conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge radiated and conducted RF fields magnetic fields and power line transient requirements...

Page 69: ...speed capacity type 8 Save secondary image capture cine loop 4D multi volume loop NOTE Restart the application before resuming clinical scanning Table 1 18 Phone numbers for Customer Assistance LOCATI...

Page 70: ...SERVICE MANUAL 1 36 Section 1 10 Customer assistance 1 10 2 System manufacturer Table 1 19 System manufacturer MANUFACTURER PHONE NUMBER FAX NUMBER GE VINGMED ULTRASOUND A S STRANDPROMENADEN 45 P O BO...

Page 71: ...erview 2 1 1 Purpose of this chapter This chapter provides the information required to plan and prepare for the setup of a VIVID E9 Included are descriptions of the facility and electrical needs to be...

Page 72: ...o 140 F Humidity Limits Operation 30 60 rH non condensing Storage and Transport 30 95 rH non condensing Air Pressure Limits Operation 700 1060 hPa Storage and Transport 700 1060 hPa Heat Dissipation 3...

Page 73: ...e room Each person in the room places an additional 300 BTU h demand on the cooling system 2 2 1 4 Lighting Bright light is needed for system installation updates and repairs However operator and pati...

Page 74: ...then a center tapped power source is required Sites with a mains power system with defined Neutral and Live The dedicated line shall consist of one phase a neutral not shared with any other circuit a...

Page 75: ...2 5 Unit power plug If the unit arrives without a power plug or with the wrong plug you must contact your GE dealer or the installation engineer must supply what is locally required Table 2 2 Electric...

Page 76: ...ral console requirements 2 2 2 6 Power stability requirements Voltage drop out Max 10 ms Power transients all applications Less than 25 of nominal peak voltage for less than 1 millisecond for any type...

Page 77: ...nce problems caused by high frequency high powered radio or video broadcast signals Ground the unit Poor grounding is the most likely reason a unit will have noisy images Check grounding of the power...

Page 78: ...C 50 to 104 F 5 to 42 7 C 41 to 108 9 F Storage 20 to 50 C 4 to 122 F 20 to 60 C 4 to 140 F Temperatures in degrees Celsius o C conversion to degrees F o F o C 9 5 32 CAUTION PAMPTE probes are designe...

Page 79: ...performed by qualified personnel The products involved and the accompanying electrical installations are highly sophisticated and special engineering competence is required All electrical work on thes...

Page 80: ...fined Neutral The dedicated line shall consist of one phase two lines not shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet Dedicated singl...

Page 81: ...inet ordered by GE for its software and proprietary manuals 2 3 4 Minimal floor plan suggestion 2 3 5 Suggested Floor Plan VIVID E9 and EchoPAC PC in Same Room Figure 2 1 Minimal Floor Plan 2 5 m x 3...

Page 82: ...es in this manner frees up the on board monitor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier and at a lower...

Page 83: ...2 3 Worksheet for DICOM Network Information VIVID E9 Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP Addresses Defaul...

Page 84: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 2 14 Section 2 3 Facility needs This page left blank to facilitate double sided printing...

Page 85: ...he actual installation and how to check and test the unit probes and external peripherals for electrical safety are included in this procedure 3 1 2 Contents in this chapter 3 1 Overview 3 1 3 2 Setup...

Page 86: ...THE UNIT CAUTION IF THE UNIT IS VERY COLD OR HOT DO NOT TURN ON ITS POWER UNTIL IT HAS HAD A CHANCE TO ACCLIMATE TO ITS OPERATING ENVIRONMENT Table 3 2 VIVID E9 Acclimate Time o C 40 35 30 25 20 15 1...

Page 87: ...ER MANUAL S SHOULD BE FULLY READ AND UNDERSTOOD BEFORE OPERATING THE VIVID E9 AND KEPT NEAR THE UNIT FOR QUICK REFERENCE CAUTION ACOUSTIC OUTPUT HAZARD ALTHOUGH THE ULTRASOUND ENERGY TRANSMITTED FROM...

Page 88: ...if the handling during transportation has been improper a set of Tilt Shock indicators have been attached to the transportation box 3 3 3 2 Position of the Tilt and Shock indicators The Tilt Shock in...

Page 89: ...the Tilt Indicator red colored inside the middle of the indicator If yes The Tilt Indicator has been activated Continue with the instructions in subsection 3 3 4 2 Damage in transportation before you...

Page 90: ...he carrier Whether noted or concealed damage MUST be reported to the carrier immediately upon discovery or in any event within 14 days after receipt and the contents and containers held for inspection...

Page 91: ...lt Indicator has triggered or is missing STEP TASK 1 If the Tilt Indicator is missing Note on the shipping papers at the time of receipt that the Tilt Indicator label is missing If the Tilt Indicator...

Page 92: ...100 230 VAC 50 60 Hz 1100 W 1 Set G G aG G G GaG G G G G G SG G G G G G G G G G aG W TYW GE Vingmed Ultrasound AS O P G G GZ ZG wvihG G G SG_ Vivid E9 GaG Vivid E9 116x78x133 mm LxWxH 128 kg 200 kg M...

Page 93: ...E9 Transportation Box Label printed on the Carton Box ICONS and SYMBOLS DESCRIPTION ICONS and SYMBOLS DESCRIPTION ICONS and SYMBOLS DESCRIPTION KEEP AIR PRESSURE BETWEEN 70 kPa and 106 kPa DO NOT STAC...

Page 94: ...On the first version of the Transportation Box only the front door is meant to be used as a ramp It has bevel cut in one end 2 Place the front door as a ramp against the rear edge of the pallet 3 Care...

Page 95: ...the wheels are swiveling when the system is inside the transportation box it may jam the system inside the box The system must be moved backwards both into and out of the transportation box Moving the...

Page 96: ...2 Setup reminders 3 3 6 Unpacking VIVID E9 from the Carton Box Table 3 8 Uncrating the VIVID E9 sheet 1 of 4 Step Task Illustration 1 Cut the straps around the crate 2 Remove the Top Cover 3 Remove t...

Page 97: ...SERVICE MANUAL Chapter 3 System setup 3 13 5 Remove the two Frames sides and the two Exit Ramp Bases 6 Install the two Exit Ramp Base on the Complete Exit Ramp the rear plate Table 3 8 Uncrating the...

Page 98: ...ANUAL 3 14 Section 3 2 Setup reminders 7 Remove the Inlay UI Top 8 Remove the Support For Monitor 9 Remove the Complete Front Protection 10 Remove the clear plastic bag from the VIVID E9 Table 3 8 Unc...

Page 99: ...rection stability when the system is rolled down the ramp from the pallet 1 Direction Dir Lock 2 Release Dir Lock and Front Brakes 3 Front Brakes 4 Rear Brakes 13 Unlock the Rear Brakes 14 Carefully m...

Page 100: ...recycling information The packing materials for VIVID E9 are recyclable The Transportation Box is made of spruce or similar material PHYTOSANITARY CERTIFICATE included in all shipments to The People s...

Page 101: ...fer to the table below Table 3 9 Packaging parts for VIVID E9 sheet 1 of 2 Item Description Qty Material Illustration 1 Export pallet 1200 x 800 1 D 2 Complete base 1 A 3 Complete column left and righ...

Page 102: ...ction 3 5 Packing materials for the Carton Box recycling information Material type A B C D BB34bc with varnish C9068 BB27c with varnish C9068 PE foam Stratocell Wood ISPM15 Table 3 9 Packaging parts f...

Page 103: ...nspection If the system has been damaged please refer to Damage in Transportation on page x in the beginning of this manual 3 6 3 EMI protection This unit has been designed to minimize the effects of...

Page 104: ...n site conditions 3 7 2 2 Physical dimensions The physical dimensions of the VIVID E9 unit are summarized in Table 3 10 3 7 2 3 Mass with monitor and peripherals 3 7 2 4 Acoustic noise level Less than...

Page 105: ...VEL WILL MOST LIKELY DESTROY THE UNIT Figure 3 4 Mains Voltage Rating label Table 3 12 Electrical specifications for VIVID E9 PART NUMBER DESCRIPTION VOLTAGE TOLERANCES POWER CONSUMPTION FREQUENCY BT...

Page 106: ...the requirements of the valid version of IEC60601 1 1 If in doubt consult the technical service department or your local representative for GE Healthcare 3 7 4 1 Connect Ethernet Connect the Ethernet...

Page 107: ...r See Section 9 20 Physio TX Parts on page 9 92 Connect the Heart Microphone via the corresponding Phono Adapter to the Phono input on the Patient I O panel 3 7 5 3 Connect Pulse Pressure Transducer C...

Page 108: ...ort three PDT probe ports Figure 3 9 and a Doppler probe port The PD probe port is compatible with the Vivid 7 probe connectors The three PDT probe ports are specific to the VIVID E9 probe connectors...

Page 109: ...to place 5 Turn the locking handle clockwise to the full vertical position to lock in place 6 Position the probe cable so that it is not resting on the floor 3 7 6 3 Disconnect a probe 1 Rotate the lo...

Page 110: ...For a detailed procedure see 4 2 2 Power shut down on page 4 8 DANGER ALWAYS CONNECT THE UNIT TO A FIXED POWER SOCKET WHICH HAS THE PROTECTIVE GROUNDING CONNECTOR DANGER NEVER USE A THREE TO TWO PRONG...

Page 111: ...Location Figure 3 10 Hospital and department name Table 3 13 Location name STEP TASK EXPECTED RESULT S 1 Select the Hospital field see Figure 3 10 and type the name of the hospital max 64 characters...

Page 112: ...MM DD YYYY format is used 3 Select the preferred Time Format see c in Figure 3 11 24 the 24 hour format is used 12 the 12 AM PM hour format is used 4 Adjust the date see a in Figure 3 11 New date is...

Page 113: ...guage Adjustments STEP TASK EXPECTED RESULT S 1 Open the Configuration Window see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 Select System if needed The System Settings window is displayed 2 Use...

Page 114: ...Online Manual Language Selection STEP TASK EXPECTED RESULT S 1 Open the Configuration Window see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 Select System if needed The System Settings window is...

Page 115: ...EXPECTED RESULT S 1 Open the Configuration Window see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 Select System if needed The System Settings window is displayed 2 Use the drop down dialog to se...

Page 116: ...Format to be used by DVR Select DVR Adjust LCD s Contrast and Backlight Intensity Alphanumeric Keyboard Setups Printer Setup Monitor Selection 3 8 3 2 Open Service Screen 1 Press Config F2 and log on...

Page 117: ...ic Keyboard configuration NOTE This Procedure is not required if the alphanumeric keyboard is a US keyboard since the default setting is set to US English keyboards 1 Select Keyboard Setup to get acce...

Page 118: ...r installation procedure for correct printer installation See 3 8 4 Optional Peripherals Peripheral Connection on page 3 36 for more information 1 Select Add Printer to start the Add Printer Installat...

Page 119: ...NOTE Installation instructions for the Digital Video Recorder DVR option is included in the DVR Installation manual Direction Number GA294614 Selecting Offline mode Follow these steps to select Offlin...

Page 120: ...ochrome Printer P95DE Digital Video Recorder DVR 3 8 4 2 External Peripherals Optional for Connection to USB One of the external units listed below may be connected to the USB port on the rear of the...

Page 121: ...nected via Ethernet TCP IP Network please refer to the respective printer s documentation The VIVID E9 supports the following network printers HP OfficeJet Pro 8000 HP OfficeJet PRO K5400dn HP OfficeJ...

Page 122: ...assword Software Option String enables a software option or a combination of software options This password is specific for each VIVID E9 3 8 5 2 Installing a Software Option 1 Press Config F2 and log...

Page 123: ...the Password Software Option Key Alphanumeric string 6 Press Save to save the new setting 7 Restart to save and activate the settings and adjustments you have done so far Figure 3 23 Type Software Opt...

Page 124: ...nd of the crossed network cable to the network connector on the VIVID E9 b Connect the other end to the network connector to the EchoPAC PC Workstation see chapter 3 in the EchoPAC PC Workstation Serv...

Page 125: ...t delivery settings 3 10 1 Introduction To be able to use the network functions when connected to a hospital network the scanner must have a proper network address Before you can set up the scanner yo...

Page 126: ...COMPUTER NAME AE TITLE FOR VIVID E9 THIS NAME IS ON THE FORM VIVIDX 00NNNN WHERE 00NNNN IS A NUMBER NNNN IS THE SCANNER S SERIAL NUMBER PORT NO DEFAULT PORT NUMBER 104 REMOTE ARCHIVE SETUP AREA IP SE...

Page 127: ...ange AE Title and or Port No edit the respective fields 2 Select Save settings to store your changes This will bring up a new Warning screen 3 Select Ok to save your changes or Cancel to return withou...

Page 128: ...use of DHCP 1 When in the TCP IP screen select Network Settings to display the Network Connections screen see Figure 3 28 Network Connections on page 3 44 2 Right click Local Area Connection and sele...

Page 129: ...L Chapter 3 System setup 3 45 3 10 5 DHCP setup cont d 3 Select Properties to display the Local Area Connection Properties 4 Select Internet Protocol TCP IP then select Properties Figure 3 29 Local Ar...

Page 130: ...the IP address Subnet mask Default gateway To save your new settings and close the open windows 1 Select OK to save the setting and close the Internet Protocol TCP IP Properties dialog 2 Select OK to...

Page 131: ...or server you must configure VIVID E9 to communicate with it In the Remote Archive Setup area of the Tcpip screen see example in Figure 3 32 enter the 1 Remote Archive IP address Default IP Address f...

Page 132: ...ow 2 Select OK to save the new settings By selecting Cancel the previous settings will be kept 3 The new settings are saved to a common settings file After a restart the settings are also included in...

Page 133: ...ters long so it must be placed within reach of this cable 3 11 2 1 Preparations for installing the CP30DW for the first time Set Switch 1 and 2 on the rear side of the printer to ON Please refer to th...

Page 134: ...on User s Manuals Installation Manuals etc for the peripheral units must be kept as part of the original system documentation This will ensure that all relevant safety and user informations are availa...

Page 135: ...oduct Locator Installation Card NOTE The Product Locator Installation Card shown may not be the same as the provided Product Locator card From the factory a sheet with five Product Locator cards for t...

Page 136: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 3 52 Section 3 12 Setup paperwork This page left blank to facilitate double sided printing...

Page 137: ...and Functional checks Section 4 1 Overview 4 1 1 Purpose of this chapter This chapter provides procedures for quickly checking major functions of the VIVID E9 scanner and diagnostics instructions usin...

Page 138: ...red An empty blank DVD RW disc ECG Pads ECG Harness CABLE ECG MARQ AHA AMERICA P N 164L0025 LEADWIRES ECG MARQ AHA AMERICA P N 164L0027 or CABLE ECG MARQ IEC EU AS P N 164L0026 LEADWIRES ECG MARQ IEC...

Page 139: ...nd cooling purposes CAUTION ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 TURN OFF THE BREAKER 2 UNPLUG TH...

Page 140: ...FIXED POWER SOCKET WHICH HAS THE PROTECTIVE GROUNDING CONNECTOR DANGER NEVER USE A THREE TO TWO PRONG ADAPTER THIS DEFEATS THE SAFETY GROUND DANGER ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND...

Page 141: ...unit can be moved slightly 3 Verify that the power cable is without any visible scratches or any sign of damage 4 Verify that the on site mains voltage is within the limits indicated on the rating lab...

Page 142: ...er a few seconds listen to the fan sound b Power is distributed to the peripherals Operator Panel Console Monitor Front End Processor and Back End Processor c Back End Processor and rest of scanner st...

Page 143: ...s been loaded either a 2D screen is displayed on the screen indicating that a probe has been connected or a No Mode screen is displayed indicating that no probe has been connected NOTE Total time used...

Page 144: ...rrent user The system remains ON and ready for a new user to log on If the Logoff button is dimmed it indicates that no user is logged on to the unit at the moment Shutdown Use this button to shut dow...

Page 145: ...e Top Console will be released so you can move the Top Console in the XY directions 2 Press once on the On Off key on the Operator Panel to display the System Exit menu 3 Select Shutdown to do a compl...

Page 146: ...ains Power Circuit Breaker located on the rear of the unit This will cut power distribution within the unit NOTICE Be sure to wait with the next step until the system has finished its shut down Failin...

Page 147: ...Console out of the locked position until the Top console is in the desired position 2 Release the button s 4 2 3 3 To lock the Top Console 1 Press and hold down the Lock and brake release button s Fi...

Page 148: ...e the console s XY mechanism frog leg The release point is located on the rear of the XY Mechanism Frog Leg 1 Insert a small tipped screwdriver or eq into the release point and push until a click is h...

Page 149: ...ed or has failed or if the mains power is turned off or disconnected or if it is a mains power failure Move the handle a few millimeters to the right side to release the Z Mechanism disengage the moto...

Page 150: ...tment 4 2 4 1 To unlock the LCD monitor Turn the release knob counter clockwise to unlock the LCD monitor The LCD monitor can be moved freely in all directions 4 2 4 2 To lock the LCD monitor Turn the...

Page 151: ...select Config choose either Procedure A or Procedure B A Procedure A On the Touch Screen 1 Select the Utility tab 2 Then select Config B Procedure B On the alphanumeric keyboard Press F2 Both method A...

Page 152: ...onnectivity settings As default the password for ADM is ulsadm Select the name ADM the password ulsadm and select Login It is possible for the administrator ADM to establish new users and set unique p...

Page 153: ...edal to release the Swivel lock 4 2 6 2 To prepare the unit to be moved 1 If not locked move the keyboard console and LCD monitor to the park position see 4 2 3 Top Console position adjustment on page...

Page 154: ...ibed below 1 If not locked move the keyboard console and LCD monitor to the park position see 4 2 3 Top Console position adjustment on page 4 11 and 4 2 4 LCD Monitor position adjustment on page 4 14...

Page 155: ...2 Tcpip see 4 2 7 6 Connectivity Recording the TCP IP settings on page 4 23 System Settings see 4 2 7 7 System Settings on page 4 25 About System Version see 4 2 7 9 About System Version on page 4 26...

Page 156: ...dures 4 2 7 2 Connectivity Dataflow 1 Select Config F2 to log on as ADM Default password ulsadm For a detailed description see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 2 Select Connectivity Da...

Page 157: ...lsadm For a detailed description see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 2 Select Connectivity Additional Outputs 3 Select Button P1 in the Button pull down menu 4 Record the settings for...

Page 158: ...ption see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 2 Select Connectivity Tools 3 If in use Record the Remote Path 4 If in use Record the Configurable Remote Path User Also ensure that you have...

Page 159: ...onnectivity Tcpip 3 Record all settings in Table 4 1 Record settings from Tcpip screen and sub screens on page 4 24 4 Select Network Settings 5 Double click Local Area Connection 6 Select Properties 7...

Page 160: ...ivity Recording the TCP IP settings cont d Table 4 1 Record settings from Tcpip screen and sub screens PARAMETER VALUE COMPUTER NAME AE TITLE PORT NO REMOTE ARCHIVE SETUP REMOTE ARCHIVE IP ADDR REMOTE...

Page 161: ...see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 2 Select System Settings 3 Record the following information for Location Hospital Department 4 Record the following information for Date and Time T...

Page 162: ...n 3 Record software and hardware versions in Table 4 2 Record Software versions on page 4 26 4 2 7 10 About Firmware Version There are no data on this screen that need to be recorded 4 2 7 11 About HW...

