Troubleshooting: Problems
Error Messages
The following table describes error messages that may appear and how to
resolve the problem.
Message
Possible Reason/Solution
“WARNING: The EEPROM data was found to be either INVALID or
uninitialized. GE factory defaults will be stored in both the EEPROM
and the monitor’s configuration memory. You will be required to re-
enter the network configuration, re-enable any password protected
features and restore all monitor settings and site-specific defaults.”
1. Restore Ethernet address and IP address.
2. Power cycle.
3. Access Boot Loader and restore all monitor settings and site-
specific defaults via Set Configuration and Options Menu.
4. If error message persists, replace the main processor FRU.
“WARNING: THIS VERSION OF BOOT CODE IS NOT
COMPATIBLE WITH THE VERSION OF MAIN CODE CURRENTLY
STORED IN FLASH. PLEASE UPDATE THE BOOT CODE.”
1. Power cycle.
2. Reload Boot Code.
3. If problems persists, replace the main processor FRU.
“Boot Flash test FAILED.”
“ERROR: The Boot Code stored in Flash is not valid. Main Code
cannot be loaded until valid Boot Code exists.”
“Main Flash test FAILED.”
“ERROR: The Main Code stored in flash is not valid.”
“Static RAM test FAILED.”
“ERROR: The SRAM memory test failed. Main Code will not be
loaded unless this test passes. Reboot the monitor to repeat testing.”
1. Power cycle.
2. If problems persists, replace the main processor FRU.
“Real Time Clock FAILED - will not start.”
“WARNING: The real time clock chip is not running. Main Code
cannot be loaded until this chip is started. Attempting to start real time
clock...”
Followed by either:
“The real time clock was started. Select Start Patient Monitoring to
load and execute Main Code.”
Or
“ERROR: Unable to start the real time clock.”
PREVENTIVE MAINTENANCE REQUIRED
Perform all of the maintenance procedures.
Internal lithium battery is LOW.
This message displays during power up. After the power-up tests, the
following message displays in the BOOT LOADER program:
ONE OR MORE POWER-UP TESTS HAVE FAILED ** SERVICE
MAY BE REQUIRED **
DO YOU WISH TO CONTINUE MONITORING ANYWAY?
The main processor battery may be depleted. Remove this
monitor from service and contact GE.
The battery switch on the main processor FRU may be off.
If problems persists, replace the main processor FRU.
Summary of Contents for Transport Pro
Page 1: ...Transport Pro Patient Monitor Service Manual Software Version 2 2024579 018 Revision A ...
Page 9: ...Revision A Transport Pro 1 1 2012659 042 1 Introduction ...
Page 12: ...1 4 Transport Pro Revision A 2012659 042 Introduction Safety Message Definitions ...
Page 13: ...Revision A Transport Pro 2 1 2012659 042 2 Equipment Overview ...
Page 28: ...2 16 Transport Pro Revision A 2012659 042 Equipment Overview Theory of Operation ...
Page 29: ...Revision A Transport Pro 3 1 2012659 042 3 Installation ...
Page 33: ...Revision A Transport Pro 4 1 2012659 042 4 Configuration ...
Page 38: ...4 6 Transport Pro Revision A 2012659 042 Configuration Monitor Settings ...
Page 39: ...Revision A Transport Pro 5 1 2012659 042 5 Maintenance ...
Page 47: ...Revision A Transport Pro 6 1 2012659 042 6 Troubleshooting ...
Page 57: ...Revision A Transport Pro 7 1 2012659 042 7 Ordering Parts ...
Page 59: ...Revision A Transport Pro 7 3 2012659 042 Ordering Parts Exploded View Exploded View 622A ...
Page 62: ...7 6 Transport Pro Revision A 2012659 042 Ordering Parts Hardware ...
Page 63: ...Revision A Transport Pro 8 1 2012659 042 8 Disassembly ...
Page 83: ...Revision A Transport Pro 9 1 2012659 042 9 Checkout ...
Page 97: ...Revision A Transport Pro A 1 2012659 042 A Technical Specifications ...
Page 100: ...A 4 Transport Pro Revision A 2012659 042 Technical Specifications External Power Supply ...
Page 101: ...Revision A Transport Pro B 1 2012659 042 B Electromagnetic Compatibility ...
Page 109: ......