4 Making the Wiring Connections
You must wire the transmitter using the factory-supplied cable, which is
available in a variety of lengths.
Note:
If you need to lengthen cables, refer to
below to splice an
extension onto the existing cable. Be sure to connect positive to
positive and negative to negative.
Use the following steps to wire the transmitter to your system:
1.
Push the female connector end on the transmitter cable into the
mating male connector on the transmitter module. Make sure the
pins are properly aligned. Then, secure the connectors together by
sliding the metal sleeve on the cable over the connectors and
turning it clockwise until it is tight.
2.
Using the flying leads at the other end of the transmitter cable,
connect the transmitter to your power supply and data acquisition
system (DAS), as shown in
. Refer to
below for a description of the leads on the factory-supplied cable.
3.
Trim any unused leads back to the outer cable jacket, to remove the
bare tinned wire and prevent accidental short circuits.
The VeriDri is now ready for operation.
Table 1: Cable Lead Descriptions
Lead Color*
Connection Description
Blue
(+) 7 to 28 VDC
Brown
(
−
) 7 to 28 VDC
Shield
Connect to ground, if desired
*
The blue and brown leads also produce a current output
equivalent to 4 to 20 mA.
Summary of Contents for VeriDri
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 8: ...Preface vi VeriDri User s Manual no content intended for this page ...
Page 10: ...Contents viii VeriDri User s Manual no content intended for this page ...
Page 30: ... no content intended for this page ...
Page 31: ......