Page 163: ...d ulsadm For a detailed description see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 2 Select Admin Disk Management 3 Record the selected Reminder Interval Every setting 4 2 7 14 Admin Backup 1 Se...

Page 164: ...m For a detailed description see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 2 Select Admin Restore 3 Record Remote Path 4 2 7 16 Admin Users 1 Select Config F2 to log on as ADM Default password...

Page 165: ...Admin System Admin 3 Record the SW Option Key one or more alphanumeric strings from the SW Option Key field Figure 4 28 System Admin Example WARNING WARNING MAKE SURE THAT THE SOFTWARE OPTION KEY ALP...

Page 166: ...rd ulsadm For a detailed description see 4 2 5 Logging on to VIVID E9 as ADM on page 4 15 2 Select Service 3 Record Video settings Format in Table 4 4 Service screen settings on page 4 30 4 Record Mon...

Page 167: ...d trackball from the keyboard 5 Wipe off any oil or dust from the trackball using a cleaner or dry cloth 6 Wipe off any oil or dust from the trackball housing rollers etc using a cleaner or cotton bud...

Page 168: ...olves removing the OP and then remove and open the Trackball as described in the procedure below OP 1 An inductive trackball is used on the first version of the Operator Panel It is only needed to cle...

Page 169: ...n page 8 125 CAUTION ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 TURN OFF THE BREAKER 2 UNPLUG THE SYSTE...

Page 170: ...res 4 2 9 5 Remove the Trackball Follow these steps to remove the Trackball 1 Unplug the cable connectors from the Trackball 2 Use the Hex key to remove the two fixing screws with shims 3 Remove the T...

Page 171: ...To get access for cleaning you need to remove the ball The ball is held in position by the Dust Gasket 1 Turn the Dust Gasket counter clockwise to unlock it 2 Remove the Dust Gasket 3 Remove the ball...

Page 172: ...it so the fixing screw holes on the Fixing Ring align with the fixing screw holes on the Trackball assembly 4 2 9 7 Install the Trackball Follow these steps to install the Trackball 1 Install the Trac...

Page 173: ...on Turn on power to VIVID E9 4 3 2 2D Mode B mode Checks 4 3 2 1 Introduction The 2D Mode is the system s default mode 4 3 2 2 Preparations 1 Connect one of the probes See Connecting Probes on page 3...

Page 174: ...ve to lower frame rate to increase resolution in image Use Frequency move to lower frequency to increase penetration Use the Reject control to reduce noise in the image Use the DDP control to optimize...

Page 175: ...are used to optimize the M mode picture Verify that all the listed controls are working as intended Adjust Horizontal sweep to optimize the display resolution Adjust Gain and TGC controls to adjust t...

Page 176: ...ss trackball to be able to select between Position and Size controls 4 Use the trackball to adjust the dimension of the ROI 4 3 4 4 Adjust the Color 2D Mode controls Adjust the Active mode gain to set...

Page 177: ...gain to set the gain in the color flow area Adjust Scale to the highest setting that provides adequate flow detection NOTE The scale value may affect FPS Low Velocity Reject and Sample Volume Adjust...

Page 178: ...nted low velocity blood flow and tissue movement In PW mode adjust Sample volume to low setting for better resolution or higher setting to more easily locate the disturbed flows Adjust the Compress se...

Page 179: ...examined 6 Press Select The instruction Size should be highlighted in the trackball status bar NOTE If the trackball control pointer is selected press trackball to be able to select between Position...

Page 180: ...An application menu for the desired probe is listed on the screen 3 Trackball to the desired application Press Select to launch the application To change application without changing the current prob...

Page 181: ...6L0028 ECG Pads 3 pc or ECG simulator 4 3 8 3 ECG Check Table 4 6 ECG Check Step Task Expected Result s 1 Connect the ECG harness to the connector on the front of the system The unit displays a straig...

Page 182: ...is displayed default mode 4 3 9 3 Adjust the Cineloop controls 1 Press Freeze The left and right markers are displayed on either side of the last detected heart cycle on the ECG trace 2 Press 2D Freez...

Page 183: ...e internal printer is controlled from the P1 and P2 keys on the VIVID E9 s Operator Panel The factory default is P1 for the internal printer P2 for external network printer Table 4 7 Printer checks St...

Page 184: ...ect one of the Physical Print Buttons that you want to configure 8 In the right pane click on the service name you just created in the Services Tab 9 Click on the button This will place this service i...

Page 185: ...ve the busy light on the DVR unit or the busy icon on screen 2 Create a patient record or open an existing one 3 Press Record on the Operator Panel A red dot is displayed in the DVR status area on the...

Page 186: ...change title or chapter 4 To find a patient record press Go to Search on the Touch Panel The Video Counter Search window is displayed 5 Select a title from the Recording title drop down menu and pres...

Page 187: ...ks function as intended 1 Press the right pedal to engage the front wheel brakes Apply a pressure on the scanner to verify that the brakes works as intended 2 Press the center pedal to release the fro...

Page 188: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 4 52 Section 4 4 Site Log Section 4 4 Site Log Table 4 9 Site Log DATE SERVICE PERSON PROBLEM COMMENTS...

Page 189: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 4 General procedures and Functional checks 4 53 Table 4 9 Site Log cont d DATE SERVICE PERSON PROBLEM COMMENTS...

Page 190: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 4 54 Section 4 4 Site Log This page left blank to facilitate double sided printing...

Page 191: ...m and the Common Service Desktop interface 5 1 2 Contents in this chapter 5 1 Overview 5 1 5 2 InSite ExC 5 4 5 3 VIVID E9 overview 5 9 5 4 Top Console with LCD monitor and Operator Panel 5 17 5 5 Mai...

Page 192: ...4D Nvidia v104 3 x v112 0 x or higher v112 1 x GA000950 Vivid E9 100 230V 4D Expert Option 19 LCD GB200001 BEP6 w 4D v104 3 4 or higher v112 0 7 or higher v112 1 x GB200003 BEP5 w 4D Nvidia Quadro 20...

Page 193: ...90 or GA200800 BEP5 w 4D Nvidia v104 2 x v104 1 x v110 0 x GA000830 VIVID E9 100 230V 2D 17 LCD GA200804 VE9 Card Rack Complete w MLA4 GB200002 BEP6 wo 4D v104 3 x v110 1 12 v112 1 x GA200900 or GA200...

Page 194: ...te ExC link at the bottom of the display screen 5 2 2 InSite ExC Icon The InSite ExC icon in the status bar change symbol and color depending on ongoing activity Clicking on the Icon brings up the InS...

Page 195: ...Is Connected Yellow Icon InSite ExC activated and Technical Support can look around on your system see temperatures and voltage information see status get files stored with Alt D and look at logs but...

Page 196: ...he bottom of the display 2 Press the Right Trackball Set Key This opens of the RFS screen which sends a service dispatch directly to GE Service after you fill in the following information Items with a...

Page 197: ...1 DRX4_TOP DRX1 4 10 C 80 C If the temperature on any of the DRX cards rise above 80 C 176 F an ARFS is generated The lower temperature limit 10 C is outside the operating temperature for the product...

Page 198: ...vice related functions depending on whether InSite has connected or not connected There are two Non Disruptive states If you see a black and white icon InSite ExC is activated but not open for Technic...

Page 199: ...n The VIVID E9 ultrasound unit is a high performance digital ultrasound imaging system with total data management The system provides image generation in 4D 2D B Mode Color Doppler Power Doppler Angio...

Page 200: ...ed on the hard drive inside the Back End Processor BEP and all necessary software is loaded from the hard drive on power up A Physio module the Patient I O is incorporated in the Back End Processor BE...

Page 201: ...s amplifies these signals Then the signals are routed to the DRX board s where the signals are A D converted The digital signals are then further processed on the DRX boards After amplification and di...

Page 202: ...onfiguration is stored on a hard disk drive inside the Back End Processor At power up all necessary software is loaded from the hard disk drive 5 3 7 The electronics VIVID E9 internal electronics are...

Page 203: ...HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 5 Components and functions theory 5 13 5 3 8 VIVID E9 interconnection diagram Figure 5 5 VIVID E9 w BEP6 interconnection diagr...

Page 204: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 5 14 Section 5 3 VIVID E9 overview 5 3 8 VIVID E9 interconnection diagram cont d Figure 5 6 VIVID E9 w BEP5 interconnection diagram...

Page 205: ...wall thickness when accurate timing information is required M mode is also known as T M mode or time motion mode Ultrasound echoes of different intensities are mapped to different gray scale values in...

Page 206: ...odalities the other being CW Doppler In spectral Doppler blood flow is presented as a scrolling display with flow velocity on the Y axis and time on the X axis The presence of spectral broadening indi...

Page 207: ...erator Panel with an ON OFF switch a touch screen and a switch board with controls for manipulating the picture quality and for use in Measure Analyze M A an alphanumeric keyboard QWERTY keyboard on a...

Page 208: ...ide the system assists in the Z axis vertical movement The brakes used for locking the Top Console s X and Y movement are operated by electrical motors 5 4 1 4 The XY Locking mechanism In addition to...

Page 209: ...ked position on page 5 19 When the Operator Panel moves into locked position a micro switch 1 is activated The micro switch activation is used by the Motor Controller to detect when the Operator Panel...

Page 210: ...block diagram MAIN CABLE TO FROM MAIN CONSOLE MAIN CABLE DVI VIDEO 48V POWER AUDIO USB FOR OP PANEL USB FOR VIDEO ON TOUCH PANEL UP DOWN BRAKE RELEASE SWITCHES Cable List Cable Part Number Descriptio...

Page 211: ...2 Operator Panel Control Panel 5 4 2 1 Operator Panel general description The Operator Panel includes an On Off switch different controls for manipulating the picture quality and controls for use in M...

Page 212: ...onal rotaries push buttons potentiometers etc It also connects to the touch panel overlay and handles the push events from this Additionally the custom keyboard also interfaces the trackball buttons T...

Page 213: ...ERVICE MANUAL Chapter 5 Components and functions theory 5 23 Figure 5 12 Top Console with Operator Panel block diagram TO FROM MAIN CONSOLE MAIN CABLE DVI VIDEO 48V POWER AUDIO USB FOR OP PANEL USB FO...

Page 214: ...s drive is also used for software installation during a software upgrade The optional second DVD drive DVR is used to record and replay video captured by the Digital Video Stream Recorder option Power...

Page 215: ...upply Temperature sensors several places inside VIVID E9 Temperature control software on the BEP 5 6 1 2 Software control Software running on the BEP samples the temperature on each sensor If the temp...

Page 216: ...has two major functions These are direction lock parking brake The third pedal is the release pedal used to release the two other functions if activated The direction lock and brake is operating on th...

Page 217: ...ition Do not change the color keys position Don t insert a card in the wrong position in the Card Rack If the power is turned on with a card placed in the wrong position the VIVID E9 will be destroyed...

Page 218: ...C 15 VDC 5 8 1 4 Input Pulser voltages TSV1P 0 VDC to 95 VDC TSV1N 0 VDC to 95 VDC TSV2P 0 VDC to 95 VDC TSV2N 0 VDC to 95 VDC These voltages comes from the Main Power Supply 5 8 1 5 Input signals RX...

Page 219: ...e transmit trigger pulse TXTRIG_L The transmit pulses are routed via the Front Plane XD bus located on the front side of the GTX GRX and Relay boards to the Relay board where they are fed to the selec...

Page 220: ...d on the VIVID E9 model and options installed The reflected signal from body structures and blood cells are routed from the probe via the Relay Board and the Front Plane to the GRX receiver boards whe...

Page 221: ...h voltage multiplexer in linear probes In addition the GFI generates the transmit trigger pulse for GTX a receive synchronization pulse TXTRIG_L used by DRX a differential ATGC voltage used by GRX glo...

Page 222: ...systems requires 256 TX channels so four TX boards with 64 TX channels each are used Each channel from the TX boards is connected to 4 elements in the probe so we can transmit on 1024 elements 256 x 4...

Page 223: ...Transmit Pulse Generators TS Voltage 1 and TS Voltage 2 from the Main Power Supply supply the transmitters with the needed voltages to generate the correct ultrasound signals Figure 5 18 Block Diagram...

Page 224: ...BT 08 systems Figure 5 20 GTX TLP 3 0 boards location 192 TX channels BT 09 and some BT 11 systems CW PROBE CONNECTOR DC VOLTAGES TX PULSER VOLTAGES TO FROM BACK END PROCESSOR OPTIONAL VPD DRX GTX TL...

Page 225: ...DC 5 8 3 4 Input TX Pulser voltages TSV1P 0 VDC to 95 VDC TSV1N 0 VDC to 95 VDC TSV2P 0 VDC to 95 VDC TSV2N 0 VDC to 95 VDC 5 8 3 5 Outputs Pulses sent via the Front Plane to the selected probe Figure...

Page 226: ..._18 C4_18 U1_18 C7_18 C3_18 C25_17 C21_17 U2_17 C24_17 C20_17 C8_39 C4_39 U1_39 C7_39 C3_39 C714 C25_38 C21_38 U2_38 TP112 C24_38 C20_38 C8_36 C4_36 C7_36 C3_36 U1_36 C25_35 C21_35 U2_35 C24_35 C20_35...

Page 227: ...ne ON OFF DS18 GREEN CPLD IF_FPGA init ON OFF DS1 GREEN IF_FPGA GTXIF_DAVID_ERROR_sig 0 OFF OFF FPGA DAVID ERROR DS2 GREEN IF_FPGA GTXIF_DAVID_ERROR_sig 1 OFF OFF FPGA DAVID ERROR DS3 GREEN IF_FPGA GT...

Page 228: ...6 TP1 C147 S3 FL3 R50 R52 R20 F2 R101 TP9 R22 R3 U18 100 25 26 U13 TP13 TP7 R111 TP6 C46 K228 K229 K230 C144 K180 K190 K181 K182 C135 K116 K126 K117 K118 C98 R102 TP10 R103 C117 U17 CR1 U15 Q10 K233 K...

Page 229: ...Voltages 15VDC 6VDC 6VDC 15VDC 100V PMXVPP Voltage for Probe MUX 100V PMXVNN Voltage for Probe MUX Voltages provided by regulators on the board 12 VDC 5 VDC 3 3 VDC 3 3 VDC 5 VDC and LVDC 3 3 V 2 5A...

Page 230: ...K240 C145 K195 K207 K208 K196 S6 K131 K143 K144 K132 FL2 D10 D9 C38 R54 R55 Q6 TP1 C147 S3 FL3 R50 R52 R20 F2 R101 TP9 R22 R3 U18 100 25 26 U13 TP13 TP7 R111 TP6 C46 K228 K229 K230 C144 K180 K190 K181...

Page 231: ...Amplification on the echo signals To support 192 analog receiving channels from the probes two different GRX boards are used in VIVID E9 64 Channel Receiver board without analog CW Doppler 128 Channe...

Page 232: ...15 VDC 5 8 5 4 Outputs After Time Variable Amplification the analog signals are sent via high level analog differential lines to the DRX board for A D conversion and beamforming When using the Pedof...

Page 233: ...the Relay Board are routed via these boards Two different Front Plane boards have been used The first model B supporting up to four TX cards with 64 channels each was used from the introduction of VIV...

Page 234: ...m 64 channels into streams of digital data and 2 Receive signal beamforming The DRX performs optimal range dependent focusing and steering to create multiple receive beams simultaneously Up to four DR...

Page 235: ...ignal data to next DRX Data from the last DRX card is sent to GFI 5 8 7 6 LEDs on the DRX board the Nathan field The Nathan field has an array of LEDs that display Nathan status There is one green LED...

Page 236: ...column will turn ON then the LEDs in the next column are lit then the LEDs in the third column and at last the LEDs in the fourth row Next the sequence will repeat If the card starts the voltages are...

Page 237: ...from the BEP s hard drive via the PCI Express bus to the GFI2 board When a scanning mode is selected or scanning parameters are adjusted the GFI board receives setup parameters from the BEP and pass t...

Page 238: ...he Back Plane as the right most board in the FE rack 5 8 8 3 Input DC Voltages 24 DC from Main LV Power via FEP Backplane 15 VDC 6VA DC BEP5V Other needed voltages are generated locally on the GFI 5 8...

Page 239: ...1V2 STATUS LIT DS6 GREEN DRX1 LIT IF DRX 1 IS MISSING DARK BEFORE STARTUP DARK IF THE FPGAs ON THE BOARDS HAVE NOT BEEN SUCCESSFULLY PROGRAMMED DS7 GREEN DRX2 LIT IF DRX 2 IS MISSING DS8 GREEN DRX3 L...

Page 240: ...ferent boards Control signals and Clocks are distributed from the GFI board to the other boards Low amplitude analog signals from GRX board to DRX board Data signals are routed from DRX board to GFI b...

Page 241: ...REVISION 5 VIVID E9 SERVICE MANUAL Chapter 5 Components and functions theory 5 51 5 8 9 2 Location in the unit The FEP Backplane is attached to the rear side of the Front End Card Rack Figure 5 36 FEP...

Page 242: ...e following BEP models have been used for VIVID E9 BEP6 was introduced in manufacturing in October 2012 The BEP6 is also called BEPY3 BEP5 was introduced at the introduction of VIVID E9 and was phased...

Page 243: ...ERVICE MANUAL Chapter 5 Components and functions theory 5 53 5 9 4 Location of the Back End Processor BEP The BEP is located on the left side inside the scanner see Figure 5 37 Figure 5 37 Back End Pr...

Page 244: ...FROM MAIN PWR SUPPLY SATA SATA IO BOARD MOTHERBOARD LED AUX PHONO ECG BEP6 w 4D and DVR USB POWER PATIENT IO LED LED FRONT PANEL FROM MAIN PWR SUPPLY CONTROL SIGNAL TO MAIN PWR SUPPLY DVI Video Audio...

Page 245: ...supply for local voltages A motherboard with RAM a processor and PCI connectors for extension cards An I O board on the rear of the BEP 5 9 6 2 Outside the EMC Enclosure House Front Panel with status...

Page 246: ...9 13 48V 17 22 25 GND 18 PWR_LED_P 21 PWR_LED_N 5 Test Connector Factory Test Pins of note 5 13 23 Ground 3 24V 4 48V 8 12V 9 5V 10 AC_FAIL_N 11 3 3V 12 5V_STDBY 16 Not Used on Vivid E9 24 PWR_SW 25...

Page 247: ...onnections cont d Table 5 10 BEP6 Top connections ITEM DESCRIPTION ILLUSTRATION 1 J1 Main PS 48V 5V In 2 J5 Front End Rack PCIe to GFI Table 5 11 BEP6 Rear connections ITEM DESCRIPTION ILLUSTRATION 1...

Page 248: ...r BEP 5 9 8 BEP5 Face and Top connections Please refer to the illustrations Figure 5 41 BEP5 Face connections Figure 5 42 BEP5 Top connections Sub woofer PCI Express to GFI SATA to DVD Main PS 48VDC 5...

Page 249: ...B distribution USB is used to communicate with and or control many functions and devices in the VIVID E9 The illustration below shows how the USB signals are distributed in BEP6 The illustration below...

Page 250: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 5 60 Section 5 9 Back End Processor BEP 5 9 8 2 SATA distribution BEP6 Figure 5 45 BEP6 SATA distribution...

Page 251: ...and functions theory 5 61 5 9 8 3 Use of Expansion Slots on BEP6 Motherboard Use of Expansion Slots listed from the rightmost slot BEP 6 0 Power Board Assembly Graphics Adapter Video Card DVR Board O...

