![GE Vivid 7 Service Manual Download Page 831](http://html1.mh-extra.com/html/ge/vivid-7/vivid-7_service-manual_4263692831.webp)
GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
Chapter 10 - Care & Maintenance
10 - 13
10-7-2
GEHC
Leakage Current Limits
The following limits are summarized for NFPA 99/UL2601-1/UL60601-1 (for USA) and IEC 60601-1
Medical Equipment Safety Standards. These limits are GEHC standards and in some cases are lower
than the above standards listed.
NOTE:
*Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the
part to determine the amount of current that will pass (or sink) to ground if a patient contacted
mains voltage.
The following tests are performed at the factory and should be performed at the site. These tests are:
grounding continuity, chassis leakage current, probe leakage current, and ECG leakage current. All
measurements are made with an electrical safety analyzer.
Table 10-12 Chassis Leakage Current Limits—Accessible Metal Surfaces
Country
Normal Condition
Open Ground
Reverse Polarity
Open Neutral
USA
N/A
0.3 mA
0.3 mA
N/A
Other
0.1 mA
0.5 mA
0.5 mA
0.5 mA
Table 10-13 Type BF Patient Leakage Limits - Non-Conductive (Floating) Surface and Cavity
Probes
Country
Normal
Condition
Open Ground
Reverse Polarity
Open Neutral
*Mains Applied
USA
0.1 mA
0.5 mA
0.5 mA
0.5 mA
5.0 mA
Other
0.1 mA
0.5 mA
0.5 mA
0.5 mA
5.0 mA
Table 10-14 Type CF Patient Leakage Limits - Surgical Probes and ECG Connections
Country
Normal
Condition
Open Ground
Reverse Polarity
Open Neutral
*Mains Applied
USA
0.01 mA
0.05mA
0.05 mA
N/A
0.05 mA
Other
0.01 mA
0.05 mA
0.05 mA
0.05 mA
0.05 mA
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......