Section 8-4 - Reloading the software
8-4-9-3
Initiate software reload from the Recovery Console
Ensure that the VIVID E9 is powered down.
1.) Depress the
On/Off
button on the
Operator Panel
. The VIVID E9 starts.
2.) Press the
Esc
button, on the alphanumeric keyboard, multiple times until the
GRUB
menu is
displayed on the screen.
3.) On the
Grub
menu, select:
Recovery Console
.
The GE Healthcare Recovery Console is displayed.
4.) Call the
Online Center
(
OLC
).
Tell
OLC
that you need a “
Challenge Code
”, and on request, you must read the
Challenge String
on the screen.
OLC
will return a password that you need to type in as the
Response
.
5.) Enter the
Response Code
(
password
) you got from
OLC
.
Figure 8-2 GRUB menu
Figure 8-3 GE Healthcare Recovery Console (Example)
GNU GRUB version 0.97 (638K lower / 2095879K upper memory)
___________________________________________________________
Windows
Recovery Console
Use the | and | keys to select which entry is highlighted.
Press enter to boot the selected OS or 'p' to enter a
password to unlock the next set of features.
USE THE ARROW DOWN KEY
TO SELECT RECOVERY
CONSOLE, THEN PRESS
ENTER.
Enter the password (from OLC).
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......