Chapter 3 - System setup
3 - 19
Section 3-6
Preparing for setup
3-6-1
Verify Customer Order
Compare items received by the customer to that which is listed on the delivery order. Report any items
that are missing, back ordered, or damaged.
3-6-2
Physical inspection
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to “Damage in Transportation“ on page
in the beginning
of this manual.
3-6-3
EMI protection
This unit has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
2-2-3 "EMI limitations" on page 2-7
for more information about EMI protection.