GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 5 - Components and functions (theory)
5 - 95
5-11-4-3
Input signals
•
Supply voltage of 5V and 12V from BEP power supply card.
The power voltages are regulated at +/-10%.
•
5V STDBY from the BEP power supply card for OPIO (Pass through).
•
PWR_ON signal from the Operator panel to the Motherboard (pass through).
•
Two USB ports for OP Panel (pass through), one USB for USB ‘hub’ and one USB for Patient IO
module.
•
Doppler Audio from GFI
•
Doppler audio mixed with Windows system sounds from motherboard
•
DVI (Digital Video) input from the video card or DVR Board.
•
AC Fail and PSON_N signal from the BEP PS.
5-11-4-4
Output signals
•
12V DC and 5V DC for Patient IO, DVD drive and High capacity drive.
•
5V STDBY from the BEP power supply card for OPIO (Pass through)
•
USB ports for Color and BW printer
•
3 x USB ports (Spare)
•
Audio Line In to the Motherboard
•
Audio to DVR board on Motherboard
•
Audio to the OP Panel for the speakers.
•
External Audio outputs
•
Audio to the sub woofer
•
DVI-D output for Main Monitor (Digital Video)
•
DVI-I output for External Monitor (only the Digital Video Signal is included)
•
Two external USB ports
•
Two spare internal USB ports
•
External Test connector
•
I
2
C bus clock and data signals to the Backplane and BEP PS.
5-11-5
Probe Connectors
The probe connectors are mounted on the Relay board.
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......