GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 87
8-6-4-4
Install the LCD Arm assembly
Follow these steps to install the LCD Arm Assembly:
1.) Carefully install the LCD Arm assembly into position, first feeding the LCD Arm cables down through
the console opening.
2.) Push the LCD Mount Lock Handle into locked position.
Figure 8-94 LCD Arm installed onto the UI Frame Upper (Arm for 17” screen illustrated)
Figure 8-95 LCD Mount Lock Handle
DO NOT ADJUST THIS NUT!
LOCKED POSITION
LCD MOUNT LOCK HANDLE
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......