Chapter 3 - System setup
3 - 27
Section 3-8
Configuration
3-8-1
Purpose of this section
This section describes how to configure the VIVID E9.
3-8-2
VIVID E9 configuration
3-8-2-1
Select System Settings screen
1.) Select
Config
(
F2
) and log on as
adm
, see
4-2-5 "Logging on to VIVID E9 as ‘ADM’" on page 4-15
.
2.) Select
System
and then select
Settings,
if needed.
3-8-2-2
Enter Location
Figure 3-10 Hospital and department name
Table 3-13 Location name
STEP
TASK
EXPECTED RESULT(S)
1
Select the Hospital field, see
, and type the name of
the hospital (max 64 characters).
The 24 first characters of this name are displayed on the scanning
screen’s title bar (after restart)
All 64 are displayed on the image properties on saved images (after
restart).
2
Select in the Department field, see
name of the department (max 64 characters).
This name will be displayed on the image properties on saved images as
soon as the unit has been restarted.
HOSPITAL NAME
DEPARTMENT
SETTINGS
SYSTEM