March 2004
2-14
Programming Site Data
Entering Reynolds
Correction Data (cont.)
3.
The menu varies, depending on your selection in Step 2.
•
If you select
Off
, no further choices are available.
•
If you select
Single
, the UTX878 will select and automatically
display the
Kinematic Viscosity
. To change the value, press [ENT].
Use the arrow keys to change the value (available in document
#914-004,
Sound Speeds and Pipe Size Data
), and press [ENT].
•
If you select
Table
, the screen displays three options:
Units, Rows
Used
and
Edit Table.
a.
If you scroll to
Units
and press [ENT], the screen displays three
more options:
Velocity, Soundspeed
and
Diagnostic
. If you
select either
Velocity
or
Soundspeed
and press [ENT], the
screen displays the measurement units (either metric or
English). Press [ESC] or [ENT] to return to the previous
screen. But if you scroll to
Diagnostic
and press [ENT], the
meter asks for the type of signal to be used, Signal Strength
Up
or Signal Strength
Dn
. Scroll to the appropriate signal, and
press [ENT].
b.
If you select
Rows Used
, the program asks for the number of
rows
you wish to use. Enter the desired number (from 2 to 20)
and press [ENT].
c.
If you select
Edit Table
, the table opens with a series of
rows
.
Scroll to the desired row, and press [ENT].
d.
For each row, the screen displays the Reynolds Correction
number (X) and the Kinematic Viscosity (KV). If you wish to
change either value, scroll to the value and press [ENT]. Use
the arrow keys to change the value (available in document
#914-004,
Sound Speeds and Pipe Size Data
), and press
[ENT].
e.
Repeat steps c and d until you have programmed all of your
available data (from 2 to 20 rows).
Press [ESC] until you return to the PROG menu, or continue pressing
[ESC] until the display screen reappears.
Summary of Contents for UTX878 Ultrasonic
Page 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Page 8: ...Chapter 1 ...
Page 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Page 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Page 23: ...Chapter 2 ...
Page 45: ...Chapter 3 ...
Page 52: ...Chapter 4 ...
Page 60: ...Chapter 5 ...
Page 65: ...Chapter 6 ...
Page 75: ...Chapter 7 ...
Page 80: ...Appendix A ...
Page 85: ...Appendix B ...
Page 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Page 88: ...Appendix C ...
Page 89: ...Data Records Site Data C 1 ...
Page 93: ...Appendix D ...
Page 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...