GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
5 - 92
Section 5-11 - Input and Output (I/O) modules
5-11-3
BEP6 I/O Board
5-11-3-1
General description
BEP6’s I/O board is the interface between the BEP and the rest of the system. It distrbutes USB to the
Rear IO, the OP and to internal units (printer and Patient IO). It distributes SATA to the DVD(s). It splits
video between the main monitor and rear video output. It also distributes audio to speakers and rear
connectors.
Figure 5-66 BEP6’s I/O Board Block Diagram
MAIN CABLE - VIDEO
MAIN CABLE -
POWER/AUDIO
SATA CABLE – DVD1
SATA CABLE – DVD2
MAIN CABLE - USB
MAIN CABLE - USB
USB CABLE – XYZ
CONTROLLER
USB CABLE –
BW PRINTER
USB CABLE - MAIN
POWER SUPPLY
POWER CABLE, DVD (2X)
AUDIO CABLE -
SUBWOOFER