7 - 20
Section 7-9 - Troubleshooting
7-9-9
System Temperature Too High
When the temperature inside the VIVID E9 increase, the fan speed will increase to cool down the
system. If the air filters become too dusty, a higher fan speed is required to keep the air stream at the
needed level. When the fan speed increase, the fan noise will also increase.
•
If the cooling air stream is insufficient to stabilize the interior temperature within the operating
margin, the system will stop operating.
-
Solution
: Clean or replace the filters.
•
If a fan is worn out, it may be noisy, or stop working.
-
Solution
: Replace the fan(s).