Section 4-3 - Functional checks
4-3-12-2
Setup and Test a Printer Service
1.) Select Utility > Connectivity
If you get a pop-up asking you to log on, select ADM.
2.) Select the Service tab.
3.) In the comb box “Select Service to Add” select “Standard Print” and click on Add.
4.) In the right pane Properties “Combo Box” Select the printer you wish to test. Set any other
parameters you desire.
5.) In the left pane “Properties” Enter a name that describes the printer and configuration you just
selected in the right pane.
6.) Select the Button tab.
7.) Select one of the “Physical Print Buttons” that you want to configure.
8.) In the right pane click on the service name you just created in the Services Tab.
9.) Click on the “>>” button. This will place this service in the PrintFlow View for the printer button you
selected.
10.)Click on Save.
You have now configured a printer service and attached it to a print button.
Now you can test the printer by pressing the print button you just configured. If you configured it for 1
row and 1 column each time you press the print button you will get a print sent to the printer. If you
configured some other combination of rows or columns you will have to push the printer button multiple
times before a print is sent to the printer.
4-3-12-3
View the Windows Printer Queues
•
Go to
Utility
>
System
>
Peripherals
•
Click on
Properties
.
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......