Page 252: ...t beep will last only 1 second BIOS also uses beeps of different frequency to indicate critical errors If BIOS detects that the CPU is overheating it may play a high pitched repeating beep while the c...

Page 253: ...ECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 5 Components and functions theory 5 63 5 9 9 Input DC Voltages NOTE The BEP is not connected to AC power The BEP gets it power supply via the...

Page 254: ...MANUAL 5 64 Section 5 9 Back End Processor BEP 5 9 10 Input Signals BEP6 Audio signal path GFI FEP Backplane BEP6 MBD a and b a IO Board Audio Out b DVR BEP5 Audio signal path GFI FEP Backplane BEP I...

Page 255: ...uted via printed circuits on the cards No cables are used Table 5 14 BEP5 Bi directional Signals Signal Name Description Signal Path PCIe PCI Express J3 BEP Motherboard Cable 4 inside BEP J5 BEP PCI E...

Page 256: ...EDs 5 9 13 1 LEDs on the BEP5 BEP6 Front Figure 5 47 Front Panel Table 5 16 LEDs on the BEP s front ITEM LED NAME DESCRIPTION 1 ACT LED HD ACTIVITY Blinks when the Hard Disk drive is active 2 SPD LED...

Page 257: ...5 67 5 9 13 2 LEDs on the BEP6 s Front Cover Table 5 17 LEDs on the BEP6 s Front Cover DESCRIPTION ILLUSTRATION The LEDs for the BEP6 s Power Supply can be viewed without removing the BEP6 s Front Co...

Page 258: ...ing Green Amber or OFF Hub3 OK Error Green LED means OK Blinking Green Amber or OFF Hub2 OK Error Green LED means OK Blinking Green Amber or OFF MC OK Error Green LED means USB micro controller is OK...

Page 259: ...Power Supply Dedicated control signals are used for controlling the BEP Power Supply 5 9 14 2 BEP6 Power Supply Block Diagram Table 5 20 BEP6 Power Supply Block Diagram ITEM DESCRIPTION ILLUSTRATION 1...

Page 260: ...Block Diagram 5 9 14 4 Location in the BEP6 The BEP6 Power Supply Card is located inside the BEP6 at the right hand side when opening the BEP Figure 5 49 BEP5 Power Supply Block Diagram Figure 5 50 B...

Page 261: ...5 9 14 5 Location in the BEP5 The BEP Power Supply is located inside at the top of the BEP5 5 9 14 6 Input Voltage BEP6 BEP5 48 VDC from the Main Power Supply 5 9 14 7 Input Signals BEP6 BEP5 48V_OK f...

Page 262: ...n of the LEDs refer to the label on the BEP6 s Front Cover or to 5 9 13 2 LEDs on the BEP6 s Front Cover on page 5 67 5 9 14 11 LEDs on BEP5 Power Supply Several LEDs are viewable through cut outs in...

Page 263: ...est Points are available through cut outs in the BEP Power supply NOTICE Be careful so you don t short circuit the signals when placing the test probe s Figure 5 53 Test Points on BEP5 Power Supply Ta...

Page 264: ...or for connection to the BEP5 Motherboard The IO Board is interfaceing between the BEP and the connectors on the rear side of VIVID E9 video audio USB BEP and the Top Console BEP and the internal prin...

Page 265: ...Ds on the BEP5 Rear Panel Two LEDs are located to the right for each USB connector The upper LED next to a USB connector is USB connected and in use Lit green The lower LED next to a USB connector is...

Page 266: ...Adapter is located in the PCI Express slot PCI Express x16 inside the BEP cabinet 5 9 16 3 Input Signals Graphics data from the BEP via the BEP s motherboard 5 9 16 4 Output Signals DVI to the LCD sc...

Page 267: ...ral description The Main Power Supply s main task is to galvanically isolate the scanner from the on site Mains Power System and to supply the various internal subsystems with AC or DC power Figure 5...

Page 268: ...DC power with several voltages 24 VDC 15 VDC 11 VDC 6V VDC Front End Rack TXPSV1 and TXPSV2 for the transmitters Front End Rack PMXVOUT for the probe channel multiplexers Back End Processor 48 VDC Ope...

Page 269: ...male plug to the wall outlet 5 10 2 2 Temperature Control The Main Power Supply is equipped with an internal fan with variable speed for temperature control Both the temperature of the air entering t...

Page 270: ...s on the BEP The USB is used to set the power supply to the correct transmit XD voltages Table 5 24 Power outputs sheet 1 of 2 CONNECTOR DESCRIPTION COMMENTS P2 115 VAC OUT ON TWO CONNECTORS PERIPHERA...

Page 271: ...y of the outputs are overloaded Over voltage Protection is provided for these voltages 3 3 V 5 Vd 5 Va 15 Va Voltage will be turned off if the temperature grows too high temperature watch dog P5 48 VD...

Page 272: ...osition Connect the mains power to the VIVID E9 and switch AC Breaker to ON position The fans inside the Main Power Supply starts on a high speed and then settle to a lower speed 5V_STB 5V DC Standby...

Page 273: ...and functions theory 5 83 5 10 3 2 AC Breaker to ON position cont d The 5Vstdby LED is lit green The ON OFF switch on the OP is lit amber color One green LED LED 1 on the BEP s motherboard is lit Fig...

Page 274: ...sequence The fans below the Front End Card Rack starts The lights in the alphanumeric keyboard and in the Lower OP panel are turned on AC Power to the B W printer is turned on Figure 5 62 ON has been...

Page 275: ...oon as the BEP Power Supply gets power from the Main Power Supply the different voltages needed for the BEP are turned on and the BEP starts to boot 1 The BIOS is loaded 2 Then the DVD drive is checke...

Page 276: ...bed below 5 10 4 2 Power Down Press the ON OFF button for a short time 1 BEP detects the contact of Power ON OFF switch 2 PSON_N goes high This trigs the Main Power Supply to shut down the output volt...

Page 277: ...hut down the output voltages 3 Controller ACFAIL_N output signal goes low 4 Controller TS_OK output signal goes low 5 Controller turns OFF the TS Transmit voltages 6 Controller turns OFF the PMX Probe...

Page 278: ...ral description NOTE The ECG functionality of the Patient I O module is not intended for patient monitoring nor to support alarm functionality This input is intended as a tool for easier synchronizati...

Page 279: ...e scanned image that is displayed is synchronized with the ECG respiration and phono traces In M Mode or Doppler the traces are synchronized to that particular mode s sweep The operator can control th...

Page 280: ...ble below The AUX is default a 1 Vpp Volt peak to peak input with a max frequency of 300 Hz It has a programmable high gain mode with a maximum input signal of 300 mVpp millivolt peak to peak The inpu...

Page 281: ...theory 5 91 5 11 2 5 Patient I O outputs USB2 Digital Trace Data Module ID PROM communication Pinout for the USB outlet on the Patient I O module Table 5 27 USB outlet on the Patient I O Module CONNE...

Page 282: ...r IO the OP and to internal units printer and Patient IO It distributes SATA to the DVD s It splits video between the main monitor and rear video output It also distributes audio to speakers and rear...

Page 283: ...h Two USB ports for the Rear IO One USB port for internal B W printer One USB port for XYZ Controller Some unused USB ports Doppler audio orginating on the GFI board Doppler audio mixed with Windows s...

Page 284: ...e between the BEP and the rest of the system It contains three USB hubs and it splits video between the main monitor and rear video output It also distributes audio to speakers and rear connectors 5 1...

Page 285: ...t from the video card or DVR Board AC Fail and PSON_N signal from the BEP PS 5 11 4 4 Output signals 12V DC and 5V DC for Patient IO DVD drive and High capacity drive 5V STDBY from the BEP power suppl...

Page 286: ...m the front of the VIVID E9 5 12 1 2 Digital Video Stream Recorder DVR Option This is a kit with a Video Recorder Card inside the BEP and a DVD R W unit used for export of the recorded video The DVD R...

Page 287: ...ments only USB Flash Cards tested for use with VIVID E9 may be used For a list of available USB Flash Cards see 9 15 6 USB Flash Card on page 9 69 The following USB Flash Cards USB Memory keys have al...

Page 288: ...time NOTE To be able to use an USB storage device like this Iomega Ultramax Desktop Hard Drive on a Vivid E9 the Vivid E9 may need a BIOS upgrade A CD with the new BIOS software is included in the kit...

Page 289: ...and service the VIVID E9 scanner is collected in the documents described in this section 5 13 1 User documentation VIVID E9 User Manual User Guide Special Probes User Guide 5 13 2 Service documentatio...

Page 290: ...sections contain a brief introduction of iLinq s features 5 14 3 1 Web Server Browser The Service Platform and other Service software use the iLinq web server and the Internet Explorer browser 5 14 3...

Page 291: ...onalization The user interface provided by the service platform is designed for GE personnel and as such is in English only There is no multi lingual capability built into the Service Interface 5 14 4...

Page 292: ...use Every access request whether successful or not will be logged into a service access log that is viewable to authorized users Section 5 15 Restart VIVID E9 after diagnostics Always shutdown the sy...

Page 293: ...his chapter describes how to adjustments the VIVID E9 6 1 2 Contents in this chapter 6 1 Overview 6 1 6 2 LCD Monitor adjustments 6 2 6 3 Test the LCD Arm and LCD Monitor range of motion 6 4 6 4 Backl...

Page 294: ...s have been optimized at the factory so normally there should be no need for any further adjustments Default Brightness for the 19 LCD screen on VIVID E9 is 89 Under special light conditions you may w...

Page 295: ...access Utilities Test Patterns to review the test patterns Figure 6 2 LCD Controls Advanced Menu Table 6 1 Advanced Menu MENU PARAMETER DEFAULT SETTING Screen Smoothing 1 2 3 4 5 3 Return Color Bright...

Page 296: ...ition Figure 6 3 Too tight If the customer finds the LCD difficult to move from a vertical to horizontal position use a 17mm wrench to loosen nuts at hinge Be sure to loosen both sides equally Start w...

Page 297: ...adjustment 6 4 1 Purpose of this section This section describes how to adjust the backlight on the 17 LCD monitor 6 4 2 Cautions and Warnings N A 6 4 3 Accessing the Main LCD and Touch Screen Backlig...

Page 298: ...t the left most rotary to adjust the LCD Backlight Push and adjust the left most rotary to adjust the LCD Blue Tint Adjust the second rotary from the left side to adjust the Touch Panel TP Backlight 6...

Page 299: ...Test Images Select Page 2 of 2 on the Utility screen Turn the second rotary from the left side to select between the Test Images The Test Images are reproduced in Table 6 2 Figure 6 8 Select Test Imag...

Page 300: ...REVISION 5 VIVID E9 SERVICE MANUAL 6 8 Section 6 4 Backlight adjustment 3 Test Green grades 4 Test Blue grades 5 Blue Circle on Black background 6 Black Table 6 2 Test Images cont d sheet 2 of 3 TEST...

Page 301: ...HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 6 Service adjustments 6 9 7 White 8 Straight Line Table 6 2 Test Images cont d sheet 3 of 3 TEST IMAGE No TEST IMAGES DESCRIPTI...

Page 302: ...rk Follow these steps to calibrate the Touch screen 1 Select Calibrate The Touch screen goes blank and display a cross in the screen s upper left corner 2 Point your finger on the centre of the cross...

Page 303: ...talling a new VIVID E9 after software has been updated or replaced when DRX boards have been interchanged when a DRX board has been replaced when the Back End Processor has been replaced 6 6 3 Front E...

Page 304: ...brakes in the froglegs to allow easy movement When the lock button is pressed and the console is pushed back into the locked position the U bolt mechanically engages the park lock similar to a car doo...

Page 305: ...he XY Assembly This is the adjustment screw NOTE When turning the hex screw do not exceed half turn increments Overtightening will prevent the console from locking into place and too loose will make t...

Page 306: ...6 7 3 XY Manual Release for Lock and Brake Mechanism and Adjustment The following procedure is intended to release and adjust the XY mechanism 6 7 3 1 XY Lock and Brake Mechanism Parts Figure 6 13 XY...

Page 307: ...e if the park lock is not working or the lock does not respond when pressing the Frogleg Controls 1 Release the lock manually 2 Remove the four screws Item 1 The screws s heads are marked with red col...

Page 308: ...is occurred in some earlier systems The nut and retainer did not return to the charged mode after unlock By hand rotate the Threaded Lead Screw on the actuator 4 counterclockwise until the Lock Nut is...

Page 309: ...nsole into the locked position until the lock is engaged Not apply any weight on the console or lean on it If the console is not in the normal resting position the lock will not engage when trying to...

Page 310: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 6 18 Section 6 7 Adjusting the XYZ Mechanism This page left blank to facilitate double sided printing...

Page 311: ...system operation Basic host system and board level diagnostics are run whenever power is applied Some Service Tools may be run at the application level 7 1 2 Contents in this chapter 7 1 Overview 7 1...

Page 312: ...t start up of the VIVID E9 and is repeated at specific pre defined intervals 7 3 2 Reset the BEP from a Hang Question It seems that the BEP is hanging it s not responding at all How do I Reset it Answ...

Page 313: ...There is a lever behind the openings in the Rear Cover Press the lever to the right to release the X Y brakes At the same time push the Upper Console down If you stop to press the lever the Z brakes e...

Page 314: ...Shortcuts SHORTCUT DESCRIPTION Alt B BOOKMARK PLACE A BOOKMARK IN LOGS Alt D SYSTEM PROBLEM REPORTING See 7 6 2 Collect a Trouble Image with Logs on page 7 11 PRINT SCREEN Make a SCREENSHOT See 7 6 2...

Page 315: ...th the new probe reload software Do not reload presets until you have tested the system with the default settings and be sure that the problem does not persist User Defined Presets can carry corruptio...

Page 316: ...bout the occurrence of the problem and the date and time it showed up Reload the software 7 4 3 2 CD DVD Drive Failures Check that the media disk capacity and speed is supported For the CD the capacit...

Page 317: ...Assembly might be defective too 7 4 4 2 No Video on LCD Display If the video is too dim has dimmed areas or there is no video at all replace the HV LCD inverter Check the cabling within the system If...

Page 318: ...work Connectivity Troubleshooting 7 5 1 First Status Select the network icon on the bottom of the screen to get a first status for the network connectivity If the status reports No network verify cabl...

Page 319: ...eed to reboot the VIVID E9 Check device configuration Clean the spooler F4 Check Connectivity configuration on the VIVID E9 If it is a printer check that the printer supports the film type and format...

Page 320: ...stem Information The following information is necessary in order to properly analyze data or images being reported as a malfunction or being returned to the manufacturer Product Name VIVID E9 Select C...

Page 321: ...ted CD R DVD R or save to the Export directory D drive for remote viewing through InSite NOTE You MUST select one of the available devices as the destination device if it is to be different than the d...

Page 322: ...then be viewed by the OLC 7 7 3 To Capture a Screen Image Using the Shortcut 1 With the desired image displayed on the screen press Ctrl and PrtSc print screen keys simultaneously 2 From the touch pan...

Page 323: ...he VIVID E9 See 4 2 1 Power ON Boot Up on page 4 4 for detailed instructions After scanner initialization is complete continue with the steps below 2 Press CONFIG on the Utility tab on the Touch Panel...

Page 324: ...If in a quiet environment it should be possible to hear the Lock Engine engage for a few seconds 8 Check if the Lock function works now If the lock still does not work try to press the TestXY button...

Page 325: ...situation This will if the situation is trapped bring up a dialog telling that the system is not responding 1 Press Restart to restart the application This will save a special debug log When the syst...

Page 326: ...MANUAL 7 16 Section 7 9 Troubleshooting 7 9 2 Unable to scan Figure 7 5 Unable to scan Unable to scan Check the probe connection YES NO Are probes connected in right way Restart the scanner Are other...

Page 327: ...ng NOTE Cables are not labeled at the XY controller so right side and left side motors are not defined When troubleshooting the XY brake motors keep this in mind The rear motors are 1 and 3 The front...

Page 328: ...release lock mechanism is located to the rightmost section of the drawer thus the force needed to release and lock the alphanumeric keyboard will increase significantly if the leftmost section of the...

Page 329: ...n screen Try a replacement USB Footswitch 7 9 8 USB Device s stopped working If a keyboard and mouse or other equipment not intended for connection to the VIVID E9 has been connected to any of the USB...

Page 330: ...stem If the air filters become too dusty a higher fan speed is required to keep the air stream at the needed level When the fan speed increase the fan noise will also increase If the cooling air strea...

Page 331: ...e VIVID E9 complies with limits as stated on the EMC label However there is no guarantee that interference will not occur in a particular installation Possible EMI sources should be identified before...

Page 332: ...from other EMI sources Special shielding may be required to eliminate interference problems caused by high frequency high powered radio or video broadcast signals Ground the unit Poor grounding is the...

Page 333: ...anner to another location and verify if the noise changes or disappear This may help you to locate an external noise source Try to move the scanner to another location inside the room another room ano...

Page 334: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 7 24 Section 7 10 Noise troubleshooting This page left blank to facilitate double sided printing...

Page 335: ...s in this chapter 8 1 Overview 8 1 8 2 Warnings and important information 8 2 8 3 Definitions of Left Right Front and Back 8 5 8 4 Reloading the software 8 6 8 5 Replacing Covers and Bumpers 8 38 8 6...

Page 336: ...still plugged into the AC Outlet WARNING WARNING BECAUSE OF THE LIMITED ACCESS TO CABINETS AND EQUIPMENT IN THE FIELD PLACING PEOPLE IN AWKWARD POSITIONS WE HAVE LIMITED THE LIFTING WEIGHT FOR ONE PE...

Page 337: ...with visible body fluids be taken or shipped from a clinic or site for example body coils or an ultrasound probe The purpose of the regulation is to protect employees in the transportation industry as...

Page 338: ...M3 3 BIT TX 20 M4 4 BIT TX 25 M5 5 BIT TX 30 M6 6 BIT TX 45 M10 7 FLAT BLADE SCREWDRIVER 3 2 mm 8 FLAT BLADE SCREWDRIVER 4 mm 9 FLAT BLADE SCREWDRIVER 6 mm 10 PHILLIPS SCREWDRIVER PH1 11 PHILLIPS SCR...

Page 339: ...ter 8 Replacement procedures 8 5 Section 8 3 Definitions of Left Right Front and Back The Figure below illustrates what is Left Right Front and Rear or Back of the VIVID E9 Figure 8 1 Definition of Le...

Page 340: ...CTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 8 6 Section 8 4 Reloading the software Section 8 4 Reloading the software 8 4 1 Purpose of this section This section describes how to reinstall the so...

Page 341: ...112 1 x v112 1 x v112 1 x GA200890 BEP5 w 4D Nvidia v104 3 x v112 0 x or higher v112 1 x GA000950 Vivid E9 100 230V 4D Expert Option 19 LCD GB200001 BEP6 w 4D v104 3 4 or higher v112 0 7 or higher v11...

Page 342: ...or GA200800 BEP5 w 4D Nvidia v104 2 x v104 1 x v110 0 x GA000830 VIVID E9 100 230V 2D 17 LCD GA200804 VE9 Card Rack Complete w MLA4 GB200002 BEP6 wo 4D v104 3 x v110 1 12 v112 1 x GA200900 or GA200805...

Page 343: ...atient Archive and User Defined Settings Password for the user ADM The default password for the user ADM is ulsadm If the password has been changed by the site you should record it for your own use be...

Page 344: ...of the Patient Database Depending on the location set up either move the images to a remote server or to removable media like DVD or CD discs As the images are moved the database will point to the new...

