GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
8-114
Section 8-8 - Peripherals
Section 8-8
Peripherals
8-8-1
B/W Video Printer Replacement Procedure
NOTE:
This section describes the installation and removal procedure for the Sony UP-895MDW B/W Video
Printer.
Figure 8-100 SONY B/W Video Printer
8-8-1-1
Tools
Use the appropriate flat and Phillips-type screw drivers.
8-8-1-2
Manpower
One person - 30 mins.
8-8-1-3
Preparations
1) Shut down the Vivid™ 4 ultrasound unit, as described in
.
2.) Remove the Connectors Panel cover as shown in the
Connector Panels Cover Removal Procedure
3.) Unpack the B/W Video Printer.
4.) Check that the B/W Video Printer AC voltage corresponds to that of your local voltage.
5.) Verify that all the DIP Switches on the rear panel of the B/W Video Printer are in the
Down
position