DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
8-28
Chapter 8 - Replacement Procedures
P R E L I M I N A R Y
8-2-12
Casters Replacement Procedure.
8-2-12-1
Tools
Allen key 6 mm
Phillips screwdriver of appropriate size
8-2-12-2
Time Required
5 min
8-2-12-3
Preparations
1.) Shut down the Venue Fit™ ultrasound unit, as described in
.
2.) Lower the cart to its lowest position.
3.) Dismount Venue Fit™ system from cart.
8-2-12-4
Casters Removal Procedure
4.) Using a flat screw driver, carefully remove the screw cap.
5.) Using Allen key 6mm, remove the holding screw and detach the caster.
8-2-12-5
Casters Installation Procedure
1.) Install new caster and secure the holding screw.
2.) Place screw cap back into its place.
3.) Mount Venue Fit™ system on cart.
Figure 8-53 Removing Cater Cap and Holding Screw
Perform the checks listed in
-
section