Page 345: ...he printer s IP Number The printer s IP number is not easily available via the VIVID E9 s menus Print out a status sheet on the printer For more information please refer to the printer s documentation...

Page 346: ...ed on the VIVID E9 has become corrupted VIVID E9 s configuration setup has issues that are difficult to troubleshoot and resolve so it may be easier to do a software reload and start the setup from sc...

Page 347: ...compressed original files from the DVD and CD were stored on a separate disk partition on the hard disk drive labeled REPOSIT Repository If you are going to reinstall the software you can install from...

Page 348: ...line Center OLC Tell OLC that you need a Challenge Code and on request you must read the Challenge String on the screen OLC will return a password that you need to type in as the Response 5 Enter the...

Page 349: ...6 Press the Enter key on the keyboard The software reload starts NOTE Time to complete the software reload is typically 15 to 25 minutes When the System Software reload is done the VIVID E9 reboots Af...

Page 350: ...CD with Printer Driver software When installing the software start with the System Software then continue with the Application Software and if included install the patch software after you loaded the...

Page 351: ...d Next the BEP starts to boot from the disc in the DVD drive Several screens will be displayed indicating the progress 5 If asked for press F and the booting continues Figure 8 5 BIOS Boot Screen BEP6...

Page 352: ...ased If backup has not been performed all data like Patient Database System Configuration and User Configurations Customer Presets will be lost Table 8 4 Descriptions of your choices WHEN TO USE DESCR...

Page 353: ...Figure 8 8 Are you really sure Delete HDD Content This will erase the whole harddisk before installation Are you really sure y n This will erase the whole harddisk before installation Are you really...

Page 354: ...ress a key to initiate the reboot NOTE Let the DVD remain in the DVD tray 3 Press any key to continue The VIVID E9 will reboot The monitor may display the message No sync for a few seconds before the...

Page 355: ...ANUAL Chapter 8 Replacement procedures 8 21 8 4 10 5 Press A Install System Software Erase all partititions cont d The VIVID E9 loads the BIOS The BEP starts to boot into PC DOS Refer to Figure 8 6 on...

Page 356: ...he copied Image file is used as source when the System software is installed on the C partition Several screens will be displayed When the installation of the System Software has finished the DVD is e...

Page 357: ...ont d Loading the Application Software 1 If not already done remove the System Software DVD and insert the applicable Application Software CD in the DVD drive 2 Press any key to continue The VIVID E9...

Page 358: ...0 5 Press A Install System Software Erase all partititions cont d When the screen below is displayed just wait and the installation continues Several screens are displayed as illustrated below The VIV...

Page 359: ...5 Press A Install System Software Erase all partititions cont d The monitor may display the message No sync for a few seconds before the BIOS Boot Screen is displayed When the screen below is displaye...

Page 360: ...em Software Erase all partititions cont d The booting continues When booted the Start Application dialog will display on the screen 3 Select Install SW to continue the installation 4 Select OK 5 Selec...

Page 361: ...Replacement procedures 8 27 8 4 10 5 Press A Install System Software Erase all partititions cont d The LoadSoftware bat file starts to run Figure 8 23 Starting the LoadSoftware bat file 1 2 3 Text in...

Page 362: ...he computer name has been generated previously the pop up screen in Figure 8 27 will be displayed 6 Enter the VIVID E9 s Serial Number The Serial Number is located on a label at the rear of the VIVID...

Page 363: ...and New Computer Name is correct or select Change if you want to change the Serial Number When confirmed the software installation starts Figure 8 27 Confirm the Serial Number CAUTION Computer Name mu...

Page 364: ...ion is complete you are prompted to restart the VIVID E9 but remove the CD first 8 Remove the CD 9 Select OK to restart When you have selected OK to restart the CD DVD drawer will close and the VIVID...

Page 365: ...ing for the Option Key is displayed 10 Type the Option Key and select OK to continue The VIVID E9 continue to boot 11 Next If any Software Patches or Virus Hardening software CDs are available continu...

Page 366: ...update the software on C Next a screen with the following text is displayed This will repair upgrade partition C Are you really sure y n 2 Select y The System Software image file is being copied from...

Page 367: ...tallation Installing the Application Software Follow these steps to install the Application Software 1 Remove the System Software DVD 2 Insert the Application Software CD 3 Press a key on the keyboard...

Page 368: ...software installation Figure 8 38 Select OK to start the software installation 6 Select OK to start the software installation Next the Application SW Installation warning is displayed on the screen F...

Page 369: ...re 8 40 Confirm the Serial Number 8 Select OK to confirm that the Serial Number is correct or select Change if you want to change the Serial Number When confirmed the unpacking of the software starts...

Page 370: ...installed the InSite ExC software is installed Figure 8 43 Installing the InSite ExC software After the InSite ExC software has been installed the installation is complete and you are prompted to rest...

Page 371: ...Based on your recordings correct the settings as needed 4 Continue with Section 3 11 Options Setup on page 3 49 Based on your recordings correct the settings as needed 5 After a software load you sho...

Page 372: ...r holes give access to the locks By inserting a 1 Phillips screwdriver 10 16 cm 4 6 inch length or similar size and length tool into the holes and bending the handle on the screwdriver down to the flo...

Page 373: ...it reaches the lock mechanism 2 Push the handle on the screwdriver downwards to release the lock 3 Repeat the steps above for the other lock 4 Lift the cover forward and up to remove it from the VIVI...

Page 374: ...ooks onto the Top Cover 3 Align and squeeze the front edge of the side cover to latch it into place 4 Position the Side Cover s side lock first 5 Position the Side Cover s front lock 6 Align and squee...

Page 375: ...ERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCEDURES 2 TURN OFF THE BREAKER 3 UN...

Page 376: ...E9 SERVICE MANUAL 8 42 Section 8 5 Replacing Covers and Bumpers 8 5 3 2 Top Cover removal procedure Follow these steps to remove the Top Cover 1 Remove two screws 2 Remove the Top Cover Figure 8 48 To...

Page 377: ...r installation procedure Follow these steps to install the Top Cover 1 Position the Top Cover onto the Front Cover at the four hooks 2 Hook Top Cover onto Front Cover 3 Install and tighten the two scr...

Page 378: ...Side Cover 2 Remove the Side Bumper from the Side Cover 8 5 4 4 Side Bumpers installation Follow these steps to install the Side Bumper 1 Align the fastening holes in the Side Bumper with the holes in...

Page 379: ...ow Figure 8 51 the Side Cover was removed to be able to view the Side Latch You don t need to remove the Side Cover to perform this procedure 8 5 5 3 Foot Rest Bumper installation Follow these steps t...

Page 380: ...VID E9 SERVICE MANUAL 8 46 Section 8 5 Replacing Covers and Bumpers 2 Push the Foot Rest Bumper down and over the pedals You may need to push down the pedals to be able to place the Foot Rest Bumper 3...

Page 381: ...Assembly to the chassis 2 Pull the upper end of the Front Cover out and upwards to free it from the pedals and the frame WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PA...

Page 382: ...nt Cover 1 Thread the Front Cover so it fits in between chassis and pedals 2 Align the Front Cover guide pins with holes in the frame 3 Fasten Front Cover with two 2 pc screws and washers NOTE Orienta...

Page 383: ...uide is attached with hatches 1 Use a flat bladed screwdriver to loosen the hatches 2 Remove the Plate Connectors w Guide WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING P...

Page 384: ...E9 SERVICE MANUAL 8 50 Section 8 5 Replacing Covers and Bumpers 8 5 7 3 Plate Connectors w Guide installation 1 Position the plate and snap it into position 2 Install Front Cover 3 Install Foot Rest...

Page 385: ...Follow these steps to install the Filter Cover 1 Inspect the Filter and clean if necessary 2 Place the Filter Cover into position and press the Filter Cover top corners until the Filter Cover locks en...

Page 386: ...ove two screws on each side of the Rear Cover 2 Lift the Rear Cover away WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOL...

Page 387: ...re the Z mechanism s manual release handle extends through the air vent on the Rear Cover 3 Position the Rear Cover into place 4 Gently pull up on the Z mechanism s manual release handle to confirm pr...

Page 388: ...the lock 3 Open the Door I O Panel 4 Hold the Door I O Panel near the upper hinge and pop hinge out and away from the upper hinge post on the Rear Cover WARNING WARNING ENERGY CONTROL AND POWER LOCKO...

Page 389: ...cont d 8 5 10 3 Install the I O Door Follow these steps to install the Door I O Panel 1 Place the Door I O Panel into position by sliding lower hinge onto lower hinge post 2 Pop the upper hinge into p...

Page 390: ...replacement 8 5 11 1 Preparations When preparing for the replacement you must perform the following steps 1 Power down the VIVID E9 2 Disconnect the Mains Power Cable from the wall outlet 3 Remove th...

Page 391: ...ook removal Follow these steps to remove the Cable Hook 1 Locate the Cable Hook s 2 Place the Rear Cover face down on a protected flat surface 3 Remove the screw securing the Cable Hook 4 Repeat step...

Page 392: ...Cable Hook 1 Position the Cable Hook on the Rear Cover 2 Install the screw to secure the Cable Hook to the Rear Cover 3 Repeat step 2 to replace the other Cable Hook if necessary 4 Install the Rear Co...

Page 393: ...Follow these steps to install the Rear Bumper 1 Place the Rear Bumper into position 2 Install the 2 screws to secure the Rear Bumper WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN...

Page 394: ...e Rear Handle 1 Install the Rear handle in position so its fastening holes are flush with the holes in the frame WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF T...

Page 395: ...N SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCEDURES 2 TURN OFF THE BREAKER 3 UNPLUG THE VIVID E9 4 MAINTAIN CONTROL OF THE POW...

Page 396: ...ver WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCEDURES 2 TUR...

Page 397: ...e DVD drive 3 Raise the console to full height 4 Remove the remaining screw on the right side that secures the Column Cover Assembly 5 Remove the 2 screws on the left side that secure the Column Cover...

Page 398: ...se steps to install the Column Cover Assembly 1 Install the Main Cable Cover so the Column Cover Assembly overlaps Main Cable Cover edges 2 Install screws to Column Cover Assembly tighten by hand 3 Po...

Page 399: ...The Main Cable Cover is held by four screws Follow these steps to remove the Main Cable Cover 1 Remove the Main Cable Cover WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING...

Page 400: ...s Main Cable Cover edges see Figure 8 66 2 With the console raised to its full height install the 3 screws to secure the Main Cable Cover and Column Cover Assembly 3 Lower the console and install the...

Page 401: ...release point and pressing until release Pull the console out to its extended position to gain access to the screws in the next step WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN...

Page 402: ...wo types of covers one type has two screws 1 and the other type has one screw 2 3 Remove the screws from the cover s to replace 4 Pull down and slide the cover away from the XY mechanism Be sure to fl...

Page 403: ...XY Frogleg motors installation procedure Follow these steps to install the Covers under XY Frogleg motors 1 Thread the cover s on the cable 2 Perform a dry fit of the covers confirm the covers face t...

Page 404: ...and pressing until release Pull the console out to its extended position to gain access to the screws in the next step WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PAR...

Page 405: ...emove the Bulkhead Cover 1 Use thumbs to press upper lock tabs toward the front of the VIVID E9 to release the top of the Bulkhead Cover or if needed use a 4 mm flat blade screwdriver to release the 2...

Page 406: ...these steps to install the Bulkhead Covers 1 Place the Bulkhead Cover s mounting angle onto the lower edge of the bulkhead opening NOTE Tuck any cables within the cover to avoid pinching the cables 2...

Page 407: ...ables to the monitor To get access to the cable connectors remove the LCD Rear Cover Follow these steps to remove the LCD Rear Cover 1 For easy access tilt the LCD Monitor forward to horizontal positi...

Page 408: ...GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 8 74 Section 8 6 Top Console Parts Replacement 2 Unscrew the two fixing screws on the rear side of the LCD Monitor assembly 3 Remove the LCD Rear Cover and...

Page 409: ...N 5 VIVID E9 SERVICE MANUAL Chapter 8 Replacement procedures 8 75 8 6 2 3 Disconnect the 17 LCD Monitor Cables 1 Disconnect the two cables 2 Tilt the monitor back to vertical position Figure 8 78 Disc...

Page 410: ...LCD Monitor assembly Two screws are used for fixing the LCD Monitor assembly to the LCD Arm Bracket 1 Loosen and remove the screws 2 Lift the LCD Monitor assembly upwards until you can lift it away fr...

Page 411: ...lace the LCD assembly onto the LCD Arm Bracket Be sure that the notches in the LCD fixing bracket is well aligned onto the corresponding positions on the LCD Arm Bracket 2 Install the two screws M5x8...

Page 412: ...the 17 LCD Monitor cont d 4 Connect the two cables Arrange the cables as illustrated in Figure 8 83 Connect cables 5 Install the LCD Rear Cover and fasten it with the fixing screws Figure 8 83 Connec...

Page 413: ...onitor assembly covers the two cables to the monitor To get access to the cable connectors remove the Cable Cover Follow these steps to remove the Cable Cover 1 For easy access tilt the LCD Monitor fo...

Page 414: ...r side of the LCD Monitor assembly 3 Remove the Monitor Cables Cover Carefully insert a screwdriver into the Removal Slot to separate the cover from the LCD Monitor assembly 4 Lift the cover away and...

Page 415: ...the 19 LCD Monitor Cables 1 Disconnect the two signal cables 2 Remove the 2 cable clips if present NOTE If you are replacing any cables you will also need to remove 2 additional cable clips at the bra...

Page 416: ...Monitor assembly to the Monitor Bracket 1 Loosen the four screws by turning each screw between one half and one turn counter clockwise You don t need to remove the screws 2 Lift the LCD Monitor assem...

Page 417: ...1 Install the LCD Monitor assembly on the Monitor Bracket Verify that all four fixing screws have engaged in their slots 2 Tighten the four screws 3 Tilt the monitor to horizontal position 4 Connect...

Page 418: ...ID E9 SERVICE MANUAL 8 84 Section 8 6 Top Console Parts Replacement 8 6 3 8 Install the Monitor Cables Cover 1 Install the Monitor Cables Cover and fasten it with the fixing screw Figure 8 91 LCD Moni...

Page 419: ...over WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCEDURES 2 TU...

Page 420: ...nnector Lock upwards then disconnect the HDMI plug 8 6 4 3 Remove the LCD Arm assembly Follow these steps to remove the LCD Arm assembly 1 Slide the LCD Mount Lock Handle into unlocked position 2 Move...

Page 421: ...he LCD Arm Assembly 1 Carefully install the LCD Arm assembly into position first feeding the LCD Arm cables down through the console opening 2 Push the LCD Mount Lock Handle into locked position Figur...

Page 422: ...nnect the LCD cables to the connectors on the Bulkhead 4 Install the Bulkhead Cover 5 Install the LCD monitor 8 6 4 5 Calibration and adjustments See Section 6 4 Backlight adjustment on page 6 5 for L...

Page 423: ...G WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCEDURES 2 TURN OFF THE...

Page 424: ...LCD Cables NOTE When handling the arm use the arm lock unless the arm is being turned Locking will help stabilize the arm NOTE LCD cable color may be different than pictured 1 Remove the two M4 x 8 s...

Page 425: ...emove the bumper cover from the end of the first arm as shown in Figure 8 99 3 Remove the covers from all three arm sections of the arm assembly a Remove the M4 x 8 screw holding the first arm section...

Page 426: ...he second arm section cover with Phillips screwdriver 4 Cut tie wraps securing cables in arm 5 Remove the cable s to be replaced 6 Mark cables After removing the cable s to be replaced according to th...

Page 427: ...tarting at the console end of the first arm section thread the replacement cable s through the arm sections toward the monitor end 2 Feed cable through opening opposite the arm lock mechanism side of...

Page 428: ...ng covers on arms test arm motion and cable stress 10 Secure the cables in the locations indicated in Figure 8 103 Band the tie wrap through the clamp 11 Replace the second arm cover 12 Replace the th...

Page 429: ...RECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 8 Replacement procedures 8 95 8 6 5 4 Calibration and adjustments See Section 6 4 Backlight adjustment on page 6 5 for LCD Monitor calibrati...

Page 430: ...RNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCEDURES 2 TURN OFF THE BREA...

Page 431: ...ssary to remove the LCD Monitor or the LCD Arm 1 Remove the 2 screws securing the LCD Mount Lock 2 Remove the LCD Mount Lock 8 6 6 3 LCD Mount Lock installation procedure Follow these steps to install...

Page 432: ...NNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIV...

Page 433: ...rear of the VIVID E9 release the console s XY mechanism by inserting a screwdriver into the release point and pressing until release Pull the console out to its extended position to gain access to th...

Page 434: ...the Upper OP Panel Touch Panel assembly slightly from the bottom and then tilt the top toward the front of the VIVID E9 There are tabs at the bottom of the Touch Panel Assembly Pull straight up on th...

Page 435: ...ce 8 6 7 3 Install the Upper OP Panel Touch Panel Assembly Follow these steps to install the Upper OP Panel Touch Panel Assembly 1 Place the Upper OP Panel Touch Panel Assembly in the frame 2 Ensure t...

Page 436: ...ion J21 USB Cable 2 part of Main Cable in position J22 Bulkhead Board USB Cable 1 from bulkhead position closest to the OP in position J24 Bulkhead Board USB Cable 2 from bulkhead position most far fr...

Page 437: ...ouch Panel Assembly by securing the 5 tabs and 2 hooks 5 Install the four screws to the Back Cover from behind 6 Install the five OP Panel Knobs Figure 8 115 Cables to Upper OP NOTICE Be careful not t...

Page 438: ...P TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENERGY C...

Page 439: ...On some VIVID E9s the openings for the USB connectors have EMC fingers attached on the right and left sides of each opening ref Figure 8 119 Be careful to not loose these EMC fingers when you remove t...

Page 440: ...e USB Connector Board to the LCD plastic Cover 7 Carefully separate the Frame w LCD and Touch Screen from the LCD plastic Cover 8 Carefully separate the Frame w LCD and Touch Screen and the Upper Beze...

Page 441: ...Upper Bezel 3 Install the fixing screws NOTE Be careful so you don t bend the EMC fingers see Figure 8 122 when installing the LCD Cover 4 Install the LCD Cover 5 Install the five 5x fixing screws 6 I...

Page 442: ...ION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR...

Page 443: ...LTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 8 Replacement procedures 8 109 8 6 9 1 Preparations cont d 4 Remove five OP Panel Knobs see Figure 8 124 Figure 8 124 Remove five...

Page 444: ...k cover by removing the three 3x fixing screws Phillips 1 and the two screws at the D SUB connector 3 16 inch nut driver see Figure 8 125 Operator Panel Upper seen from the rear side on page 8 110 3 L...

Page 445: ...of the connector Disconnect the connector right side in Figure 8 126 5 Disconnect the plug on the black cable from the connector on the Main Controller board NOTE The Main Controller Board does not c...

Page 446: ...hese steps to install the Main Controller Board 1 Position the Main Controller Board with the USB board plugged in so it aligns with the fixing screw holes in the LCD frame 2 Install the five 5x screw...

Page 447: ...CTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT F...

Page 448: ...THCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 8 114 Section 8 6 Top Console Parts Replacement 8 6 10 1 Preparations cont d 5 Remove the Operator Panel Upper 6 Remove the Main Controller...

Page 449: ...ce with the solder side up The USB Connector Board is plugged into the Main Controller Board 2 Pull the USB Connector Board upwards to disconnect it from the Main Controller Board 8 6 10 3 USB Connect...

Page 450: ...STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENE...

