GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 4 - General procedures and Functional checks
4 - 47
4-3-10
Back End Processor checks
•
If all the previous tests have been passed successfully, the Back End Processor is most likely OK.
4-3-11
Operator Panel Test
•
The Operator Panel is tested when the VIVID E9 is powered up as part of the start-up scripts, run
at every start-up.
4-3-12
Peripheral checks
4-3-12-1
Printer checks
The internal printer is controlled from the
P1
and
P2
keys on the VIVID E9’s Operator Panel.
The factory default is:
•
P1
for the internal printer
•
P2
for external (network) printer
Table 4-7
Printer checks
Step
Task
Expected Result(s)
1.
When scanning in 2D Color Mode, Press
Freeze
to
stop image acquisition.
Image scanning stops with the last picture on the
screen.
2.
Press
P1
on the Operator Panel
The image displayed on the screen is printed on the
assigned printer.
3.
Press
P2
on the Operator Panel
The image displayed on the screen is printed on the
assigned printer.
4.
Check if the print quality on the pictures from both
printers are of expected quality.
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......