GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 37
8-4-10-7
Loading Software Patches
If a CD with a software patch is included, insert the disc in the drive, restart the VIVID E9 and follow the
on-screen instructions.
Please continue with
8-4-12 "Verifications after the software has been re-loaded" on page 8-37
.
8-4-11
Setup after Software loading
NOTE:
Use this procedure if all partitions on the HDD have been erased during the software load.
NOTE:
If only the software on C:\ has been replaced or updated, please go to:
after the software has been re-loaded" on page 8-37
1.) Restore the Patient Archive and System Configurations from the backup you made before the
software loading.
-
For instructions, please see “Data Backup and Restore” in the User Manual/User Guide.
2.) With your recordings from before the software loading, available, continue with the setup
instructions starting in:
Section 3-8 "Configuration" on page 3-27
. Correct the settings as needed.
3.) Continue with:
Section 3-10 "Connectivity setup" on page 3-41
. Based on your recordings, correct
the settings as needed.
4.) Continue with:
Section 3-11 "Options Setup" on page 3-49
. Based on your recordings, correct the
settings as needed.
5.) After a software load, you should always calibrate the Front End, as described in:
Offset Calibration (Front End Alignment)" on page 6-11
.
6.) Install the network printer (if any). Please refer to the Printer Driver Installation Manual.
8-4-12
Verifications after the software has been re-loaded
NOTE:
DO NOT use this procedure if all partitions on the HDD have been erased during the software
load, but refer to:
8-4-11 "Setup after Software loading" on page 8-37
NOTE:
Use this procedure if only the software on C:\ has been updated or reloaded.
1.) With your recordings from before the software loading available, continue with the setup instructions
starting in:
Section 3-8 "Configuration" on page 3-27
. Correct the settings if needed.
2.) Continue with:
Section 3-10 "Connectivity setup" on page 3-41
. Based on your recordings, correct
the settings if needed.
3.) Continue with:
Section 3-11 "Options Setup" on page 3-49
. Based on your recordings, correct the
settings as needed.
4.) After a software load, you should always calibrate the Front End, as described in:
Offset Calibration (Front End Alignment)" on page 6-11
.
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......