Page 451: ...fixing screws 3 Remove the High Voltage Inverter Board with Cable 8 6 11 3 High Voltage Inverter Board with Cable installation procedure 1 Align the board s fixing holes to the respective holes on the...

Page 452: ...oval The Upper Bezel is what is left when you have removed the Frame w LCD and Touch Screen Use the procedure in 8 6 8 2 Remove the Frame w LCD and Touch Screen on page 8 105 to remove the Upper Bezel...

Page 453: ...7 Remove the Bulkhead Board 8 Remove the Bulkhead Plate WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN...

Page 454: ...Bulkhead Plate so it aligns with the holes for the fixing screws 3 Install the seven fixing screws as described below The two upper screws are M6 x 20 Torque 8 5 Nm The next screw is M6 x 45 Torque 8...

Page 455: ...own the VIVID E9 2 Disconnect the Mains Power Cable from the wall outlet 3 Separate and remove the Storage Tray WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF TH...

Page 456: ...s Holder 8 6 15 Knobs for Encoders and Slidepots replacement 8 6 15 1 Tools No tools are needed for this procedure 8 6 15 2 Knobs for Encoders and Slidepots removal procedure NOTE If you are going to...

Page 457: ...E MANUAL Chapter 8 Replacement procedures 8 123 8 6 15 3 Knobs for Encoders and Slidepots installation procedure Install the knobs one by one Refer to the illustration above for the correct position f...

Page 458: ...ISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENERGY CONTROL AND P...

Page 459: ...the screws in the next step 2 The screws are available from the underside of the Operator Panel tray Unscrew and remove the four screws used for fixing the Operator Panel assembly to the Operator Pane...

Page 460: ...4 On the Bulkhead Bracket loosen the ESD wire from the Lower Panel 5 Pull and lift the Operator Panel assembly up and away Be careful to not destroy the fingers on each side of the Operator Panel Lowe...

Page 461: ...g brackets points as illustrated below Since this feature is for the EMI compatibility the braid have to be exposed under the bracket and the cables have to be well locked by the bracket 4 Install the...

Page 462: ...S WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR TH...

Page 463: ...n the correct position and are positioned on different locations on the top and the bottom sides of the ring When used for the Vivid E9 install it so the fixing screw holes on the Fixing Ring allign w...

Page 464: ...URSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPM...

Page 465: ...ower with the front up 3 Unscrew and remove 4 screws 4 Remove the Encoder Board 8 6 18 3 Encoder Board installation procedure 1 Position the Encoder board so the holes for the fixing screws aligns 2 I...

Page 466: ...NS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROS...

Page 467: ...h Board with Elastomer removal 1 Disconnect the cable to the Encoder Board 2 Remove the 26 fixing screws 3 Remove the Lower Switch Board with Elastomer Store it in an ESD safe place Figure 8 146 Lower...

Page 468: ...h Board with Elastomer installation 1 Verify that all switch cups are in their correct positions 2 Position the Lower Switch Board with Elastomer in its correct position 3 Install the 26 fixing screws...

Page 469: ...48 Lower Bezel WARNING WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNEC...

Page 470: ...ion on a table or other place where you can keep them until you start the reinstallation If you are going to install a new Lower Bezel you should move the Switch Cups over to the new Lower Bezel one b...

Page 471: ...in Figure 8 149 as a reference NOTE THE ILLUSTRATION BELOW IS A MIRRORED ILLUSTRATION USE IT FOR REFERENCE WHEN INSTALLING THE SWITCH CUPS 2 Install the Lower Switch Board with Elastomer 3 Install the...

Page 472: ...RATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONN...

Page 473: ...securing the Main Cable 2 Remove the 5 remaining screws securing the Bulkhead Bracket 3 Remove the Bulkhead Bracket 4 From underneath the Frame pull the Main Cables through the Frame opening to free t...

Page 474: ...or Panel Frame 2 Install the Upper Frame 3 Install the plastic Plate Washers on both side of the Bulkhead Bracket 4 Position the Bulkhead Bracket 5 Install the 5 screws to secure the Bulkhead Bracket...

Page 475: ...strated in Figure 8 153 The Trackball USB Cable is used on the Operator Panel Lower The USB Video Board Flex Cable and the HV Inverter Cable are both used on the Operator Panel Upper Figure 8 153 The...

Page 476: ...USB Video Board Flex Cable The USB Video Board Flex Cable is the connection between the USB Video Board and the LCD Touch Display 1 Remove the Rear Cover on the Operator Panel Upper 2 Disconnect both...

Page 477: ...wer 8 6 22 7 Disconnect the HV Inverter Cable The HV Inverter Cable is the cable between the HV Inverter board and the Main Controller board 1 Remove the Rear Cover on the Operator Panel Upper 2 Disco...

Page 478: ...UITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FO...

Page 479: ...ocedure The Alphanumeric Keyboard is installed on a rail below the Operator Panel Follow the steps below to remove the Alphanumeric Keyboard assembly 1 Disconnect the cable plug that is connected to t...

Page 480: ...rated in Figure 8 156 Alphanumeric Keyboard seen from below on page 8 145 4 Move the Alphanumeric Keyboard assembly so it un gage then lift it away 8 6 23 4 Alphanumeric Keyboard assembly installation...

Page 481: ...D PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING...

Page 482: ...has two holes where you can access two of the fixing tabs to release them 2 Insert a thin screwdriver in one of the provided holes and use it to release the tabs buy pressing as illustrated with the a...

Page 483: ...yboard using the current screw and two additional washers tooth 3 2mm Use one washer under and one washer over the cable shoe Torque 2 3 Nm 2 Install the A N keyboard into the A N Keyboard Bottom Encl...

Page 484: ...the Ground Cable C in the opening of the A N Keyboard Bottom Enclosure 4 Install the A N Keyboard Top Enclosure onto the keyboard by snapping the eight small clips on the Top Enclosure onto the keyboa...

Page 485: ...uch Panel assembly 5 Remove the Lower Operator Panel 6 Remove the Alphanumeric Keyboard Assembly CAUTION ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS...

Page 486: ...the Wagon AN Drawer Sheet Met Assembly 8 6 25 3 Wagon AN Drawer Sheet Met Assy installation 1 Position the Wagon AN Drawer Sheet Met Assy below the UI Console as illustrated in Figure 8 162 Wagon AN...

Page 487: ...t Met Assy 2 Slide the Wagon A N Drawer Sheet Met Assy out of the J Rail 3 If needed repeat this procedure for the other J Rail Assy Figure 8 164 J Rail Assy WARNING WARNING ENERGY CONTROL AND POWER L...

Page 488: ...3 Bend down the tab 4 If needed repeat the three steps above for the other J Rail Assy 5 Position the Wagon AN Drawer Sheet Met Assy below the UI Console as illustrated in Figure 8 162 Wagon AN Drawe...

Page 489: ...ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENERGY CONTROL AND POWER...

Page 490: ...ard 3 Remove the Speaker 4 Repeat steps 1 3 for the other Speaker assembly if necessary 8 6 27 3 Install the Speaker Assembly Follow these steps to install the Speakers 1 Position the Speaker 2 Instal...

Page 491: ...STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENER...

Page 492: ...ION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 8 158 Section 8 6 Top Console Parts Replacement 8 6 28 1 Preparations cont d 4 Remove five knobs see Figure 8 167 5 Remove the Upper OP Panel Figure 8 1...

Page 493: ...Unscrew and remove two screws 3 Remove the Bulkhead Board 8 6 28 3 Install the Bulkhead Board Follow these steps to install the Bulkhead Board 1 Slide the Bulkhead Board into the correct position 2 I...

Page 494: ...ve the Bulkhead board 8 Remove the Operator Panel Lower WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN...

Page 495: ...ad Plate so it aligns with the holes for the fixing screws 2 Install the seven fixing screws as described below The two upper screws are M6 x 20 Torque 8 5 Nm The next screw is M6 x 45 Torque 8 5 Nm T...

Page 496: ...been locked with Lock Tite so you may need to apply a little extra force to unscrew the screw 2 Remove the Cable Hook Twin 8 6 30 2 Install the Cable Hook Twin 1 Position the Cable Hook Twin as shown...

Page 497: ...tallation procedure 1 Position the Non Magnetic Touch Latch and fasten it with one screw 2 Install the other screw Ensure that the Non Magnetic Touch Latch works as intended WARNING WARNING ENERGY CON...

Page 498: ...e Parts Replacement 8 6 32 Handle Left Top Handle Right Top replacement 8 6 32 1 Preparations 1 Power down the VIVID E9 and disconnect the Mains Power Cable 2 Remove the buttons below the Touch Panel...

Page 499: ...165 indicates where the hatches are located 1 Use a thin screwdriver to release the hatches one by one as illustrated in Figure 8 173 Release Hatches on page 8 165 until you can remove the Handle Lef...

Page 500: ...d Handle Right Top Figure 8 174 The Palm Rest ASSY WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 V...

Page 501: ...e the Palm Rest ASSY 8 6 33 3 Palm Rest ASSY installation Follow these steps to install the Palm Rest ASSY 1 Ensure the XYZ Buttons cable runs along the lower tray cable channel so that the cable is n...

Page 502: ...lowing types of inserts are available Probe Holder Insert STD Probe Holder Insert 3D Probe Holder Softinsert Doppler 8 6 34 2 Manpower One person 1 minute 8 6 34 3 Tools No tools needed 8 6 34 4 Prepa...

Page 503: ...t procedures 8 169 8 6 35 Replacing the Gel Cup The Gel Cup is a soft insert used for storing the gel bottle on the scanner so it is easy to access during scanning For replacement see 8 6 34 Replacing...

Page 504: ...ilize the console Toggle one of the buttons marked 2 to reposition the Z vertical location of the upper console Toggle up to raise or down to lower the console Figure 8 177 Top Console adjustment cont...

Page 505: ...of the Lower Op Panel frame handle 2 Connect the cable to the XYZ Buttons Frame 3 Ensure the XYZ Buttons Frame cable runs along the Lower Op Panel cable channel so that the cable is not pinched when t...

Page 506: ...ONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTRO...

Page 507: ...ton IF Board installation procedure 1 Install the Button IF Board into the UI Frame Lower by pressing it under the small hooks 2 Ensure the XYZ Buttons Frame cable runs along the Lower Op Panel cable...

Page 508: ...er 9 Remove the Bulkhead Board 10 Remove the Bulkhead Plate 11 Remove the Upper Frame Assembly 12 Remove the Lower Frame Assembly WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SER...

Page 509: ...8 7 1 2 XY Mechanism removal procedure 1 From the rear side of the VIVID E9 remove the four 4x fixing screws 2 Remove the XY mechanism Figure 8 180 Operator Panels have been removed Figure 8 181 Remo...

Page 510: ...Lower Frame Assembly NOTE Ensure that Bumper Z enters the slide at the back before tighten the screws 4 Install the Upper Frame Assembly 5 Install the Bulkhead Plate 6 Install the Bulkhead Board 7 In...

Page 511: ...down the VIVID E9 3 Disconnect the Mains Power Cable from the wall outlet 4 Disconnect all probes and external I O cabling Figure 8 182 XY Brake Assy WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT...

Page 512: ...screw and remove the fixing screw for the XY Brake Assy 5 Remove the brake retainer screw 6 Disconnect the brake motor wire connector 7 Using a 3 mm L Allen wrench slide the short L end of the wrench...

Page 513: ...procedure cont d 8 Gently but firmly pull down on the long end of the allen wrench The brake assembly will slide out of the slot it mates with in the brake shoe NOTE Figure 8 186 shows the XY brake re...

Page 514: ...t be replaced 1 Position the XY Brake Assy into the arm You may need to turn the axle on the motor by hand to adjust the lever so it fits 2 Install the fixing screw for the XY Brake Assy 3 Connect the...

Page 515: ...k Lock removal 1 Remove four screws 2 Carefully pull out the XY Park Lock assembly 3 Disconnect the cable at the connector 4 Remove the XY Park Lock assembly 8 7 3 3 XY Park Lock installation 1 Instal...

Page 516: ...1 Remove the Upper Frame Assembly 12 Remove the Lower Frame Assembly 13 Remove the XY Frogleg Assembly 14 Remove the XYZ Control Box Figure 8 189 Z Mechanism WARNING WARNING ENERGY CONTROL AND POWER L...

Page 517: ...7 4 2 Z Mechanism removal procedure The Z Mechanism is fastened with several screws 1 Remove four 4x screws from the lower part of the Z Mechanism 2 Remove four 4x screws 3 Lift the Z Mechanism away...

Page 518: ...l two 2x M6x16 Countersunk and two 2x M6x16 screws 3 Install four 4x M6x16 lower part of unit 4 Install the XY Frogleg Mechanism 5 Install the Lower Frame Assembly 6 Install the Upper Frame Assembly 7...

Page 519: ...ure 8 193 Drive Gear Assembly WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OU...

Page 520: ...operate the Z release lever when pulling or move the Top Console slightly up or down to make the wheel disengage from the gear 8 7 5 3 Drive Gear Assembly installation procedure 1 Position the Drive...

Page 521: ...dures 8 187 8 7 5 3 Drive Gear Assembly installation procedure cont d 3 Connect the cable from the motor to the XYX Control Box 4 Power up the unit and verify that the XYZ function as intended 5 Power...

Page 522: ...re 8 197 XYZ Control Assembly WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OU...

Page 523: ...Control Assembly forwards up and away from the bracket 8 7 6 3 XYZ Control Assembly installation procedure 1 Thread the XYZ Control Assembly into position 2 Install the three screws at the top rear M6...

Page 524: ...Purpose of this section This section describes how to replace the replaceable parts in the Main Console 8 8 2 Rear Filter and handle type Bottom Filter replacement 8 8 2 1 Preparations When preparing...

Page 525: ...air filters which need to be cleaned The top air filter is located on the back of the VIVID E9 below the power cord and the bottom air filter is located underneath the VIVID E9 Follow these steps to r...

Page 526: ...the filter B Bottom Filter Remove the Filter Assembly by lowering the handle NOTE The handle for the bottom filter is located in the same location for both Filter Assemblies 4 Clean the Rear Filter a...

Page 527: ...use a clean soft brush or vacuum 6 Reinstall the rear filter and tuck the edges of the filter under the Rear Bumper and Rear Cover Reinstall Filter Cover 7 Reinstall Bottom Filter by positioning the...

Page 528: ...removal procedure Grab the Filter and remove it CAUTION ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 TURN...

Page 529: ...placement procedures 8 195 8 8 3 3 Filter installation procedure Follow these steps to install the Filter 1 Install the Filter 2 Tuck the edges of the filter under the Rear Bumper and Rear Cover 3 Ins...

Page 530: ...tab can be accessed on the right side of the VIVID E9 8 8 4 2 Preparations 1 Power down the VIVID E9 2 Disconnect the Mains Power Cable from the wall outlet 3 Kneeling down on the right side of the V...

Page 531: ...ter by rinsing with water allow it to dry completely before re installing When dry install the filter as described next 8 8 4 5 Filter installation procedure Install the filter at the bottom of the VI...

Page 532: ...ARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCEDURES 2 TURN OFF THE BRE...

Page 533: ...rd Rack Cover 2 Unscrew the thumb screws of the Card Rack Cover 3 Remove the Card Rack Cover 4 Pull the Fan Assembly out of the frame 8 8 5 3 Fan Assembly installation procedure Follow these steps to...

Page 534: ...rvice manual 2 Remove the rubber bushings that holds the old fan screens 3 Remove the old fan screen and fan screen frame 8 8 6 2 Install the new Fan Screen and Fan Screen Frame 1 Place the new fan sc...

Page 535: ...cedures 8 201 8 8 6 2 Install the new Fan Screen and Fan Screen Frame cont d 3 Hold the frame screen firmly down onto the fan 4 Press the rivet pin firmly down to secure the fan screen 5 Install the F...

Page 536: ...ONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENERGY CONTROL AND POWER LOCKO...

Page 537: ...perator Panel Upper 2 Disconnect the HDMI cable from the Bulkhead 3 Disconnect the cable from the Z switch Up Down control 4 If not already done unscrew the fixing screws on the Ground GND Clamp on th...

Page 538: ...2 03 86 0LFUR RQWUROOHU 2 3RUWV 93 86 86 86 6 1 VWE 9 5 8 2 287 21 2 6 3 5 2 2 5 9 9 8 2 5 5 5 7 7 7 5 5 5 3 1 6 1 8 2 02725 5 9 02725 175 5 219 57 56 32 5 725 255 7 21 62 7 21 1 9 19 57 5 21752 5 5...

Page 539: ...ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENERGY CONTROL AND POWER...

Page 540: ...MANUAL 8 206 Section 8 8 Main Console parts replacement 8 8 8 2 Sub Woofer removal procedure Follow these steps to remove the Sub woofer 1 Remove the six 6x hex key screws securing the Sub woofer to...

Page 541: ...in position at the base of the chassis 2 Install the six 6x hex key screws securing the Sub woofer to the chassis 3 Replace the BEP 8 8 8 4 Verification Perform the following steps to verify that the...

Page 542: ...WARNING WARNING AT LEAST TWO PERSONS ARE NEEDED WHEN REPLACING CASTERS WHEELS OR ADJUSTING BRAKES WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE...

Page 543: ...2 6 Activate the brakes 7 The VIVID E9 is now nearly balanced between one Front and one Rear Caster 8 Make the VIVID E9 rest on both Front Casters and lift the Rear Caster Put the Wooden Wedge under t...

Page 544: ...nstall the fixing bolt M12 X 40 mm Use a 10 mm HEX key Torque 130 Nm 3 Remove the Wooden Wedge 4 Roll the VIVID E9 off the Bevel Edged Board 5 To replace the other Rear Caster repeat all the steps sta...

Page 545: ...t Rest Bumper WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCED...

Page 546: ...igure 8 212 6 The VIVID E9 is now nearly balanced between one Front and one Rear Caster 7 Make the VIVID E9 rest on both Rear Casters and lift the Front Caster 8 Put the Wooden Wedge under the chassis...

Page 547: ...the fastening screw 2 Install the three fixing screws M8 x 20 mm Torque 20 5 Nm 3 Remove the Wooden Wedge 4 Roll the VIVID E9 off the Bevel Edged Board 5 To replace the other Front Caster repeat all...

Page 548: ...ront Cover Figure 8 214 Pedal Mechanism Assembly WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOL...

Page 549: ...Torque 20 5 Nm 2 Connect the rods by snapping on the ball joints 3 Verify that the pedals work as intended a Push down brake pedal from free position release it and verify the pedal to return quickly...

Page 550: ...x key screws from below 1 Unscrew and remove the four screws 2 Remove the Brake Pedal WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EX...

Page 551: ...the four fixing screws 8 9 5 Pedal Release replacement This pedal placed in the middle is fixed with three screws With this in mind please refer to 8 9 4 Brake Pedal replacement on page 8 216 for repl...

Page 552: ...VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WA...

Page 553: ...CARD POSITION ABBREVIATION CARD NAME TOTAL NUMBER OF CARDS PER SYSTEM BT 09 BT 11 BT 12 SW v108 x x SW v110 0 0 to v110 1 3 SW v110 1 4 or higher SW v112 0 x 1 RLY RELAY BOARD 1x 1x 1x 1x 2 GRX64 ANA...

Page 554: ...11 DRX DIGITAL RECEIVER BOARD 12 GFI GLOBAL RADIO FREQUENCY INTERFACE 1x 1x 1x 1x N A FRONT PLANE XD BUS FOR GTX TLP3 THE FRONT PLANE BOARDS CONNECT TO THE BACK OF THE RELAY BOARD THE GTX BOARD S AND...

