3.) Write down the following settings as it will be necessary to restore them at the end of the process:
-
TCP/IP address
-
Network printer (if present)
-
Wireless network settings (if present)
-
InSite ExC settings
8-3-2
Preparation and Notes for Software Upgrade Procedure
•
When
updating
the system software from a previous version, it is necessary to contact your local
OTR department to receive an appropriate software password. Make sure you specify the system
serial number located on the Venue Fit™ unit.
•
It is
recommended
to backup the data and configuration before starting the
update
procedure.
At the site, perform the following steps before you start the upgrade procedure:
1.) Perform Disk Management for all database records and select the "Copy" option
(for instructions, refer to the Venue Fit™ User Manual on your Document Media).
NOTE:
It is recommended that Disk Management be executed using the "move" option to a network
repository.
2.) Perform a full back-up for Patient Archive and System Configuration
(for instructions, refer to the Venue Fit™ User Manual on your Document Media).
3.) Write down the following settings as it will be necessary to restore them at the end of the process:
-
TCP/IP address
-
Network printer (if present)
-
Wireless network settings (if present)
-
InSite settings
WARNING
WARNING
REMOVE ALL EXTERNAL DEVICES SUCH AS PRINTERS AND USB CONNECTIONS
BEFORE STARTING THE UPDATE PROCEDURE.