Page 555: ...TEST POINTS TO VERIFY ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER IS TURNED OFF W...

Page 556: ...s used to ground the PCIe GFI cable to the FEP cover 1 Unscrew the fixing screw for the cover for the PCIe GFI cable 2 Remove the cover for the PCIe cable 3 Unplug the PCIe Cable 4 Unscrew the FEP Cov...

Page 557: ...ighten the finger screws 1 Install the FEP Cover and fasten it with the ten fixing screws Use your fingers to tighten the screws 2 Plug in the PCIe cable in the GFI board The connector on the GFI card...

Page 558: ...DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 8 224 Section 8 10 Front End Processor FEP Card Cage parts replacement 8 10 3 Front Plane XD BUS replacement Figure 8 219 Front Plane cards FRONT...

Page 559: ...ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER IS TURNED OFF WHEN THE CORD IS STILL...

Page 560: ...teps to remove the Front Plane Boards NOTICE Read through the removal and installation steps completely before performing Perform the steps carefully to avoid damage to the Front Plane Board connector...

Page 561: ...oard and GTX Board s are returned to proper position before installing the Front Plane Board 4 Repeat the same also for the RX Board 5 Holding the upper and lower edges of the upper Front Plane Board...

Page 562: ...oards 2 Holding the upper and lower edges of the board with both hands carefully install the lower Front Plane Board Ensure that you do not bend any of the connector pins during the installation Be su...

Page 563: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 8 Replacement procedures 8 229 8 10 4 Relay Board RLY replacement Figure 8 220 Relay Board location RELAY BOARD...

Page 564: ...NO TEST POINTS TO VERIFY ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER IS TURNED OFF...

Page 565: ...the Relay Board with the rails and push it Push the card to the left so it enters correct in the mechanical slide before you push it in position so it is seated in the Back Plane connectors 2 Careful...

Page 566: ...ARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 8 232 Section 8 10 Front End Processor FEP Card Cage parts replacement 8 10 5 Receiver Board GRX replacement Figure 8 221 GRX Boards location...

Page 567: ...O TEST POINTS TO VERIFY ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER IS TURNED OFF...

Page 568: ...on nearest the GFI the 64 channel board GRX64 is positioned in the left hand position Follow these steps to install the GRX Board 1 Carefully align the GRX Board with the rails and push it in till it...

Page 569: ...35 8 10 6 Transmitter Board GTX replacement 8 10 6 1 Overview Two different GTX board models have been used GTX TLP 3 0 This board has 64 TX channels GTX TLP192 This board has 192 TX channels NOTE Whe...

Page 570: ...RIFY ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER IS TURNED OFF WHEN THE CORD IS ST...

Page 571: ...HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 8 Replacement procedures 8 237 8 10 6 2 Preparations cont d 8 Remove the FEP Cover 9 Carefully remove the two Front Plane Boar...

Page 572: ...10 6 4 GTX Board installation procedure Follow these steps to install the GTX Board NOTE If a GTX TLP192 is used it must be positioned in the right most TX slot 1 Carefully align the GTX Board with t...

Page 573: ...O TEST POINTS TO VERIFY ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER IS TURNED OFF...

Page 574: ...Board installation procedure Follow these steps to install the DRX Board 1 Carefully align the DRX Board with the rails and push it in till it is seated in the Back Plane connectors 2 Install the DRX...

Page 575: ...F THE POWER PLUG 5 WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO TEST POINTS TO VERIFY ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE...

Page 576: ...rd 1 Carefully align the GFI Board with the rails and push it in till it is seated in the Back Plane connectors 2 Install the FEP Cover 3 Install the FEP Cover s fixing screws 4 Plug in the PCIe GFI C...

Page 577: ...ISCHARGE AS THERE ARE NO TEST POINTS TO VERIFY ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF T...

Page 578: ...E AS THERE ARE NO TEST POINTS TO VERIFY ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWE...

Page 579: ...and Service settings 4 If possible wipe the HDD partitions as described here Reload the system software from DVD Select A to overwrite all content on HDD 5 Power down the VIVID E9 6 Disconnect the Ma...

Page 580: ...to the top of the BEP WARNING WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD...

Page 581: ...ger screws on the top of the BEP Cover On newer BEP5 there are three Phillips screws one on the top and one on each side of the BEP Cover On older BEP5 there is one finger screw at the top of the BEP...

Page 582: ...BEP is a BEP6 you can move over the Graphics Adapter If the BEP you removed was a BEP5 and the new BEP is a BEP5 you can move over the Graphics Adapter If the BEP you removed was a BEP5 and the new BE...

Page 583: ...e on each side of the BEP Cover On older BEP5 there is one finger screw at the top of the BEP Cover 5 Replace the Printer Bracket at the top of the BEP Be sure the lip on the underside of the bracket...

Page 584: ...VIVID E9 SERVICE MANUAL 8 250 Section 8 11 Back End Processor BEP parts replacement 8 11 1 6 Calibration and adjustments Calibrate the Front End A D converters as described in Section 6 6 DC Offset C...

Page 585: ...Cover aside WARNING WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTIO...

Page 586: ...of the BEP6 Power Board Assembly 2 Remove the two screws that secure the BEP6 Power Board Assembly 3 Slide the BEP6 Power Board Assembly out of the BEP You will need to apply some force use the two br...

Page 587: ...P frame 3 Plug in the cable plug in the connector near the top of the BEP6 Power Board Assembly 4 Install the BEP Cover as described in these steps a Insert the base of the BEP Cover inside the base o...

Page 588: ...er Supply 2 Gently insert a flat screwdriver between the power supply and the BEP frame to create enough space to get hold on the power supply 3 Pull the Power Supply toward you while gently rocking t...

Page 589: ...pply to the BEP frame 3 Install the BEP Cover as described in these steps a Insert the base of the BEP Cover inside the base of the BEP Be sure the bottom lip of the BEP Cover rests inside the BEP b T...

Page 590: ...OLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING BEFORE YOU DISPOSE OF THE HARD DRIVE MAKE SURE YOU REMOVE ALL PATIENT DATA FROM THE HARD DRIVE GIVEN THAT THE...

Page 591: ...ains Power Cable from the wall outlet 6 Disconnect all external cables probes ECG physio ethernet audio video USB 7 Remove the Left Side Cover 8 Remove the BEP Cover a Loosen the BEP Cover s fixing sc...

Page 592: ...3 2 BEP6 Hard Disk Drive HDD removal procedure Follow these steps to remove the HDD 1 Disconnect the cables connecting the HDD 2 Remove the two screws that secures the HDD Bracket to the BEP s frame...

Page 593: ...top of the BEP Cover toward the VIVID E9 c Install the fixing screw s for the BEP Cover On BEP6 there are two finger screws on the top of the BEP Cover On newer BEP5 there are three Phillips screws on...

Page 594: ...e HDD removal procedure Follow these steps to remove the HDD 1 Remove the screw that secures the HDD bracket 2 Swing the HDD bracket out 3 Disconnect the cables connecting the HDD 4 Lift the bracket u...

Page 595: ...g screw s for the BEP Cover On BEP6 there are two finger screws on the top of the BEP Cover On newer BEP5 there are three Phillips screws one on the top and one on each side of the BEP Cover On older...

Page 596: ...BT 12 with BEP5 without 4D Prolink Add 2 card NOTE On BEP6 without 4D the graphics circuits Intel HD Graphics and Intel Graphics Media Accelerator are included on the BEP s motherboard A Video Bypass...

Page 597: ...SENSITIVE EQUIPMENT WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT...

Page 598: ...e the Left Side Cover 5 Remove the BEP Cover a Loosen the BEP Cover s fixing screw s On BEP6 there are two finger screws on the top of the BEP Cover On newer BEP5 there are three Phillips screws one o...

Page 599: ...rocedure 2 Unscrew the fixing screws from the connector at top of the Graphics Adapter If DVR is installed the flex between the Graphics Adapter and the DVR board uses finger screws Without DVR the fl...

Page 600: ...ollow these steps to install the new Graphics Adapter 1 Plug in the new Graphics Adapter 2 Insert and fasten the fixing screw 3 Plug in the cables you removed earlier Connections if DVR is installed i...

Page 601: ...nside the BEP b Tilt the top of the BEP Cover toward the VIVID E9 c Install the fixing screw s for the BEP Cover On BEP6 there are two finger screws on the top of the BEP Cover On newer BEP5 there are...

Page 602: ...described in these steps a Insert the base of the BEP Cover inside the base of the BEP Be sure the bottom lip of the BEP Cover rests inside the BEP b Tilt the top of the BEP Cover toward the VIVID E9...

Page 603: ...EP s motherboard 2 If a locking mechanism is installed on the connector on the BEP s motherboard push the plastic slider down to release the Graphics Adapter 3 Unplug and remove the Graphics Adapter S...

Page 604: ...Graphics Adapter Add2 3 DVR Board stipulated position to the left installed to the right 4 DVI Out cable 5 DVI Flex Main cable 6 SATA cable Don t disconnect in this procedure Figure 8 238 Cables on th...

Page 605: ...Quadro FX1800 Nvidia Quadro 2000D BEP5 Figure 8 239 Cables on the Nvidia Quadro FX1800 Nvidia Quadro 2000D card BEP5 1 I O Unit 2 Graphics Adapter Nvidia Quadro FX1800 Nvidia Quadro 2000D 3 DVR Board...

Page 606: ...the BEP Cover toward the VIVID E9 c Install the fixing screw s for the BEP Cover On BEP6 there are two finger screws on the top of the BEP Cover On newer BEP5 there are three Phillips screws one on t...

Page 607: ...NECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT...

Page 608: ...s one finger screw at the top of the BEP Cover b Tilt the top of the BEP Cover away from the BEP c Set the BEP Cover aside NOTE BEP5 Remove the HDD to get better access Ref 8 11 3 Hard Disk Drive HDD...

Page 609: ...stall the BEP Cover as described in these steps a Insert the base of the BEP Cover inside the base of the BEP Be sure the bottom lip of the BEP Cover rests inside the BEP b Tilt the top of the BEP Cov...

Page 610: ...AS THERE ARE NO TEST POINTS TO VERIFY ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER...

Page 611: ...d one on each side of the BEP Cover On older BEP5 there is one finger screw at the top of the BEP Cover b Tilt the top of the BEP Cover away from the BEP c Set the BEP Cover aside 6 Remove the BEP and...

Page 612: ...e thin rubber from the rubber fastner 6 Plug in the fan s plug to the connector on the motherboard 7 Install the BEP in the VIVID E9 8 You may want to power up the VIVID E9 for a short time just to ch...

Page 613: ...e 1 Power down the VIVID E9 2 Disconnect the Mains Power Cable from the wall outlet 3 Disconnect all probes and external cables 4 Remove the Left Side Cover 5 Remove the Top Cover 6 Remove the B W Pri...

Page 614: ...on the BEP6 Power Board 2 Install the J1 connector in the frame with the two screws from below and the nuts above 3 Install the HDD Bracket Ensure that the BEP Power In Cable is routed above the HDD...

Page 615: ...4 Remove the Left Side Cover 5 Remove the BEP Cover a Loosen the BEP Cover s fixing screw s On BEP6 there are two finger screws on the top of the BEP Cover On newer BEP5 there are three Phillips scre...

Page 616: ...4 Remove the Left Side Cover 5 Remove the BEP Cover a Loosen the BEP Cover s fixing screw s On BEP6 there are two finger screws on the top of the BEP Cover On newer BEP5 there are three Phillips scre...

Page 617: ...e Cable 1 Connect the two connectors Power and USB to the Patient I O 2 Connect the other end of the cable to CN USB11 on the BEP s motherboard 3 Install the BEP Power Board 4 Install the BEP Cover 5...

Page 618: ...ws on the top of the BEP Cover On newer BEP5 there are three Phillips screws one on the top and one on each side of the BEP Cover On older BEP5 there is one finger screw at the top of the BEP Cover b...

Page 619: ...ws on the top of the BEP Cover On newer BEP5 there are three Phillips screws one on the top and one on each side of the BEP Cover On older BEP5 there is one finger screw at the top of the BEP Cover b...

Page 620: ...lex es 4D with DVR Install the DVI to Samtec Jumper Flex between CN DV12 on the MBD and DVI OUT on the DVR board 4D without DVR Install the DVI to Samtec Jumper Flex between CN DV12 on the MBD and the...

Page 621: ...he top of the BEP Cover b Tilt the top of the BEP Cover away from the BEP c Set the BEP Cover aside 6 Remove the BEP and place it on a table Related information 4 2 2 Power shut down on page 4 8 8 5 2...

Page 622: ...d one on each side of the BEP Cover On older BEP5 there is one finger screw at the top of the BEP Cover b Tilt the top of the BEP Cover away from the BEP c Set the BEP Cover aside 6 Related informatio...

Page 623: ...ne on the top and one on each side of the BEP Cover On older BEP5 there is one finger screw at the top of the BEP Cover b Tilt the top of the BEP Cover away from the BEP c Set the BEP Cover aside 6 Re...

Page 624: ...s When preparing for the replacement you must perform the following steps 1 Power down the VIVID E9 2 Disconnect the Mains Power Cable from the wall outlet 3 Disconnect all probes 4 Remove the Left Si...

Page 625: ...ow remove the BEP6 to GFI and Card Rack Backplane Cable 8 11 15 3 BEP6 to GFI and Card Rack Backplane Cable installation 1 Connect the Boundary Scan Cable to the FEP Backplane Connector 2 Connect the...

Page 626: ...VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WA...

Page 627: ...er supply 2 Remove the three screws that secure the Main Power Supply 3 Ease the Main Power Supply away from the fang bracket 4 Pull slide the entire Main Power Supply backwards until the connectors t...

Page 628: ...t the top of the Main Power Supply toward the rear of the Card Cage until the Main Power Supply is vertical 2 Slide the Main Power Supply forward to seat the connectors on the Card Cage 3 Engage the t...

Page 629: ...ED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER NEAR THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT WARNING WARNING ENERGY CONTROL AND POW...

Page 630: ...remove the Patient I O module 1 Inside the BEP there is an opening for the Power and USB cables to the Patient I O module Disconnect the Power cable connector and the USB cable connector from the Pati...

Page 631: ...rests inside the BEP b Tilt the top of the BEP Cover toward the VIVID E9 c Install the fixing screw s for the BEP Cover On BEP6 there are two finger screws on the top of the BEP Cover On newer BEP5 t...

Page 632: ...Cover aside WARNING WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTIO...

Page 633: ...w these steps to remove the BEP6 Side I O Board Assembly 1 Disconnect the cables connected to the face of the BEP6 Side I O Board Assembly 2 Remove the four screws connecting the BEP6 Side I O Board A...

Page 634: ...4 Reconnect the I O cabling 5 Install the BEP Cover as described in these steps a Insert the base of the BEP Cover inside the base of the BEP Be sure the bottom lip of the BEP Cover rests inside the B...

Page 635: ...ard Assembly 1 Disconnect the cables inside the BEP that connect to the I O Flat cable from the Video Graphics board 2 Remove the five screws connecting the I O to the BEP 3 Reach inside the BEP frame...

Page 636: ...e BEP to the I O Board Flat cable from the Video Graphics board 5 Reconnect the I O cabling 6 Install the BEP Cover as described in these steps a Insert the base of the BEP Cover inside the base of th...

Page 637: ...VIVID E9 2 Disconnect the Mains Power Cable from the wall outlet 3 Disconnect all probes and external I O cabling 4 Remove the Side Covers 5 Remove the Top Cover WARNING WARNING ENERGY CONTROL AND POW...

Page 638: ...ring the left side of the DVD R W drive 5 Disconnect the cables to the DVD R W drive 6 Slide the DVD R W drive out the front of the VIVID E9 7 If Power Connection is different on new and old DVD R W d...

Page 639: ...ELECTRICAL SHOCK VIVID E9 MUST BE TURNED OFF AVOID ALL CONTACT WITH ELECTRICAL CONTACTS CONDUCTORS AND COMPONENTS ALWAYS USE NON CONDUCTIVE HANDLES DESIGNED FOR THE REMOVAL AND REPLACEMENT OF ESD SENS...

Page 640: ...sen the BEP Cover s fixing screw s On BEP6 there are two finger screws on the top of the BEP Cover On newer BEP5 there are three Phillips screws one on the top and one on each side of the BEP Cover On...

Page 641: ...screws from the connector at top of the DVR The flex between the Graphics Adapter and the DVR board uses finger screws The flex from the motherboard is fixed with Hex screws 3 Disconnect the audio ja...

Page 642: ...gital Video Recorder from a BEP5 1 Locate the DVR Board behind the HDD 2 Disconnect the audio jack SATA DVI flex cables where they attach to the DVR Board 3 Remove the DVR Board from the motherboard s...

Page 643: ...DVR Board in the correct slot 2 Insert and fasten the fixing screw 3 Plug in the cables you removed earlier Connections if the DVR Board is installed in a VIVID E9 with 4D Connections if the DVR Boar...

Page 644: ...ure the bottom lip of the BEP Cover rests inside the BEP b Tilt the top of the BEP Cover toward the VIVID E9 c Install the fixing screw s for the BEP Cover On BEP6 there are two finger screws on the t...

Page 645: ...these steps a Insert the base of the BEP Cover inside the base of the BEP Be sure the bottom lip of the BEP Cover rests inside the BEP b Tilt the top of the BEP Cover toward the VIVID E9 c Install th...

Page 646: ...ING ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9 WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCEDURES 2 TURN OFF THE BREAKE...

Page 647: ...hite Digital Graphic Printer 1 Slide the printer into the printer bracket until the face is flush with the VIVID E9 2 Tighten the printer bracket wing nut to secure the printer in the printer bracket...

Page 648: ...l please refer to the Printer Driver Installation Manual Direction Number GA294652 revision 3 or higher for installation and setup instructions General installation instructions 1 Connect the USB cabl...

Page 649: ...t the GFI PCIe Cable from the top of the BEP 2 Disconnect the GFI PCIe Cable from the GFI board as described in these steps a On the Front End Card Rack remove the cover for the GFI PCIe Cable connect...

Page 650: ...end of the GFI PCIe Cable to the connector on top of the BEP Ensure that the GFI PCIe Cable is kept out of the drawed box area in Figure 8 259 as there is a risk that the Z Mechanism damage the cable...

Page 651: ...IVID E9 9 6 9 6 Covers and Bumpers 9 15 9 7 Top Console parts 9 18 9 8 XYZ parts 9 50 9 9 Main Console parts 9 52 9 10 Casters Wheels parts 9 54 9 11 Front End Processor FEP Card Rack parts 9 56 9 12...

Page 652: ...2 Definitions of Left Right Front and Back Section 9 2 Definitions of Left Right Front and Back The Figure below illustrates what is Left Right Front and Rear or Back of the VIVID E9 Figure 9 1 Defini...

Page 653: ...ge 5 52 OP OPERATOR PANEL BEPY3 SEE DESCRIPTION IN 5 9 2 Introduction on page 5 52 PC PERSONAL COMPUTER Back End Processor BEP5 SEE DESCRIPTION IN 5 9 2 Introduction on page 5 52 PCIe PCI Express BEP6...

Page 654: ...er v112 0 x or higher v112 1 x v112 1 x v112 1 x GA200890 BEP5 w 4D Nvidia v104 3 x v112 0 x or higher v112 1 x GA000950 Vivid E9 100 230V 4D Expert Option 19 LCD GB200001 BEP6 w 4D v104 3 4 or higher...

Page 655: ...00800 BEP5 w 4D Nvidia v104 2 x v104 1 x v110 0 x GA000830 VIVID E9 100 230V 2D 17 LCD GA200804 VE9 Card Rack Complete w MLA4 GB200002 BEP6 wo 4D v104 3 x v110 1 12 v112 1 x GA200900 or GA200805 BEP5...

Page 656: ...ME PART NUMBER DESCRIPTION QTY CRU FRU System Software 1 VIVID E9 SYSTEM SOFTWARE BEP6 GB200015 SYSTEM SOFTWARE INSTALLATION DVD GHOST FOR BEP6 LATEST SYSTEM SOFTWARE 2012 SEP 14 v104 3 4 1 N Y 2 VIVI...

Page 657: ...1 N Y 9 VIVID E9 APPLICATION SOFTWARE v 112 1 0 GA200965 10 APPLICATION SOFTWARE INSTALLATION CD v 112 1 0 The CD is labeled Version 112 Revision 1 0 1 N Y 10 VIVID E9 APPLICATION SOFTWARE v 112 0 7 G...

Page 658: ...atch GA200973 02 FOR THE FOLLOWING SYSTEM SOFTWARE VERSION v104 3 x The patches must be installed after the installation of system software and application software This security patch CD includes the...

Page 659: ...em software v 104 3 2 for BEPY1 BEP5 1 N Y 4 GA200895 02 Vivid E9 System SW 104 3 1 for BEPY1 BEP5 1 N Y 5 GA200895 01 Vivid E9 System SW 104 3 0 for BEPY1 BEP5 1 N Y 6 VIVID E9 SYSTEM SOFTWARE GA2008...

Page 660: ...10 1 1 1 N Y 4 GA200956 03 Service Part Application software v 110 1 3 1 N Y 5 GA200956 04 Service Part Application software v 110 1 4 1 N Y 6 GA200956 05 Service Part Application software v 110 1 5 1...

Page 661: ...em software and application software This security patch CD includes the following patches MS11 006 Vulnerability in Windows Shell Graphics Processing Could Allow Remote Code Execution MS11 019 Vulner...

Page 662: ...BEP5 BIOS Load CD Release version A563E121 GA200725 FOR VIVID E9 BEP5 1 N Y 2 SET SERIAL NUMBER GA200649 Latest version per 2010 11 03 v1 2 GA200649 REV 3 1 N Y 3 PRINTER INSTALLER GA200652 Printer Dr...

Page 663: ...System Software v 104 0 2 1 N Y Application Software 5 VIVID E9 APPLICATION SOFTWARE GA200355 Application software installation CD Was delivered with new VIVID E9s Application software version v108 1...

Page 664: ...re WINDOWS PATCHES INCLUDED MS08 067 MS09 001 MS09 22 MS10 007 MS10 020 KB967715 MS10 046 MS10 054 Updated 2011 SEP 01 1 N Y 16 Patch for 108 1 6 GA200814 01 Install this patch after application softw...

Page 665: ...Table 9 12 COVERS VIVID E9 sheet 1 of 3 ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 COVER FRONT GA307022 FRONT COVER 1 N Y 2 PLATE CONNECTORS W GUIDE GA307056 1 N Y 3 COVER LEFT ASM GA200107...

Page 666: ...7 Door I O PANEL GA307046 1 N Y 8 CABLE HOOK GA307047 2 N Y 9 BUMPER REAR GA307008 1 N Y 10 COVER FILTER GA307515 FILTER COVER 1 TBD Y 11 BUMPER FOOTREST ASSY GA200005 FRONT BUMPER FOOTREST 1 N Y 12...

Page 667: ...FT TOP METAL FINISH GA307233 1 N Y 18 HANDLE RIGHT TOP METAL FINISH GA307235 1 N Y 19 BULKHEAD COVER GA307061 1 N Y 20 BW PRINTER FILLER BOX 5309088 USED ON UNITS WITHOUT INTERNAL B W PRINTER 1 N Y 21...

Page 668: ...9 7 1 LCD Monitor parts Table 9 13 LCD parts ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 17 INCH MONITOR AND PARTS 1 LCD MONITOR 17 GA200550 17 INCH LCD MONITOR 1 N Y 2 LCD ARM FOR 17 LCD 51991...

Page 669: ...IPTION QTY CRU FRU OPERATOR PANEL LOWER 5 1 OPERATOR PANEL LOWER 5 GB200030 This is the latest version of the Operator Panel Lower On this OP the Top Locking Plate including the Rubber Dust Filtering...

Page 670: ...uter side of the Operator Panel for cleaning purposes do not need to get inside the panel The new Laser Trackball Assembly can be used with old versions of Operator Panel Upper 1 N N Table 9 14 OP 5 p...

Page 671: ...R DESCRIPTION QTY CRU FRU OPERATOR PANEL UPPER 4 1 OPERATOR PANEL UPPER 4 GA200865 Replaces GA200822 GA200720 and GA200365 1 N Y 2 UPPER BEZEL GA200437 This part is compatible with all OP versions 1 N...

Page 672: ...ator Panel Cable Kit for Vivid E9 and Logiq E9 on page 9 34 1 N Y 7 USB Connector Board for Upper OP Panel 5207000 41 1 N Y OPERATOR PANEL LOWER 4 8 OPERATOR PANEL LOWER 4 GA200953 Phased out of manuf...

Page 673: ...R 3 GA200822 OBSOLETE OBSOLETE Use GA200865 as the replacement 1 N Y 2 UPPER BEZEL GA200437 This part is compatible with all OP versions 1 N Y 3 Frame w LCD and TouchScreen GA200439 GA200439 is going...

Page 674: ...for 5207000 30 on OP 3 Install with the old video cable on OP 3 7 Main Ctrl Bd w USB Video Bd Cable 5207000 30 This part is going obsolete If unavailable you may use 5207000 40 as a replacement but in...

Page 675: ...9 SERVICE MANUAL Chapter 9 Renewal parts 9 25 OPERATOR PANEL LOWER 3 9 OPERATOR PANEL LOWER 3 GA200823 OBSOLETE OBSOLETE Use GA200953 as the replacement 1 N Y Table 9 16 OP 3 parts cont d sheet 3 of 3...

Page 676: ...ION QTY CRU FRU OPERATOR PANEL UPPER 2 1 OPERATOR PANEL UPPER 2 GA200720 OBSOLETE OBSOLETE Use GA200865 as the replacement 1 N Y 2 UPPER BEZEL GA200437 This part is compatible with all OP versions 1 N...

Page 677: ...E9 SERVICE MANUAL Chapter 9 Renewal parts 9 27 5 OPERATOR PANEL LOWER 2 GA200755 OBSOLETE OBSOLETE Use GA200953 or GB200030 as the replacement 1 N Y Table 9 17 OP 2 specific parts cont d sheet 2 of 2...

Page 678: ...PERATOR PANEL UPPER GA200365 OBSOLETE OBSOLETE Use GA200865 as the replacement 1 N Y 2 UPPER BEZEL GA200437 This part is compatible with all OP versions 1 N Y 3 OPERATOR PANEL LOWER 1 GA200394 1 N Y 4...

Page 679: ...29 5 USB Connector Board 2 GA200717 NOTE This board is compatible with both GA200448 and GA200718 1 N Y 6 USB Connector Board GA200441 NOTE This board is only compatible with GA200442 1 N Y Table 9 1...

Page 680: ...Aux Board 5207000 50 Replaces 5207000 40 when it becomes obsolete 1 N Y 2 Main board Video board Video Cable USB Aux Board 5207000 40 For OP 4 May be used as a replacement for 5207000 30 on OP 3 Inst...

Page 681: ...rd and Cables GA200448 OBSOLETE OBSOLETE Use 5207000 40 as the replacement Note Use the old LCD cable NOTE GA200448 is compatible with the USB Connector Board models GA200441 and GA200717 1 N Y 7 USB...

Page 682: ...WITH CABLE GA200442 1 N Y PCB FRUs on the Lower OP 10 Lower Switch Board with Elastomer GA200440 1 N Y 11 Encoder Board GA200443 1 N Y 12 Encoder w Push Button 066E2306 Replaces GA200447 5 N Y 13 BUT...

Page 683: ...ce Trackball Switch Cable for connecting the switches located around the Trackball to the USB interface Screws washers for fixing the Trackball to the Lower Operator Panel 1 N Y 2 Dust Gasket and Fixi...

Page 684: ...le For connection between Trackball and Main Controller Board Trackball Switch Cable First version For connecting the switches located around the Trackball to the USB interface Trackball Switch Cable...

Page 685: ...nel Cable Kit for Vivid E9 and Logiq E9 cont d sheet 2 of 2 ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU Trackball USB Cable For connection between Trackball and Main Controller Board USB Video...

Page 686: ...D E9 SERVICE MANUAL 9 36 Section 9 7 Top Console parts 9 7 6 Button Kits and Knobs Table 9 22 Button Kits and Knobs sheet 1 of 7 ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 Button Kit Danish...

Page 687: ...5 VIVID E9 SERVICE MANUAL Chapter 9 Renewal parts 9 37 3 Button Kit Universal English 066E3221 1 N Y 4 Button Kit Finnish 066E3229 1 N Y Table 9 22 Button Kits and Knobs cont d sheet 2 of 7 ITEM PART...

Page 688: ...ION 5 VIVID E9 SERVICE MANUAL 9 38 Section 9 7 Top Console parts 5 Button Kit French 066E3222 1 N Y 6 Button Kit German 066E3223 1 N Y Table 9 22 Button Kits and Knobs cont d sheet 3 of 7 ITEM PART NA...

Page 689: ...ION 5 VIVID E9 SERVICE MANUAL Chapter 9 Renewal parts 9 39 7 Button Kit Spanish 066E3224 1 N Y 8 Button Kit Portuguese 066E3225 1 N Y Table 9 22 Button Kits and Knobs cont d sheet 4 of 7 ITEM PART NAM...

Page 690: ...N 5 VIVID E9 SERVICE MANUAL 9 40 Section 9 7 Top Console parts 9 Button Kit Italian 066E3226 1 N Y 10 Button Kit Swedish 066E3227 1 N Y Table 9 22 Button Kits and Knobs cont d sheet 5 of 7 ITEM PART N...

Page 691: ...Y GA200270 SWITCHES ASSY FOR XY CONTROL This ASSY includes the following parts Up Down Button Board GA200285 Button XY Break Button Z Lift Bracket Buttons 2 N Y 13 UP DOWN BUTTON BOARD GA200285 Circui...

Page 692: ...VISION 5 VIVID E9 SERVICE MANUAL 9 42 Section 9 7 Top Console parts 14 Knobs Encoders and Slidepots GA200445 KNOB KIT FOR OP 1 Y Y Table 9 22 Button Kits and Knobs cont d sheet 7 of 7 ITEM PART NAME P...

Page 693: ...n be used on all VIVID E9 systems Table 9 23 Other Common Top Console parts sheet 1 of 7 ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 FRAME UI UPPER ASM GA200392 1 N Y 2 USB Socket Cover 52070...

Page 694: ...EZEL GA200437 1 N Y 7 LCD MOUNT LOCK ASSY GA200302 1 N Y 8 Frame w LCD and TouchScreen 5207000 39 1 N Y 9 Frame w LCD and TouchScreen GA200439 GA200439 is going obsolete 10 LOWER BEZEL GA200438 1 N Y...

Page 695: ...0 2 N Y 12 Cable A N Keyboard USB 12V GA200368 1 N Y 13 CABLE HOOK TWIN GA307069 2 N Y 14 PROBEHOLDER INSERT 3D GA307073 PROBEHOLDER SOFTINSERT FOR 3D 4D PROBES 1 Y Y 15 PROBEHOLDER INSERT STD GA30707...

Page 696: ...OPPLER PEDOF PROBE 1 Y Y 17 GEL CUP GA307239 1 Y Y 18 SOFT INSERT GEL CUP GA307676 1 Y Y 19 A N KEYBOARD ENCLOSURE GA200683 The A N Keyboard Encloser includes these parts A N Keyboard Enclosure Bottom...

Page 697: ...A N Keyboard Danish 066E3210 1 N Y 22 A N Keyboard Dutch 066E3211 1 N Y 23 A N Keyboard French 066E3202 1 N Y 24 A N Keyboard German 066E3203 1 N Y 25 A N Keyboard Spanish 066E3204 1 N Y 26 A N Keybo...

Page 698: ...d Norwegian 066E3208 1 N Y 29 WAGON AN DRAWER SHEET MET ASSY GA200304 1 N Y 30 J RAIL LINING J RAIL ASSY GA200544 2 N Y 31 NON MAGNETIC TOUCH LATCH PR 21P 080X1424 2 N Y 32 SPRING AN LATCH GA307643 1...

Page 699: ...lete GA200450 BOTTOM VIEW 1 N Y 35 PALM REST ASSY GA200605 PALM REST PALM REST PAD 2 1 N Y 36 PALM REST GA307063 PALM REST PAD 1 N Y 37 Bulkhead Plate Extended GA307059 1 N Y 38 BRACKET LOCK HDMI 2 GA...

Page 700: ...MBER DESCRIPTION QTY CRU FRU 1 XY MECHANISM GA200946 FROG LEG Replaces GA200036 NOTE This part requires that either the XYZ controller Part Number GA200795 or Part number GA200994 is installed 1 N Y 2...

Page 701: ...d GA200644 1 N Y 8 XYZ CONTROL ASM GA200795 Replaces GA200644 Can be used with both GA200036 and GA200946 1 N Y 9 XYZ CONTROL ASM GA200644 1 N Y 10 XY BRAKE ASSY GA200952 Replaces GA200535 NOTE This p...

Page 702: ...eels parts on page 9 54 4 COVERS SEE Section 9 6 Covers and Bumpers on page 9 15 5 BULKHEAD BOARD ASSY GA200290 1 N Y 6 Fan for VE9 cardcage 098A0093 Spare part for VE9 Fan Tray Fan Drawer 1 N Y 7 FAN...

Page 703: ...HANDLE LOCATED AT BOTTOM OF VIVID E9 5316340 2 REPLACES GA200828 1 N Y GA200828 DUST FILTER WITH NYLON STIP LOCATED AT BOTTOM OF VIVID E9 GA200828 REPLACES 5316340 5316340 11 FILTER DUST REAR GA307351...

Page 704: ...t 1 of 2 ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 Wheel cap for VE9 casters GA200245 and GA200246 GA307888 Replacement CAP with snap on locks 8 N Y 2 CASTERS FRONT W BRAKE AND LOCK GA20024...

Page 705: ...8 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 9 Renewal parts 9 55 6 PEDAL RELEASE GA307053 1 N Y 7 PEDAL DIR LOCK GA307054 1 N Y Table 9 26 Casters Wheels parts cont d sheet 2 of 2 ITEM PART NAME PART...

Page 706: ...LS WITHOUT ANALOG DOPPLER 1 N Y 4 GRX64 GA200300 RECEIVER BOARD 64 CHANNELS WITHOUT ANALOG DOPPLER 1 N Y 5 GRX128 WITH CW GB200020 NOTE This board will be phased into production and phased in as a Ser...

Page 707: ...Renewal parts 9 57 12 DRX3 BOARD MLA4 VERSION 5301040 4 DIGITAL RECEIVER BOARD 4 MLA DO NOT MIX WITH 5301040 5 3 N Y 13 GFI 2 5161631 RADIO FREQURENCY INTERFACE 1 N Y Table 9 27 Front End Processor C...

Page 708: ...R 1 N Y 4 GRX64 GA200300 RECEIVER BOARD 64 CHANNELS WITHOUT ANALOG DOPPLER 1 N Y 5 GRX128 WITH CW GB200020 NOTE This board will be phased into production and phased in as a Service Part when GA200105...

Page 709: ...A200300 becomes unavailable RECEIVER BOARD 64 CHANNELS WITHOUT ANALOG DOPPLER 1 N Y 4 GRX64 GA200300 RECEIVER BOARD 64 CHANNELS WITHOUT ANALOG DOPPLER 1 N Y 5 GRX128 WITH CW GB200020 NOTE This board w...

Page 710: ...pare Parts 9 60 9 12 3 BEP5 Spare Parts 9 62 9 12 2 BEP6 Spare Parts Table 9 30 BEP6 Spare Parts sheet 1 of 2 ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 BEP6 0 WITHOUT 4D GB200002 BACK END P...

Page 711: ...E9s with BEP6 and 4D 1 N Y 8 BEP6 X Video ByPass Board 5433408 90 1 N Y 9 Intel HD Graphics and Intel Graphics Media Accelerator BEP6 N A Included on the BEP6 s Motherboard 1 N N 10 Digital Video Reco...

Page 712: ...Software requirement v108 0 0 and higher 1 N Y 4 GRAPHICS ADAPTER 066E0361 NVIDIA Quadro 2000D Used on VIVID E9s with 4D System Software requirement v104 3 3 or higher This card will be phased in as a...

Page 713: ...d GB200003 1 N Y 14 BEP FRONT PANEL ASSEMBLY WITHOUT USB FREY 5301222 3 For 5145000 10 GA200800 and GA200805 1 N Y 15 RJ 45 Coupler and Flange Category 6 5176472 2 For all BEPs 1 N Y 16 BEP Power Supp...

Page 714: ...Supply Table 9 32 Main Power Supply ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 Main power supply Vivid E9 GA200730 03 Replaces GA200730 1 N Y 2 Vivid E9 Main Power Supply GA200730 Replaces G...

Page 715: ...ART NUMBER DESCRIPTION QTY CRU FRU 1 PATIENT IO GB200010 Replaces GA200240 1 N Y 2 PATIENT IO GA200240 PATIENT IN OUT 1 N Y 3 BEP6 0 SIDE IO BOARD ASSEMBLY 5433408 1 I O ASSEMBLY For GB200001 and GB20...

Page 716: ...066E0700 4 DVD Drive SATA LG GH20NS15 066E0700 5 DVD Interface Board 5301204 1 N Y Table 9 35 Printer B W Internal ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 B W PRINTER GA100980 066E2961 MI...

Page 717: ...DIGITAL GRAPHIC COLOR PRINTER USB EXTERNAL REPLACES SONY UP D23MD 1 Y Y 6 COLOR VIDEO PRINTER UP D23MD FOR VIVID E PRODUCTS GA100677 066E2958 OBSOLETE SONY UP D23MD DIGITAL GRAPHIC COLOR PRINTER USB...

Page 718: ...66E0429 5 HP OFFICEJET PRO K5400 KIT SPARE PART HP OfficeJet Pro K5400dn EQ100158 066E0426 Replaced by EY100153 066E0427 1 Y Y 6 COLOR LASER PRINTER 100 127 VAC ASSEMBLY SPARE PART 066E0418 OBSOLETE O...

Page 719: ...N Y 3 DVR Board See Section 9 12 Back End Processor BEP Spare Parts on page 9 60 Digital Video Disc Recorder Board is installed in BEP 4 Cable DVD Power FREY Cable K See 9 17 6 Back End Processor BEP...

Page 720: ...llation manual and BIOS software for Vivid E9 066E0661 is the complete unit without installation manual and BIOS software For more information see Iomega Ultramax Desktop Hard Drive Installation Manua...

Page 721: ...1 4 H4000ZD 1 Y N 4 MAINS POWER CABLE JAPAN 5148381 2 H4000ZK 1 Y N 5 MAINS POWER CABLE NORTH AMERICA 5148381 H4000ZA 1 Y N 6 MAINS POWER CABLE AUSTRALIA 5323129 5148381 6 H4000ZG 5323129 replaces 514...

Page 722: ...A200225 CABLE T XYZ CONTROLLER LOCK MOTOR 1 N Y 3 CABLE POWER CONTROLLER BEP FOR VIVID E9 GA200715 CABLE I J1 MAIN POWER SUPPLY XYZ CONTROLLER The upper version of the cable was introduced 2011 GA2007...

Page 723: ...LE DVD POWER FULLY SHIELDED 5266407 CABLE K J30 DVD 1 or DVD 2 2 N Y 12 CABLE DVD eSATA FULLY SHIELDED 5270048 CABLE M J2 DVD 1 or DVD 2 2 N Y 13 MAIN CABLE ASSEMBLY FREY 5272357 CABLE D FROGLEG CABLE...

Page 724: ...RATOR PANEL CABLE KIT GA200446 1 N Y 16 Cable A N Keyboard USB 12V GA200368 1 N Y 17 CABLE MAIN 5272357 CABLE D FROGLEG CABLE DVI VIDEO BEP OP PANEL 48V POWER USB6 BEP OP PANEL AUDIO ON OFF USB5 BEP O...

Page 725: ...EAD 5248610 CABLE G USB short cable inside the UI J24 P7 J25 P8 2 N Y 21 CABLE XYZ UP DOWN GA200311 CABLE U UP DOWN LOCK BUTTONS 2 N Y 22 LCD Monitor Arm Cable Kit GA200668 Powerd_USB_cable_frey L9 19...

Page 726: ...00335 CABLE B XYZ CONTROLLER INNER FROG BRAKES 2 N Y 2 BRAKE ARM LEFT RIGHT CABLE GA200334 CABLE C INNER BRAKE OUTER BRAKE 2 N Y 3 MAIN CABLE XYZ SWITCH SIGNALS SEE MAIN CABLE PART OF MAIN CABLE CABLE...

Page 727: ...les ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 MAINS POWER CABLE See Section 9 16 Mains Power Cables VIVID E9 on page 9 71 2 CABLE USB A B HIGH SPEED 5194492 2 CABLE H J4 MAIN POWER SUPPLY 1...

Page 728: ...s VIVID E9 9 17 5 Front End Processor FEP cables Table 9 46 Front End Processor FEP cables ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 PC2GFI CARD CABLE 5152290 PCI EXPRESS CABLE GFI to BEP J...

Page 729: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 9 Renewal parts 9 79 9 17 6 Back End Processor BEP cables 9 17 6 1 BEP6 cables Figure 9 2 BEP6 internal cables...

Page 730: ...CABLE INCLUDED IN BEP6 0 FRONT IO ASSEMBLY WITHOUT USB PORT 5433408 41 1 N N 3 SATA HARD DRIVE CABLE COMES WITH HDD 5433408 50 1 N N 4 POWER HARD DRIVE CABLE COMES WITH HDD 5433408 50 1 N N 5 DVR TO S...

Page 731: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 9 Renewal parts 9 81 9 17 6 1 BEP6 cables cont d Figure 9 3 BEP6 External Cables...

Page 732: ...BLE B XYZ CONTROLLER INNER FROG BRAKES 2 N Y C FROG BRAKE ARM GA200334 CABLE C INNER BRAKE OUTER BRAKE 2 N Y D MAIN CABLE ASSEMBLY 5272357 1 N Y D1 MAIN CABLE POWER AUDIO Part of 5272357 MAIN CABLE AS...

Page 733: ...BLE USB BEP TO XYZ MOTOR CONTROLLER 5194492 6 USB CABLE XYZ CONTROLLER 1 N Y M CABLE DVD eSATA FULLY SHIELDED 5270048 SATA CABLE DVD CABLE M J2 DVD 2 J3 DVD 1 1 or 2 N Y N BEP6 TO GFI AND CARDRACK BAC...

Page 734: ...E9 SERVICE MANUAL 9 84 Section 9 17 Internal Cables VIVID E9 R POWER CABLE BW PRINTER 5194493 USED ONLY IF PRINTER INSTALLED 1 N Y Table 9 48 BEP6 External Cables and Harnesses cont d sheet 3 of 3 IT...

Page 735: ...OFER GA200333 CABLE P J33 SUBWOOFER 1 N Y 2 POWER CABLE DVD 5266407 CABLE K J30 DVD 1 1 N Y 3 POWER CABLE DVD 5266407 CABLE K J29 DVD 2 1 N Y 4 MAIN CABLE POWER AUDIO SEE MAIN CABLE CABLE D1 PART OF M...

Page 736: ...4 SATA CABLE DVD 1 5270048 CABLE M J2 DVD 1 1 N Y 5 CABLE BEP 6 5391509 1 N Y 6 CABLE USB BEP6 MAIN POWER SUPPLY 5194492 8 7 POWER CABLE BEP 48V 5194495 CABLE I J1 MAIN POWER SUPPLY XYZ CONTROLLER 1...

Page 737: ...BEP FRONT PANEL ICHIRO 5193726 1 N Y 11 HARNESS BEP MOTHERBOARD IOBOARD YGGDRASIL IMPROVED AUDIO 5193725 3 1 N Y 12 Harness BEP Power Ygg GA200886 For GA200890 GA200900 and GB200003 1 N Y 13 HARNESS...

Page 738: ...ISION 5 VIVID E9 SERVICE MANUAL 9 88 Section 9 17 Internal Cables VIVID E9 14 Bracket Lock BEP Top Connectors GA200984 1 N Y Table 9 49 Back End Processor BEP cables sheet 4 of 4 cont d ITEM PART NAME...

Page 739: ...s Cables Table 9 50 Peripherals Cables ITEM PART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 POWER CABLE BW PRINTER 5194493 CABLE R MAIN POWER SUPPLY BW PRINTER 1 N Y 2 POWER CABLE DVD 5266407 CABLE K...

Page 740: ...ION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL 9 90 Section 9 17 Internal Cables VIVID E9 5 SATA CABLE DVD 5270048 CABLE M J3 DVD 2 1 N Y Table 9 50 Peripherals Cables ITEM PART NAME PART NUMBER DESC...

Page 741: ...ART NAME PART NUMBER DESCRIPTION QTY CRU FRU 1 ECG CABLE SET FC200389 CONTENT CABLE CABLE ECG NICOLAY ADAPTER RCA MALE BNC FEMALE ADAPTER RCA FEMALE 6 ADAPTER RCA FEMALE 3 1 Y Y 1 CABLE ECG MARQ AHA A...

Page 742: ...E9 Nameplate Periwinkle Upper OP Panel GA314992 For Pro configuration N Y 3 Vivid E9 Nameplate Bronze Upper OP Panel GA314812 New for BT 11 N Y 4 Vivid E9 Nameplate Silver Upper OP Panel 5255315 BT09...

Page 743: ...NCED OPTIONS 1 LVO CONTRAST IMAGING H45561MY 2 VASCULAR CONTRAST H45561MZ 3 AFI PRODUCTIVITY PACKAGE H45561GX 4 IMT H45561GY 5 Stress H45561NC 6 LV Mass Only H45561NA 7 4D Strain and LV Mass H45561NB...

Page 744: ...9 GA200618 Label DVR GA314384 Flex Cable DVI to SAMTEC Video Jumper 5428990 Cable DVR Audio BEP6 5435462 SATA Cable DVR to BEP6 MB 5439827 2 16 DVR option VIVID E9 For BEP5 GA200859 KIT INCLUDES Digit...

Page 745: ...DVR Option for Vivid E9 H45551NS GB200048 Includes parts for installing DVR into both BEP5 and BEP6 KIT INCLUDES Digital Video Recorder Circuit Board 5135840 DVI Flex DVR jumper Yggdrasil BEP 5197217...

Page 746: ...NETWORK OPTIONS 7 DICOM CONNECTIVITY PACKAGE H45551NR OTHER 8 DVR option VIVID E9 GA200859 KIT INCLUDES Digital Video Recorder Circuit Board 5135840 DVI Flex DVR jumper Yggdrasil BEP 5197217 DVI Flex...

Page 747: ...ch GA092904 1 N N 4 BT12 Vivid E9 User manual Danish GA092905 1 N N 5 BT12 Vivid E9 User manual Dutch GA092906 1 N N 6 BT12 Vivid E9 User manual English GA092907 1 N N 7 BT12 Vivid E9 User manual Esto...

Page 748: ...sh GA092927 1 N N 29 BT12 Vivid E9 User manual Turkish GA092928 1 N N 30 BT12 Vivid E9 Reference manual English GA092929 REFERENCE MANUAL ENGLISH 1 N N 31 Remote Software Download and Installation man...

Page 749: ...R GUIDE DANISH GA092771 1 N N 10 VIVID E9 USER MANUAL JAPANESE GA092767 1 N N 11 VIVID E9 USER MANUAL CHINESE SIMPLIFIED GA092768 1 N N 12 VIVID E9 USER GUIDE POLISH GA092772 1 N N 13 VIVID E9 USER GU...

Page 750: ...30 Remote Software Download and Installation manual GB094004 English 1 N N SERVICE MANUALS 1 VIVID E9 UNPACKING PACKING PROCEDURE ENGLISH GA050318 1 N N 2 VIVID E9 SERVICE MANUAL ENGLISH GA091568 1 N...

Page 751: ...ORTUGESE GA092409 1 N N 10 VIVID E9 USER MANUAL JAPANESE GA092410 1 N N 11 VIVID E9 USER MANUAL CHINESE SIMPLIFIED GA092411 1 N N 12 VIVID E9 USER MANUAL SWEDISH GA092412 1 N N 13 VIVID E9 USER MANUAL...

Page 752: ...D E9 SERVICE MANUAL 9 102 Section 9 22 Product Manuals for VIVID E9 31 VIVID E9 USER REFER MANUAL GA092429 REFERENCEMANUAL ENGLISH 1 N N Table 9 59 Product Manuals for VIVID E9 BT 09 cont d sheet 2 of...

Page 753: ...ish Dutch Estonia Finnish Greek Hungarian Indonesian Italian Japanese Korean Latvian Lithuanian Norwegian Polish Portugese Romanian Russian Slovenian Serbian Slovakian Spanish Swedish Turkish 1 N N Ta...

Page 754: ...r Manual Korean KX192084 1 N N 10 TEE Probes User Manual Serbian KX192240 1 N N 11 TEE Probes User Manual Bulgarian KX192241 1 N N 12 TEE probes User manual Croatian KX192410 1 N N 13 TEE PROBES USER...

Page 755: ...es User manual Indonesian KX192663 1 N N 12 TEE Probes Accessories User manual Italian KX192323 1 N N 13 TEE Probes Accessories User manual Japanese KX192326 1 N N 14 TEE Probes Accessories User manua...

Page 756: ...KX192060 1 N N 7 IO Probes User manual Norwegian KX192061 1 N N 8 IO Probes User manual Danish KX192062 1 N N 9 IO Probes User manual Polish KX192063 1 N N 10 IO Probes User manual Finnish KX192064 1...

Page 757: ...ire additional tasks and or inspections at a different frequency than listed in this manual 10 1 2 Purpose of this chapter This chapter describes Care Maintenance on the VIVID E9 and peripherals These...

Page 758: ...ng must occur regularly The same tests are performed during each period so that changes can be monitored over time and effective corrective action can be taken Testing results corrective action and th...

Page 759: ...u use your VIVID E9 for an average patient load 10 12 per day and not use it as a primary mobile unit which is transported between diagnostic facilities NOTE If conditions exist which exceed typical u...

Page 760: ...eal Probe Leakage Current Checks also after corrective maintenance Surgical Probe Leakage Current Checks also after corrective maintenance Functional Checks also after corrective maintenance Table 10...

Page 761: ...1568 REVISION 5 VIVID E9 SERVICE MANUAL Chapter 10 Care maintenance 10 5 Section 10 4 Tools required NOTE A list of required tools for servicing the VIVID E9 please refer to 8 2 5 Tools needed for ser...

Page 762: ...ription 1 Ask Listen Ask the customer if they have any problems or questions about the equipment 2 Power up Turn the VIVID E9 power on and verify that all fans and peripherals turn on Watch the displa...

Page 763: ...ation PW and CW if available Check the basic system controls that affect this mode of operation 4 M Mode Verify basic M Mode operation Check the basic controls that affect this mode of operation 5 4D...

Page 764: ...B W video page printer Clean heads and covers if necessary 3 Color Printer Verify hardcopy output of the Color video page printer Clean heads and covers if necessary 4 DICOM Verify that DICOM is func...

Page 765: ...cracks or damage 9 Op Panel Movement Verify ease of Operator Panel Operator Control Panel movement in all acceptable directions Ensure that it latches in position as required 10 Op Panel Lights Check...

Page 766: ...nnection receptacle Prior to cleaning turn OFF power to the VIVID E9 and disconnect the mains cable Table 10 7 VIVID E9 System General Cleaning sheet 1 of 3 Step Item Description 1 LCD Cabinet On a we...

Page 767: ...Clean the probe holders with warm water and a damp cloth to remove all traces of gel Soaking may be required to remove excess gel 7 System Cabinet On a weekly basis moisten a soft non abrasive folded...

Page 768: ...n TEE Probe Care Card Avoid too much moisture on the cloth towel as this may cause cleaning fluids to flow along the cable and into the connector handle Always clean the cable from the connector towar...

Page 769: ...d November 2010 Please refer to the respective air filter replacement and cleaning instructions 8 8 2 Rear Filter and handle type Bottom Filter replacement on page 8 190 or 8 8 3 Rear Air Filter repla...

Page 770: ...ury and equipment damage Failure to properly handle or maintain a probe may also void its warranty Any evidence of wear indicates the probe cannot be used Do a visual check of the probe pins and VIVID...

Page 771: ...a soft towel WARNING WARNING TO HELP PROTECT YOURSELF FROM BLOOD BORNE DISEASES WEAR APPROVED DISPOSABLE GLOVES THESE ARE MADE OF NITRILE DERIVED FROM VEGETABLE STARCH TO PREVENT ALLERGIC LATEX REACT...

Page 772: ...cleaner disinfectant as recommended in TEE Probe Care Card Avoid too much moisture on the cloth towel as this may cause cleaning fluids to flow along the cable and into the connector handle Always cle...

Page 773: ...a cloth or towel with only the smallest amont of cleaner disinfectant to wipe the connector 2 Avoid the connector pin area 3 Wipe off residual chemicals with a cloth or towel lightly moistened with c...

Page 774: ...tissue sprayed with a recommended disinfectant Use additional precautions e g gloves and gown when decontaminating an infected probe High level disinfection High level Disinfection destroys vegetative...

Page 775: ...ch was in contact with the germicide according to the germicide manufacturer s instructions 4 Wipe dry with a soft towel or air dry the probe Figure 10 2 Probe immersion levels A 3V D 4V D M5S D 6S 12...

Page 776: ...TER THAN 30 VOLTS 1 FOLLOW LOCK OUT TAG OUT PROCEDURES 2 TURN OFF THE BREAKER 3 UNPLUG THE VIVID E9 4 MAINTAIN CONTROL OF THE POWER PLUG 5 WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS T...

Page 777: ...uipment users must ensure that safety inspections are performed whenever damage is suspected and at least every 12 months in accordance with local authorities and facility procedures Do not use the VI...

Page 778: ...ISOLATION THE AMBER LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER IS TURNED OFF WHEN THE CORD IS STILL...

Page 779: ...Parts or accessories that contact the patient to perform their function For ultrasound equipment this includes transducers and ECG leads Type BF Body Floating or non conductive ultrasound probes which...

Page 780: ...the condition where the Neutral grounded supply conductor and the Grounding protective earth conductor are reversed If later tests indicate high leakage currents this should be suspected as a possibl...

Page 781: ...surements should be made under the test conditions specified in Table 10 9 on page 10 22 or Table 10 10 on page 10 22 as applicable Record the highest reading of current When using the Microguard or a...

Page 782: ...oring equipment and is grounded by touching some other grounded surface Measurements should be made under the test conditions specified in Table 10 9 on page 10 22 or Table 10 10 on page 10 22 as appl...

Page 783: ...n Figure 10 5 Test Circuit for Measuring Patient Lead Leakage POWER OUTLET H Black N White G Green Internal circuitry Insulating surface Polarity reversing switch Appliance power switch use both off a...

Page 784: ...6 7 Isolated Patient Lead Source Leakage Lead to Lead Select and test each of the ECG lead positions except ALL on the LEAD selector testing each to the power and ground condition combinations found i...

Page 785: ...rent is expected on each probe depending on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from differences in line voltage and test...

Page 786: ...ll have some amount of leakage current dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from differences in line voltage a...

Page 787: ...ighest current reading 9 Follow the test conditions and test limits described in Table 10 9 on page 10 22 or Table 10 10 on page 10 22 as applicable for every probe The test passes when all readings m...

Page 788: ...be or the VIVID E9 NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from...

Page 789: ...ime 3 2 B back end processor bi directional signals 5 65 block diagram 5 54 cables 9 79 description 5 52 5 55 expansion slots on motherboard 5 62 fan replacement 8 276 front module replacement 8 273 f...

Page 790: ...pherals 9 89 top console 9 74 XYZ controller 9 76 calibration touch screen 6 10 capture a screen image 7 12 card rack spare parts 9 56 care and maintenance 10 1 casters and brakes description 5 26 rep...

Page 791: ...dure warnings 1 32 data management moving all images 8 10 date and time adjustments 3 28 DC Offset calibration 6 11 definitions of Left Right Front and Back 8 5 department name 3 27 depth of ultrasoun...

Page 792: ...ter bottom cleaning 8 197 cleaning 10 13 rear filter cover replacement 8 51 foot rest bumper replacement 8 45 footswitch 3 36 5 97 forced power downpower forced power down 5 87 frame UI upper replacem...

Page 793: ...nventions 1 6 important precautions language policy i initiate a request for service RFS 5 6 input and output I O modules description 5 88 InSite ExC definitions 5 8 description 5 4 icon 5 4 6 11 menu...

Page 794: ...5 24 parts replacement 8 190 spare parts 9 52 main controller board replacement 8 108 Main LCD and Touch Screen backlight adjustments 6 5 main power supply 5 77 block diagram 5 78 cables 9 77 parts re...

Page 795: ...7 Patient I O Physio 5 88 patient I O assembly replacement 8 295 pedal dir lock replacement 8 217 mechanism replacement 8 214 release replacement 8 217 pedals 4 17 periodic maintenance inspections 10...

Page 796: ...ver and transmitter signal control 5 31 Receiver Board 5 41 receiver board replacement 8 232 Receiver Board GRX 5 41 receiver signal path 5 30 receiving and unpacking the equipment 3 3 receiving the u...

Page 797: ...rinter service 4 48 DHCP 3 44 digital vide recorder DVR 3 35 select offline mode 3 35 options 3 49 paperwork 3 50 reminders 3 2 warnings 3 2 shock and tilt indicators 3 4 shortcut keys 7 4 Shutdown 4...

Page 798: ...1 subsystem 5 29 transmitter board description 5 32 replacement 8 235 transportation damage x transportation box label carton box 3 9 wooden box 3 8 transporting the unit by vehicle 4 18 trouble image...

Page 799: ...re parts 9 54 when there s too much leakage current 10 32 when to load or reload the software 8 12 why do maintenance 10 2 width of ultrasound scanner 3 20 windows printer queues 4 48 woofer replaceme...

Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...

Page 801: ......

Page 802: ...2008 2010 2012 General Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...